CPS Locator 3D Plus Delivery Catheter

K250828 · CenterPoint Systems, LLC · DQY · Jun 27, 2025 · Cardiovascular

Device Facts

Record IDK250828
Device NameCPS Locator 3D Plus Delivery Catheter
ApplicantCenterPoint Systems, LLC
Product CodeDQY · Cardiovascular
Decision DateJun 27, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads.

Device Story

Single-use percutaneous delivery catheter; used for introducing catheters and pacing/defibrillation leads into vasculature. Device includes catheter, dilator, and hemostatic valve. Distal radiopaque tip allows fluoroscopic visualization. Physician advances catheter/dilator over 0.035" guidewire to target location; dilator removed; lead/catheter inserted through lumen. Device features slittable design for removal after placement. Used in clinical settings by physicians. Benefits include facilitated lead/catheter placement and removal via slitting mechanism.

Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-1), sterilization (ISO 11137), packaging (ANSI/AAMI/ISO 11607-1), simulated use, valve leak, tensile strength, dimensional verification, and flush testing.

Technological Characteristics

Percutaneous catheter; PEBAX shaft with Barium Sulfate and colorants; braid-reinforced; hydrophilic liner; radiopaque distal tip; hemostatic valve. Dimensions: 10.2F OD, 8.4F ID, 42-45cm length. 0.035" guidewire compatible. E-Beam sterilization (SAL 10^-6). Single-use.

Indications for Use

Indicated for the introduction of various types of catheters and pacing or defibrillator leads into the vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 27, 2025 CenterPoint Systems LLC Conner Johnson Senior Regulatory Specialist 3338 Parkway Blvd West Valley City, Utah 84119 Re: K250828 Trade/Device Name: CPS Locator 3D Plus Delivery Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dated: June 5, 2025 Received: June 5, 2025 Dear Conner Johnson: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250828 - Conner Johnson Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K250828 - Conner Johnson Page 3 Sincerely, JESSICA L. BATISTA-S Digitally signed by JESSICA L. BATISTA-S Date: 2025.06.27 12:09:33 -04'00' for - Sara Royce - Assistant Director - Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices - Office of Cardiovascular Devices - Office of Product Evaluation and Quality - Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250828 | | | Device Name CPS Locator 3D Plus Delivery Catheter | | | Indications for Use (Describe) The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} K250828 Page 1 of 5 # 510(k) Summary ## 1.1 Submitter Name: CenterPoint Systems LLC Address: 3338 Parkway Blvd West Valley City UT Phone: 801-602-1923 Contact Person: Conner Johnson, Senior Regulatory Specialist Date Prepared: 18 March 2025 ## 1.2 Device Name of Device: CPS Locator 3D Plus Delivery Catheter Classification Name: Percutaneous Catheter Regulatory Class: Class II per 21 CFR 870.1250 Product Code: DQY ## 1.3 Predicate Device **Predicate Name and 510(k) Number**: CPS Locator 3D Delivery Catheter, K230363 This predicate has not been subject to a design-related recall. ## 1.4 Device Description The modified Delivery Catheter (CPS Locator 3D Delivery Catheter) is a single-use percutaneous catheter indicated for the introduction of various types of catheters and pacing or defibrillation leads. The modified Delivery Catheter is packaged with a dilator for introduction into the vasculature. Proximally, the modified Delivery Catheter is equipped with a hemostatic valve, and the distal soft, rounded, radiopaque tip facilitates imaging under fluoroscopy. The modified Delivery Catheter is designed to be slittable, thereby allowing its removal after device placement. A variety of curves and lengths are available to accommodate various anatomies and different locations. {5} K250828 Page 2 of 5 # 1.5 Indications for Use The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads. # 1.6 Comparison of Technological Characteristics with the Predicate Devices The Proposed Device and Predicate Device are similar in indications for use, intended use, technological characteristics, and principles of operation. The differences between the Proposed Device and the Predicate Device are minor and raise no different questions of safety and effectiveness. In accordance with 21CFR807.92(a)(6) a summary of how the technological characteristics of the Proposed Device compares to the Predicate Device is provided below. | Feature | CPS Locator 3D Plus Delivery Catheter (subject device) | Primary Predicate: CPS Locator 3D Delivery Catheter (K230363) | Same / Different between Proposed & Predicates | | --- | --- | --- | --- | | Intended Use | Percutaneous catheter for the delivery of leads | Percutaneous catheter for the delivery of leads | Same | | Indications for Use | The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads. | The Delivery Catheter is indicated for the introduction of various types of catheters and pacing or defibrillator leads. | Same | | Product Code | DQY | DQY | Same | | Regulation number | 21 CFR 870.1250 | 21 CFR 870.1250 | Same | | Prescription Device | Yes | Yes | Same | | Catheter Type | Percutaneous Catheter | Percutaneous Catheter | Same | | Guidewire Compatibility | 0.035” | 0.035” | Same | | Outer Diameter | 10.2F | 9F | Substantially Equivalent | | Inner Diameter | 8.4F | 7F | Substantially Equivalent | | Working Length | 42cm, 45cm | 42cm, 45cm | Same | | Components Provided | Catheter, Dilator, Luer cap | Catheter, Dilator, Luer cap | Same | | Hydrophilic Liner | Yes | Yes | Same | {6} | Feature | CPS Locator 3D Plus Delivery Catheter (subject device) | Primary Predicate: CPS Locator 3D Delivery Catheter (K230363) | Same / Different between Proposed & Predicates | | --- | --- | --- | --- | | Radiopaque Distal Tip | Yes | Yes | Same | | Valve | Yes | Yes | Same | | Shaft Materials | PEBAX with Barium Sulfate and colorants | PEBAX with Barium Sulfate and colorants | Same | | Braid Reinforcement | Yes | Yes | Same | | Dilator | Yes | Yes | Same | | Multiple Distal End Shapes Available | Yes | Yes | Same | | Sterility | E-Beam Sterilization, SAL 10^{-6} | E-Beam Sterilization, SAL 10^{-6} | Same | | Number of Uses | Single patient use | Single patient use | Same | | Principles of Operation | After venous access is gained, the catheter and dilator are advanced over a guidewire to the desired location. The dilator is removed and a catheter or lead is placed through the Delivery Catheter. The Delivery Catheter may be removed by slitting. | After venous access is gained, the catheter and dilator are advanced over a guidewire to the desired location. The dilator is removed and a catheter or lead is placed through the Delivery Catheter. The Delivery Catheter may be removed by slitting. | Same | The modified Delivery Catheter is used for the same intended use in the same anatomical location using the same principles of operation as the predicate device. K250828 Page 3 of 5 {7} # 1.7 Performance Data All necessary performance testing has been conducted on the CPS Locator 3D Plus Delivery Catheter to assure substantial equivalence to the predicate devices and to demonstrate the device performs as intended. All testing was performed on test units representative of finished devices. The device passed the following tests, which were conducted in accordance with noted standards: - Biocompatibility testing per FDA Final Guidance Document, "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (2023) - Sterilization validation per ISO 11137 - Packaging validation per ANSI/AAMI/ISO 11607-1 - Visual Inspection - Simulated use testing, including use/compatibility with ancillary devices - Valve liquid leak test - Tensile tests - Sheath and Dilator Dimensional verification, including OD/ID, working length - Flush test K250828 Page 4 of 5 {8} # 1.8 Conclusions Based on the similarity of the subject and predicate devices in terms of the intended use, principle of operation and overall technological characteristics, and the results of the testing summarized in section 1.7, the CPS Locator 3D Plus Delivery Catheter is substantially equivalent to the predicate device (K230363). K250828 Page 5 of 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...