VISTA BRITE TIP CATHETERS
Device Facts
| Record ID | K963962 |
|---|---|
| Device Name | VISTA BRITE TIP CATHETERS |
| Applicant | Cordis Corp. |
| Product Code | DQY · Cardiovascular |
| Decision Date | May 7, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1250 |
| Device Class | Class 2 |
Indications for Use
The guiding catheter is intended for use for intravascular introduction of interventional/diagnostic devices into the coronary, peripheral and neurovasculature.
Device Story
Single-lumen guiding catheter; nylon body reinforced with stainless steel braid wire; variable durometer nylon transition segments for stiffness gradient; soft pellethane tip with radiopaque filler. Used by physicians in clinical settings for intravascular access; facilitates delivery of interventional or diagnostic devices. Device provides trackability and support during procedures.
Clinical Evidence
Bench testing only; biocompatibility testing successfully completed.
Technological Characteristics
Single-lumen catheter; nylon body; stainless steel braid wire reinforcement; variable durometer nylon transition segments; pellethane tip with radiopaque filler. Non-powered, mechanical device.
Indications for Use
Indicated for patients requiring intravascular introduction of interventional or diagnostic devices into coronary, peripheral, or neurovasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
- Cordis Vista Brite Tip
- Cordis Endovascular Systems, ENVOY Guiding Catheters
Related Devices
- K965211 — VISTA BRITE TIP CATHETERS · Cordis Corp. · Mar 11, 1997
- K962830 — VISTA BRITE TIP CATHETERS · Cordis Corp. · Oct 17, 1996
- K971572 — VISTA BRITE TIP CATHETERS · Cordis Corp. · Jul 14, 1997
- K962362 — CORDIS ENVOY & VISTA BRITE TIP GUIDING CATHETERS · Cordis Corp. · Aug 8, 1996
- K992673 — VISTA BRITE TIP CATHETERS · Cordis Corp. · Oct 12, 1999