K133631 · Spectranetics Corporation · DRE · Mar 27, 2014 · Cardiovascular
Device Facts
Record ID
K133631
Device Name
LEXIPLIANT DILATOR SHEATH SET
Applicant
Spectranetics Corporation
Product Code
DRE · Cardiovascular
Decision Date
Mar 27, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1310
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LexiPliant Dilator Sheath Set is intended for use in patients requiring the percutaneous dilatation of tissue to facilitate the removal of cardiac leads, indwelling catheters, and foreign objects.
Device Story
LexiPliant Dilator Sheath Set consists of inner and outer polymer sheaths used individually or in telescoping fashion. Inner sheath features beveled tips; outer sheath features one beveled and one blunt tip. Device introduced at cardiac pacemaker or defibrillator pocket site; advanced over lead or foreign object. Rotation and progression of sheaths dilate tissue binding the object within vasculature. Once tissue is dilated, object is removed via traction. Used in clinical settings by physicians. Benefits include facilitated removal of indwelling cardiac devices and foreign objects.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including dimensional verification, tensile, cantilever bend, dilatation, tip compression, torque, sheath cycle, kink fatigue, radiopacity, and biocompatibility testing (including in vivo dog model for thrombogenicity).
Technological Characteristics
Polymer sheath set; telescoping design. Features beveled and blunt tips. Sterilization via product adoption equivalency per AAMI TIR:28-2009. Biocompatibility testing per ISO standards. Mechanical testing includes tensile, torque, and fatigue assessments.
Indications for Use
Indicated for patients requiring percutaneous tissue dilatation to facilitate removal of cardiac leads, indwelling catheters, and foreign objects.
Regulatory Classification
Identification
A vessel dilator for percutaneous catheterization is a device which is placed over the guide wire to enlarge the opening in the vessel, and which is then removed before sliding the catheter over the guide wire.
K142546 — TightRail Rotating Dilator Sheath/TightRail Mini Dilator Sheath · Spectranetics, Inc. · Sep 25, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
### MAR 2 7 2
510(k) Summary, K133631, page
**Spectranetics**
*Always Reaching Farther*
## 510(k) Summary
This 510(k) summary was prepared in accordance with 21 CFR 807.92(c)
Prepared on March 11. 2014
| 510(k) Submitter / Holder: | Spectranetics<br>9965 Federal Drive<br>Colorado Springs, CO 80921.3617<br>Establishment Registration No: 3007284006 |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Ms. Pharoah Garma<br>Regulatory Affairs Manager<br>Office: 719.447.2385<br>Fax: 719.447.2040<br>Email: Pharoah.Garma@SPNC.com |
#### Subject Device
| Device Trade Name: | |
|----------------------------|--|
| Device Common Name: | |
| Device Class: | |
| Classification Regulation: | |
| Regulation Description: | |
| Product Code: | |
| 510(k) Type: | |
| Model Numbers: | |
LexiPliant™ Dilator Sheath Set Dilator Sheath 21 CFR 870 1310 Vessel dilator for percutaneous catheterization DRE Traditional Standard: 550-008, 550-010, 550-011, 550-013 Long: 555-508, 555-510, 555-511, 555-513
#### Predicate Device
The LEXiPliant Dilator Sheath Sets were compared to the following legally marketed predicate device:
| 510(k) Number: | K0923 |
|---------------------|---------|
| Manufacturer: | Spectra |
| Trade Name: | VisiShe |
| Device Common Name: | Dilator |
78 (cleared on November 02, 2009) anetics Corp. eath™ Dilator Sheath Sheath
#### Device Description
The LexiPliant Dilator Sheath Set family consists of an inner and outer polymer sheath that may be used individually or as a set in a telescoping fashion. The inner sheath terminates with beveled tips. The outer sheath terminates with one beveled tip and one blunt tip. The sheaths are designed to dilate tissue while facilitating the removal of cardiac leads, indwelling catheters and foreign objects. The device is introduced at a cardiac pacemaker or defibrillator pocket's implantation site and then advanced over the lead or object that is to be extracted. The rotation and progression of the inner and outer sheaths causes dilation of the tissue binding the lead or object within the vasculature. Upon dilation of the surrounding tissue, the lead or object can be removed by traction.
#### Intended and Indications for Use
The LexiPliant Dilator Sheath Set is intended for use in patients requiring the percutaneous dilatation of tissue to facilitate the removal of cardiac leads, indwelling catheters, and foreign objects.
{1}------------------------------------------------
#### Technological Characteristics
The LexiPliant Dilator Sheath Set features similar design and functional characteristics as the predicate device (K092378 - VisiSheath Dilator Sheath Set). The results of design verification and validation testing demonstrate that the subject device is as safe and clinically effective as the predicate device.
#### Performance Data
The following testing was conducted to validate and verify that the subject device met was substantially equivalent to the predicate device:
Design Verification and Validation Testing
- . Dimensional Verification
- Tensile .
- Cantilever Bend .
- . Dilatation
- Tip Compression .
- Torque Tip �
- Sheath Cycle .
- Kink Fatigue .
- Radiopacity ●
#### Sterilization
- . Product adoption equivalency per AAMI TIR:28-2009
#### Biocompatibility
- Physiochemical .
- Cytotoxicity ●
- Sensitization .
- Irritation/Intracutaneous Reactivity .
- Acute Systemic Toxicity .
- C3a and SC5b-9 Complement Activation .
- Indirect and direct Hemolysis .
- In Vivo Thromobogenicity-Dog Model .
- Material Mediated Pyrogenicity .
- . Genotoxicity – Ames Test
#### Preclinical and Clinical Data:
Preclinical and clinical data was not required to demonstrate substantial equivalence. The design characteristics of the subject device are similar to the predicate. The design verification test results demonstrated that the subject device is as safe and clinically effective as the predicate device.
#### Substantial Equivalence
Based on the similarities in design between the subject and predicate device, and the performance data, the LexiPliant Dilator Sheath Set is substantially equivalent to the VisiSheath Dilator Sheath Set (K092378).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and tail. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 27, 2014
Specranetics Corporation % Pharoah Garma Regulatory Affairs Manager 9965 Federal Drive Colorado Springs, CO 80921-3617
Re: K133631
> Trade/Device Name: LexiPliant™ Dilator Sheath Set Regulation Number: 21 CFR 870.1310 Regulation Name: Vessel dilator for percutaneous catheterization Regulatory Class: Class II Product Code: DRE Dated: February 25, 2014 Received: February 26, 2014
Dear Ms. Garma,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Pharoah Garma
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K133617
Device Name LexiPliant™ Dilator Sheath Set
Indications for Use (Describe)
The LexiPlianTM Dilator Sheath Set is intended for use in patients requiring the perculaneous dilatation of tissue to facilitate the removal of cardiac leads, indwelling catheters, and foreign objects.
# Bram D. Zuckerman -S 2014.03.27 16:38:50 -04'00'
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.