K170671 · Pressure Products Medical Device Manufacturing, LLC · DRC · Oct 20, 2017 · Cardiovascular
Device Facts
Record ID
K170671
Device Name
SafeSept Transseptal Guidewire
Applicant
Pressure Products Medical Device Manufacturing, LLC
Product Code
DRC · Cardiovascular
Decision Date
Oct 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1390
Device Class
Class 2
Indications for Use
The SafeSept is indicated for use in procedures where access to the transseptal technique is desired. The SafeSept Transseptal Guidewire is intended for single use only.
Device Story
SafeSept Transseptal Guidewire is a 0.014-inch diameter, 135cm long nitinol wire used in healthcare facilities for transseptal heart procedures. Inserted via femoral vein, it is advanced to the fossa ovalis supported by a needle and dilator. The sharp distal tip perforates the fossa ovalis; once unsupported, the wire becomes atraumatic. A radiopaque coil allows fluoroscopic visualization in the left atrium, while proximal marker bands indicate tip position relative to the needle. The device guides the needle, dilator, and introducer from the right to the left atrium. It benefits patients by facilitating safe transseptal access for cardiac interventions. Used by physicians in hospital settings.
Clinical Evidence
Bench testing only. Testing included visual/dimensional inspection, particulate testing (USP <788>), surface finish, corrosion resistance, tensile strength, fracture/flex testing, curve integrity, torque testing, and radiopacity. Biocompatibility testing performed per ISO 10993 standards, including cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, and hemocompatibility (including in-vivo dog thromboresistance).
Technological Characteristics
Nitinol super-elastic wire; 0.014" diameter; 135cm length. Features radiopaque coil and printed marker bands. Sterilized via 100% ethylene oxide (ANSI/AAMI/ISO 11135-1) to SAL 10^-6. Single-use. Mechanical device; no software or electronics.
Indications for Use
Indicated for patients requiring access to the left atrium via transseptal technique. For single use only.
Regulatory Classification
Identification
A trocar is a sharp-pointed instrument used with a cannula for piercing a vessel or chamber to facilitate insertion of the cannula.
Special Controls
*Classification.* Class II (special controls). Except for trocars that are reprocessed for multiple use, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K220414 — VersaCross Connect Transseptal Dilator · Baylis Medical Company, Inc. · May 9, 2022
K241042 — SafeSharp Transseptal Needle (CLICK071, CLICK171, CLICK098, CLICK198) · Pressure Products Medical Device Manufacturing, LLC · May 17, 2024
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 20, 2017
Pressure Products Medical Device Manufacturing LLC Andrew Armour Managing Director 1 School Street Morton, Pennsylvania 19070
Re: K170671
Trade/Device Name: Safesept Transseptal Guidewire Regulation Number: 21 CFR 870.1390 Regulation Name: Trocar Regulatory Class: Class II Product Code: DRC Dated: September 20, 2017 Received: September 21, 2017
Dear Mr. Armour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Nicole G. Ibrahim -S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170671
Device Name SafeSept Transseptal Guidewire
Indications for Use (Describe)
The SafeSept is indicated for use in procedures where access to the transseptal technique is desired. The SafeSept Transseptal Guidewire is intended for single use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Submitter
Pressure Products Medical Device Manufacturing LLC. 1 School Street Morton, PA 19070 Phone: 610-285-9858 Fax: 610-285-9859
Contact Person: Andrew Armour Prepared: October 19, 2017
#### Identification of the Device
Proprietary-Trade Name: SafeSept® Transseptal Guidewire Device Class: Class II Classification Name: Trocar (CFR 870.1390) Common/Usual Name: Transseptal Guidewire Product Code: DRC
#### Equivalent Legally Marketed Devices
Oscor, SafeSept® Transseptal Trocar Guidewire, K081986
#### Description of the Devices
The SafeSept® Transseptal Guidewire is 0.014″ in diameter and 135cm long and consists of a flexible memory wire material called nitinol. The components of the SafeSept include the 0.014″ transseptal guidewire, dispenser and straightener, and sterile packaging and labeling. There is one model number for the SafeSept, SS-135. The SafeSept is sterilized by 100% ethylene oxide cycle and is for single-use only. The SafeSept is used in a healthcare facility/hospital.
The SafeSept is used in transseptal procedures to gain access to the left atrium through the right side of the heart. The device is used by inserting the device through the femoral vein. The SafeSept is then advanced to the fossa ovalis with the support of the needle and dilator. The SafeSept's sharp distal tip is able to perforate through the fossa ovalis, a thin wall between the right and left atrium, when it is supported by a transseptal needle and dilator. When the wire is no longer supported, it is atraumatic and operates as a typical 0.014″ diameter guidewire. A radiopaque coil located along the shaft allows for fluoroscopic visualization of the wire within the left atrium and proximal marker bands help determine the location of the guidewire tip in relation to the tip of the needle. The guidewire's duration in the body is less than 24 hours.
#### Indications for Use
The SafeSept® is indicated for use in procedures where access to the left atrium via the transseptal technique is desired. The SafeSept Transseptal Guidewire is intended for single use only.
The Indications for Use statement for the SafeSept transseptal guidewire device is identical to the predicate device. Both devices have the same intended use and are used in conjunction with a transseptal needle to create the primary puncture in the interatrial septum. The devices guide the needle, dilator, and introducer through the septum from the right side of the heart to the left side.
{4}------------------------------------------------
# Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of this device are the predicate device. The main difference between the subject device and the predicate device is the length of the guidewire. The predicate SafeSept device is 120cm in length and the currently marketed SafeSept is 135cm. The subject and predicate device are based on the follow technological elements:
- The guidewire is manufactured from nitinol wire that is super-elastic so that it can have a 'J' curve with a sharp tip
- Radiopaque coil used by the physician to locate SafeSept and to guide supporting devices like the transseptal needle, dilator, and sheath across the fossa
- . Device inserted through the transseptal needle so that it can be used in the transseptal procedure
- . Creates the primary puncture of the fossa with the sharp tip
- . Coil is used to dilate the fossa further so that the needle can then be inserted through the fossa rather then used as the primary puncture
- . Printed markers are along the body for the physician to use as a guide to understand how far the SafeSept is in relation to the supporting devices
## Performance Testing
The following performance tests were performed in support of substantial equivalence to the predicate:
- Visual and dimensional inspection
- Particulate testing (USP <788>, Light Obscuration Method) ●
- Surface finish assessment
- Corrosion resistance
- Guidewire tensile strength
- Fracture and Flex testing
- Curve Integrity testing
- Torque Testing
- Packaging testing
- Radiopacity assessment
In addition, the sterilization conditions have been validated in accordance with ANSI/AAMI/ISO 11135-1, Sterilization of health care products - Ethylene oxide - Part 1: Requirements for development, validation and routine control of sterilization process for medical device is sterilized to a SAL of 10-6. The results of the testing demonstrated the subject device performed equivalently to the predicate device.
## Biocompatibility Testing
The biocompatibility evaluation for the SafeSept was conducted in accordance with FDA 510(k) Memorandum- #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing' " June 16, 2016, and International Standard ISO 10993-"Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by the FDA. The following tests were performed:
- ISO 10993-5 Cytotoxicity
- ISO 10993-10 Sensitization
- ISO 10993-10 Irritation/Intracutaneous Reactivity .
{5}------------------------------------------------
- ISO 10993-11 Acute Systemic Toxicity ●
- . ISO 10993-11 Pyrogenicity
- ISO 10993-3 Genotoxicity ●
- . ISO 10993-4 Hemocompatibility - ASTM Hemolysis Complete
- ISO 10993-4 Hemocompatibility - In-vivo Dog Thromboresistance
- ISO 10993-4 Hemocompatibility Complement Activation Complete with C3a & SC5b-9 ●
The SafeSept transseptal guidewire is considered an external communicating device with circulating blood contact and limited exposure (less than 24 hours). The SafeSept met the requirements set forth in ISO-10993.
## Conclusion
When compared to the predicate device, the SafeSept transseptal guidewire is substantially equivalent in design, technological characteristics, materials and performance testing.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.