The FlexCath Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The sheath deflection facilitates catheter positioning.
Device Story
FlexCath Steerable Sheath & Dilator is a deflectable catheter introducer; used to facilitate percutaneous placement of catheters into vasculature and heart chambers. Device consists of a shaft and handle; includes a dilator. Operated by physicians in clinical settings. Sheath deflection allows for precise catheter positioning. Subject device incorporates modifications to the 12F sheath tip and valve assembly. Performance verified through mechanical, bench, and animal testing to ensure tip robustness, hemostasis, and deflection capabilities.
Clinical Evidence
Bench and animal testing only. Testing included hemostasis valve leak testing, hub-to-shaft tensile testing, sheath tip robustness after multiple deployments, air aspiration/flushing, tip strength, and deflection testing per ISO 11070. Biocompatibility testing included USP 33 <661> physiochemical testing, ISO 10993-5 cytotoxicity, and ISO 10993-4 hemolysis, supplemented by leveraged data from the predicate device.
Technological Characteristics
Steerable catheter introducer; 12F sheath size. Materials evaluated per ISO 10993 (biocompatibility) and USP 33 <661>. Mechanical testing per ISO 11070. Sterilization cycle validated per AAMI TIR28:2001. Features deflectable shaft and hemostasis valve assembly.
Indications for Use
Indicated for percutaneous catheter introduction into the vasculature and heart chambers to facilitate catheter positioning.
Regulatory Classification
Identification
A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.
K112541 — Z FLEX-270TM STEERABLE SHEATH · Great Batch Medical · Feb 7, 2012
Submission Summary (Full Text)
{0}------------------------------------------------
1 2010 SEP
:
#### 510(k) Summary - K102176 4.0
| Submitter: | Medtronic CryoCath LP<br>16771 Chemin Ste-Marie<br>Kirkland, Quebec<br>H9H 5H3, CANADA |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kari Lahti<br>Regulatory Affairs Program Manager<br>Telephone: 763.526.1766<br>Fax: 651.367.0575 |
| Date Prepared: | August 2, 2010 - Revised August 31, 2010 |
| Trade Name: | FlexCath® Steerable Sheath & Dilator |
| Classification: | Class II<br>Steerable Catheter, 21 CFR § 870.1280 |
| Product Code: | DRA |
| Predicate Device(s): | The subject device is equivalent to the following devices:<br>• K081049 - FlexCath Steerable Sheath & Dilator, Models<br>3FC10, 3FC12 |
| Device Description: | The FlexCath Steerable Sheath is a deflectable catheter introducer<br>used to facilitate placement of a catheter through the skin into the<br>artery or vein.<br>It is comprised of the following two (2) main sections: the shaft and<br>the handle.<br>A dilator is included with each sheath.<br>This application addresses changes to the 12F FlexCath sheath tip<br>and valve assembly. |
| Intended Use: | The FlexCath Steerable Sheath is intended for percutaneous<br>catheter introduction into the vasculature and into the chambers of<br>the heart. The sheath deflection facilitates catheter positioning. |
| Functional and<br>Safety Testing: | To verify that the modified device design met its functional and<br>performance requirements, representative finished sterilized<br>samples of the subject device underwent the following in vivo, ex<br>vivo animal testing, biocompatibility and mechanical testing in<br>accordance with applicable industry standards and FDA guidance.<br>Testing included:<br>• Hemostasis valve leak testing (per ISO 11070)<br>• Hub-to shaft tensile (per ISO 11070) |
- Shaft leak testing (per ISO 11070)
.
:
{1}------------------------------------------------
- · Assessment of the force required to detach the dilator luer lock from the sheath valve cap
- · Force required to remove the dilator from the sheath
- · Tensile testing of bond joints (in addition to those required by ISO 11070) including:
- valve hub to valve cap o
- o hub body to extension
- tube bond and stopcock to extension tube joints o
- · Sheath tip robustness testing after multiple catheter deployments/retractions, both in bench and animal models
- · Air aspiration and flushing ability
- · Tip Strength
- · Deflection testing
- · Assessed impact of changes in relation to current sterilization cycle in accordance with AAMI TIR28:2001.
#### Biocompatibility Testing:
Biocompatibility of the modified hub material for the subject device included:
- · Physiochemical testing USP 33 <661>
- · Cytotoxicity testing per ISO 10993-5 (2009)
- Hemolysis testing (human activated and indirect) per ISO 10993-4 (2002 Amd 2006)
Leveraged biocompatibility of the predicate device included:
- · Kligman Maximization Test per ISO 10993-10 (2002)
- · Rabbit Pyrogen Test -- Material Meditated per ISO 10993-11 (2006)
- · Intracutaneous Test per ISO 10993-10 (2002)
- · Systemic Injection Test per ISO 10993-11 (2006)
- · Complement Activation Assay per ISO 10993-4 (2002)
{2}------------------------------------------------
- UPTT per ISO 10993-4 (2002) •
- · Thrombogenicity In vivo canine per ISO 10993-4 (2002)
The changes to the subject FlexCath Steerable Sheath do not affect · Conclusion: the intended use of the device, do not alter the fundamental scientific technology of the device, or raise new issues of safety or effectiveness. The FlexCath Steerable Sheath is therefore · substantially equivalent to the predicate FlexCath Steerable Sheath.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three overlapping wings, representing the department's mission to protect the health of all Americans. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medtronic CyroCath LP c/o Ms. Kari Lahti Regulatory Affairs Program Manager 16771 Chemn St-Marie Kirkland, Quebec Canada H9H 5H3
SEP 1 2010
K102176 Re:
Trade/Device Name: FlexCath® Steerable Sheath & Dilator Common Name: Steerable Catheter Regulation Number: 21 CFR 870.1280 Regulatory Class: II Product Code: DRA Dated: August 2, 2010 Received: August 3, 2010
#### Dear Ms. Lahti:
We have reviewed your Section 510(k) premarket notification of intent to market the devices we nave reviewed your bection 510(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrease) to tegans to the Medical Device Amendments, or to commerce prices that have been reclassified in accordance with the provisions of the Federal recomments (DMA) and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosment Act (Act) that to not require approvince to the general controls provisions of the Act. The Y ou may, merelore, market the devices of the more of the manual registration, listing of
general controls provisions of the Act include requirements for annual register and general controls provisions of the rice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Prease note. CDICT does not overance interesting must be truthful and not misleading.
{4}------------------------------------------------
#### Page 2 - Ms. Kari Lahti
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Zuckerman, M.D. Bram D Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# K102176
# 3.0 Indications for Use Statement
510(k) Number (if known): K_102176
Device Name: FlexCath® Steerable Sheath and Dilator
#### Indications for Use:
The FlexCath Steerable Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart. The Sheath deflection facilitates catheter positioning.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Division of Cardiovascular Devices
510(k) Number K102176
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.