← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K250972

# Primum Hydrophilic Guiding Catheter (K250972)

_Pendracare · DQY · Jun 29, 2025 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K250972

## Device Facts

- **Applicant:** Pendracare
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Jun 29, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Guiding Catheter is intended to be used in the coronary or peripheral vascular system.

## Device Story

Flexible, single-use percutaneous catheter; reinforced body with stainless steel braid; hydrophilic coating; luer hub; strain relief; intermediate and soft distal tip. Available in 5F, 6F, 7F, 8F sizes with various preformed shapes (Judkins, Amplatz, etc.). Optional in-line side holes for vessel perfusion. Operated manually by physicians in clinical settings. Inserted via guidewire/introducer sheath to target vascular location. Provides conduit for diagnostic/therapeutic devices. Facilitates vessel access and device placement; aids in clinical procedures within coronary/peripheral vasculature.

## Clinical Evidence

Bench testing only. Performance verified via ISO 10993 biocompatibility testing (haemocompatibility, cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity) and in-vitro mechanical testing (shape conformance, kink resistance, radial stiffness, coating integrity, torque strength, pull force, radiopacity). No new clinical data generated; relies on clinical post-market surveillance data from predicate devices.

## Technological Characteristics

Reinforced catheter body (inner liner, stainless steel 304V braid, outer jacket). Materials: Grilamid TR55LX/L25, Vestamid L2140/D16/E62-S3, Pebax 3533 SA01, BiOCl. Hydrophilic coating. Dimensions: 5F-8F. Sterilization: Ethylene Oxide. Manual operation. No software/electronics.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Convey 5F Guiding Catheters ([K120585](/device/K120585.md))
- Convey 6F Guiding Catheters ([K120585](/device/K120585.md))
- Convey 7F Guiding Catheters ([K132197](/device/K132197.md))
- Convey 8F Guiding Catheters ([K132197](/device/K132197.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

June 29, 2025

PendraCare
Erendira Rodriguez
Director of Quality Assurance and Regulatory Affairs
Van der Waalspark 20-22
Leek, 9351VC
Netherlands

Re: K250972
Trade/Device Name: Primum Hydrophilic Guiding Catheter
Regulation Number: 21 CFR 870.1250
Regulation Name: Percutaneous Catheter
Regulatory Class: Class II
Product Code: DQY
Dated: March 31, 2025
Received: March 31, 2025

Dear Erendira Rodriguez:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K250972 - Erendira Rodriguez
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K250972 - Erendira Rodriguez
Page 3

Sincerely,

Jenny R. Katsnelson -S

Digitally signed by Jenny R. Katsnelson -S
Date: 2025.06.29 23:22:10
-04'00'

for Lydia Glaw
Assistant Director
DHT2C: Division of Coronary and Peripheral Intervention Devices
OHT2: Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K250972  |   |
|  Device Name Primum Hydrophilic Guiding Catheter  |   |
|  Indications for Use (Describe) The Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Guiding Catheter is intended to be used in the coronary or peripheral vascular system.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

K250972
Page 1 of 9

# 510(k) Summary

Per 21 CFR 807.92

## A. Submitter information (807.92 (a) (1))

**Prepared by:**

Eréndira Rodríguez
Quality Assurance and Regulatory Affairs Director
PendraCare International B.V.
Tel +31 85 0073 000

**Owner/Operator:**

**Manufacturer (510(k) Applicant)**

PendraCare International B.V.
Van der Waalspark 22
9351VC, Leek
The Netherlands
Registration: 3005635992
FEI: 3005635992

**Sterilization Facility**

Sterigenics Belgium Petit Rechain S.A.
Zoning Industriel De Petit-Rechain
Avenue Andre Ernst 21
Verviers Liege, BE B-4800
Registration Number: 3005941719
FEI Number: 3002807111

**Contact Person:**

Eréndira Rodríguez
Quality Assurance and Regulatory Affairs Director
PendraCare International B.V.
Van der Waalspark 22
9351VC, Leek
The Netherlands
Tel +31 85 0073 000
Email: erendira.rodriguez@welling.com

**Date Prepared:**

March, 2025

## B. Device Name

Trade or proprietary name: Primum Guiding Catheter
Common or usual name: Percutaneous Catheter
Classification name: Class II, 21 CFR 870.1250
Review Panel: Cardiovascular
Product Code: DQY

{5}

K250972
Page 2 of 9

## C. Predicate Device information

The legally marketed device which the 510(k) used for claiming equivalence are

- Convey 5F Guiding Catheters (K120585, August 8, 2012)
- Convey 6F Guiding Catheters (K120585, August 8, 2012)

PendraCare International B.V., Leek, The Netherlands, is the legal manufacturer of both the predicate device, Convey Guiding Catheter, and the submission subject device, the Primum Guiding Catheter. The Convey Guiding Catheter is a spin-off of the Primum Guiding Catheter, thus the products are identical with the only difference being the trade name.

## D. Device Description

The Hydrophilic Guiding Catheter consists of a reinforced body with a hub and strain relief at the proximal end and an intermediate and soft tip at the distal end. The distal part of the catheter features a specific tip shape. A part of the catheter body features a hydrophilic coating. The (distal part of the) catheter body is provided either with or without "in-line" side holes.

The guiding catheter is a flexible plastic tube featuring a luer hub, strain relief, a body, an intermediate tip, and a soft tip. The body and the intermediate tip consist of an inner liner (basecoat) and an outer jacket (topcoat) reinforced with a tightly wound stainless steel braid wire in between the layers. The central lumen of the catheter is used for the percutaneous, transluminal passage and placement of guidewires, diagnostic and therapeutic devices within the vascular system. After the catheter is inserted through the skin using a dilator, a sheath (introducer) and guide wire is brought into position. Subsequently, a guidewire is advanced through its lumen and tracked over by a diagnostic device and/or a therapeutic devices to the intended location. The distal section of the catheters has a variety of preformed shapes (e.g. Judkins Left (abbreviated as JL), Judkins Right (JR), Amplatz (AL), Multi-purpose (MP), hockey stick) to facilitate placement of the catheter tip in the desired target vessel. Some catheter models feature two (2) small "in-line" side holes in the intermediate tip section to maintain perfusion of the target vessel. This device is a single-use device (i.e., single patient, single procedure, single purpose use). After finalizing the procedure, the catheter is withdrawn, removed and discarded.

{6}

K250972
Page 3 of 9

E. Intended use of Device

The Guiding Catheter is intended to provide a pathway through which therapeutic and diagnostic devices are introduced into the vascular system.

F. Indications for Use

The Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Guiding Catheter is intended to be used in the coronary or peripheral vascular system.

G. Technological Characteristics / Substantial Equivalence

The Predicate device, Convey Guiding Catheter, is a spin-off of the Primum Guiding Catheter, with the only difference being the trade names. The products are identical in design, technical and biological specifications.

The Primum Guiding Catheters (5F, 6F, 7F and 8F) incorporate substantially equivalent device materials, packaging materials and design, intended use, fundamental technology (operating principle &amp; mechanism of action), labeling and

manufacturing processes, and sterilization process as those featured in the legally marketed predicated devices, the Convey 5F &amp; 6F Guiding Catheters (K120585, August 8, 2012) and Convey 7F &amp; 8F Guiding Catheters (K132197, July 11, 2013).

|  Device Characteristic | New device | Primary Predicate | Reference  |
| --- | --- | --- | --- |
|  Manufacturer | PendraCare International B.V. | PendraCare International B.V. | PendraCare International B.V.  |
|  Trade name device | Primum 5F, 6F, 7F and 8F Guiding Catheter K250972 | Convey 5F, 6F Guiding Catheter K120585 | Convey 7F, 8F Guiding Catheter K132197  |
|  Intended use/Indications for Use | The Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Guiding Catheter is intended to be used in the coronary or peripheral vascular system. | The Convey Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Convey Guiding Catheter is intended to be used in the coronary or peripheral vascular system. | The Convey Guiding Catheter is designed to provide a pathway through which therapeutic and diagnostic devices are introduced. The Convey Guiding Catheter is intended to be used in the coronary or peripheral vascular system.  |

{7}

K250972
Page 4 of 9

|  Operation Principle | Manually operated. Manual process. | Same | Same  |
| --- | --- | --- | --- |
|  Design/Construction | Hub, strain Relief, body, braiding wire between outer & inner body layer, Intermediate tip, Soft Tip, Side holes (optional), Hydrophilic coated body, uncoated body. | Same | Same  |
|  Materials | Hub
Hub:
96.15% Grilamid TR55LX
3.85% Colorant white

Hub Printing:
Hot stamp foil Grey (5F)
Hot stamp foil green (6F)
Hot stamp foil Orange (7F)
Hot stamp foil Blue (8F)

Strain relief:
98% Grilamid L25 W40
2% Colorant white

Body
Base coat:
40% Vestamid L2140
40% Vestamid D16
20% Fusabond E204

Braiding:
Flat rolled Stainless Steel
304V

Top coat:
38% Vestamid L2140
35% Vestamid E62-S3
23% [Vestamid D16 (80%)+BiOCl (BironLF2000)(20%)]
4% Masterbatch colorant Blue (HT-MAB PA 51130)

Coating (external):
Primer – Hydrophilic Coating

Intermediate tip-part-1*
Base coat:
40% Vestamid L2140 | Same | Same  |

{8}

K250972
Page 5 of 9

|  40% Vestamid D16
20% Fusabond E204

Braiding:
Flat rolled Stainless steel
304V

Intermediate tip material:
49% [40% Pebax 3533SA01 + 60% BiOCl]
49% Pebax 3533 SA01
2% Masterbatch colorant
Blue (HT-MAB PA 51130)

Intermediate tip
Part-2*
Filler Ring:
98% [40% Pebax 3533SA01 + 60% BiOCl]
2% Masterbatch colorant
Blue (HT-MAB PA 51130)

Braiding:
Flat rolled Stainless steel
304V

Intermediate tip material:
49% [40% Pebax 3533SA01 + 60% BiOCl]
49% Pebax 3533 SA01
2% Masterbatch colorant
Blue (HT-MAB PA 51130)

Soft tip:
Polyether block-amide /
Pebax 3533 SA01 (40%) BiOCl
(60%)

Packaging Materials
Mounting card:
240 grs GZ (SBS) bleached
carton
Pouch:
PET/PE Foil / Film,
transparent
• Top: 12 μm
Laminated Polyester
(PET)
• Bottom: 50 μm
Peelable  |
| --- |

5

{9}

K250972
Page 6 of 9

|   | Polyethylene PE (96%) + 4% EVA, Tyvek® 1073B
• uncoated Spunbonded non-woven polyethylene (HDPE)
• natural white
No printing

Hang-tab:
• Plastic sheet material PETG 635 μm
• Adhesive: Hang-tite 203-A

Box:
Box (1) & Box (5), Box (5) 125cm:
• 400/380/355 grams GC1 (or GC2) carton & white (or crème) back-layer
• No Printing

Labelling
IFU:
Woodfree Offset Paper, white, 70 grams – profijt

Inner / Pouch Label (SB)
Outer / Box Label (SB):
Mactac’s Velvet wood free coated paper,
1 X MP 196 acrylic adhesive emulsion
Century 6

Inner/pouch & Outer/Box label Ink:
Armor’s Thermal Transfer Ribbon APR6, Wax/Resin, Black Carbon D1-547.

End label:
80 gr flat wood free paper w/ permanent adhesive, white, no printing
85 gr flat wood free paper w/ permanent adhesive, white, no printing |  |   |
| --- | --- | --- | --- |

6

{10}

K250972
Page 7 of 9

|  Package | Unit Pouch
Shelf Box
Shipping Carton | Same | Same  |
| --- | --- | --- | --- |
|  Specifications * | See Dimension table below. | Same | Same  |
|  Shelf life | 3 years (36 months) | Same | Same  |
|  Disposable single use | Yes | Same | Same  |

*Specifications Dimension Table for Primum Guiding Catheter &amp; Convey Guiding Catheter

|  Primum Guiding Catheter & Convey Guiding Catheter  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Dimension Table (Nominal) [mm]  |   |   |   |   |   |
|  Description |   | 5F | 6F | 7F | 8F  |
|  A | Soft Tip Length | 2 | 2 | 2 | 2  |
|  B | Intermediate Tip Length | 15 | 15 | 15 | 15  |
|  C | Distal Position First Side Hole | n/a | 27 | 27 | 27  |
|  D | Distance between Side Holes | n/a | 4 | 4 | 4  |
|  E | Hydrophilic Coating End Position, proximal from distal Tip | 17-87 | 17-87 | 17-87 | 17-87  |
|  F | Hydrophilic Coating Start Position, proximal from distal Tip | 750 | 750 | 750 | 750  |
|  G | Usable Catheter Length (90/100/125cm) | 900/1000 | 900/1000/1250 | 900/1000 | 900/1000  |
|  H | Strain Relief Length | 25 | 25 | 25 | 25  |
|  I | Total Length (90/100/125cm) | 972/1072 | 972/1072/1322 | 972/1072 | 972/1072  |
|  ØJ | Side Hole Diameter | n/a | 0.77 | 0.77 | 0.94  |
|  ØOD | Outer Diameter Body / Soft Tip | ≤1.82 / 1.89 | ≤2.11 / 2.20 | ≤2.46 / 2.54 | ≤2.78 / 2.87  |
|  ØID | Inner Diameter Body, Hub / Soft Tip | ≥1.45 / 1.43 | ≥1.80 / 1.78 | ≥2.05 / 2.03 | ≥2.29 / 2.27  |

7

{11}

K250972
Page 8 of 9

# H. Summary biocompatibility and bench testing

The following biocompatibility tests were considered for the Primum Guiding Catheter considering the categorization of the Material Characterization for medical application per ISO 10993-1 and USP.

Categorization by Nature of body contact:
- External communicating device:
- Circulating blood: devices that contact circulating blood.

Categorization by Duration of Contact:
- Limited Exposure (A): Devices whose single or multiple use or contact is up to 24h.

ISO 10993-4: Haemocompatibility
- Hemolysis
- In vitro Haemocompatibility
- Coagulation Test Prothrombin Time Assay (PT)
- Coagulation Unactivated Partial Thromboplastic in Time Assay (UPTT)

ISO 10993-5: Cytotoxicity

ISO 10993-7: Ethylene Oxide Sterilization Residuals

ISO 10993-10: Sensitization

ISO 10993-10: Irritation/Intracutaneous Reactivity

ISO 10993-11: Acute Systemic Toxicity

ISO 10993-11: Material Meditated Pyrogenicity

USP &lt;661&gt;: Packing Plastic containers leachables.

USP &lt;85&gt; Endotoxin-Mediated Pyrogenicity

The following in-vitro performances tests (following simulated use, if applicable) were considered for the Primum Guiding Catheters
- Shape Conformance (Shape retention)
- Inner Diameter
- Outer Diameter
- Catheter Usable Length
- Coating Length
- C-Kink (Bending Kink Diameter)
- Euler Kink (axial Kink Displacement)
- Radial Stiffness (Collapse)
- Coating Integrity (Visual Inspection)
- Outer Friction &amp; Wear (Coating Integrity – functional tests)
- Three Point Bending Test (ending Stiffness Body)
- Pull Force (after simulated use)
- Radiopacity
- Torque Strength.
- Particulate Testing

{12}

K250972
Page 9 of 9

I. Clinical and Nonclinical Performance Data and Conclusions

The Guiding Catheter is designed to provide a pathway through which therapeutic devices are introduced. The Guiding Catheter is intended to be used in the coronary or peripheral vascular system.

The only difference between the subject Device, Primum Guiding Catheter, and Predicate Device, Convey Guiding Catheter, is the trade name. Both devices are the same generation of guiding catheter developed by the manufacturer.

Therefore, the data presented in Convey 5F Guiding Catheters (K120585, August 8, 2012), Convey 6F Guiding Catheters (K120585, August 8, 2012), Convey 7F Guiding Catheters (K132197, July 11, 2013), and Convey 8F Guiding Catheters (K132197, July 11, 2013) on non-clinical design verification (bench) tests, design validation testing, as well as clinical post market surveillance (comparison) studies, including investigator-driven registries, and biocompatibility evaluation tests per ISO 10993-series of standards are all applicable to the Primum Guiding Catheter.

The data provided data for the FDA market clearances of the Predicate device, Convey Guiding Catheter, showed reasonable assurance that the proposed devices were designed and tested to assure conformance to the requirements for its intended use, which are the same data applicable to the Primum Guiding Catheter. No new safety or performance issues were raised during the testing.

J. Conclusion

The Primum Guiding Catheters are substantially equivalent to the Convey Guiding Catheters, which were cleared by FDA under premarket notifications K120585 (August 8, 2012) for 5F and 6F.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K250972](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K250972)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
