← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K231608

# Teleport XT Microcatheter (K231608)

_OrbusNeich Medical (Shenzhen) Co., Ltd. · DQY · Jan 7, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K231608

## Device Facts

- **Applicant:** OrbusNeich Medical (Shenzhen) Co., Ltd.
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Jan 7, 2024
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Teleport XT microcatheters are indicated for: - supporting and facilitating the placement of guidewires in the coronary and peripheral vasculature. - exchanging guidewires in the coronary and peripheral vasculature. - the delivery of contrast media into the coronary, peripheral, and abdominal vasculature.

## Device Story

Single-lumen, over-the-wire (OTW) microcatheter; used in coronary, peripheral, and abdominal vasculature. Facilitates guidewire placement/exchange and contrast media delivery. Operated by physicians in clinical settings. Features 2.1F distal OD; 90cm, 135cm, or 150cm working lengths; hydrophilic outer coating; fluoropolymer-lined lumen. Compatible with 0.014-inch guidewires. Provides mechanical support for navigation through vasculature; contrast delivery via hub. Benefits patient by enabling access to complex vascular lesions.

## Clinical Evidence

Bench testing only. Performance data included sterilization, shelf-life, particulate evaluation, visual/dimensional inspection, media flow rate, simulated use, vacuum leakage, shaft burst pressure, guidewire compatibility, coating integrity, flexibility, kinking, corrosion resistance, torque strength, tensile strength, radiopacity, and pouch integrity. All test results met acceptance criteria.

## Technological Characteristics

Single-lumen OTW catheter; 2.1F distal OD; hydrophilic polymer outer coating; fluoropolymer-lined lumen. Compatible with 0.014" guidewires and 4F guiding catheters. Max pressure 300 psi. Sterilization via Ethylene Oxide (EO).

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Teleport Microcatheter ([K182360](/device/K182360.md))

## Reference Devices

- ASAHI Corsair Pro XS ([K182420](/device/K182420.md))
- Turnpike LP Catheter ([K191560](/device/K191560.md))
- Mamba Flex Microcatheter ([K171452](/device/K171452.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 7, 2024

OrbusNeich Medical (Shenzhen) Co., Ltd. Jerry Cheung Senior Director of Regulatory Affairs No.1 Jinkui Road, Futian Free Trade Zone Shenzhen, Guangdong 518038 China

## Re: K231608

Trade/Device Name: Teleport XT Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: December 8, 2023 Received: December 8, 2023

Dear Jerry Cheung:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

{1}------------------------------------------------

Page 2

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Lydia S. Glaw -S

Digitally signed by Lydia S. Glaw -S Date: 2024.01.07 10:56:35 -05'00'

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K231608

Device Name Teleport XT Microcatheter

Indications for Use (Describe)

The Teleport XT microcatheters are indicated for:

- · supporting and facilitating the placement of guidewires in the coronary and peripheral vasculature.
- · exchanging guidewires in the coronary and peripheral vasculature.
- · the delivery of contrast media into the coronary, peripheral, and abdominal vasculature.

| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;">                     ☒ Prescription Use (Part 21 CFR 801 Subpart D)                 </span> <span>                     ☐ Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services
   
Food and Drug Administration
   
Office of Chief Information Officer
   
Paperwork Reduction Act (PRA) Staff
   
PRAStaff@fda.hhs.gov

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

FORM FDA 3881 (6/20)

{3}------------------------------------------------

## 510(k) Summary

This 510(k) Summary is submitted in accordance with 21 CFR 807.92.

| Submitter:                        | OrbusNeich Medical (Shenzhen) Co., Ltd.<br>No.1 Jinkui Road<br>Futian Free Trade Zone<br>Shenzhen 518038, China                                                                                                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                   | Phone: +86-755-83580181-8330<br>Fax: +86-755-83580169                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Contact Person:                   | Name: Jerry Cheung<br>Job Title: Senior Director of Regulatory Affairs<br>Email: jcheung@orbusneich.com                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Date Prepared:                    | May 31, 2023                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device:                           | Name of Device: Teleport XT Microcatheter<br>Common Name: Microcatheter<br>Classification Name: Percutaneous catheter (21<br>CFR<br>870.1250)<br>Regulatory Class: II<br>Product Code: DQY                                                                                                                                                                                                                                                                                                                              |
| Predicate Device:                 | Teleport Microcatheter (K182360, DQY, cleared Nov 09,<br>2018)<br>This predicate has not been subject to a design-related recall.                                                                                                                                                                                                                                                                                                                                                                                       |
| Reference Devices:                | ASAHI Corsair Pro XS (K182420, DQY, cleared Dec 20,<br>2018)<br>Turnpike LP Catheter (K191560, DQY, cleared Aug 09,<br>2019)<br>Mamba Flex Microcatheter (K171452, DQY, cleared Aug<br>21, 2017)                                                                                                                                                                                                                                                                                                                        |
| Device Description:               | Teleport XT Microcatheter is a single lumen OTW catheter<br>offered in one size (2.1F, distal OD) with working lengths of<br>90cm, 135cm or 150cm, designed for use in the coronary and<br>peripheral vasculature. The shaft profile gradually changes<br>from 2.8F (0.0370") to 2.1F (0.0280"). The catheter consists<br>of five primary sections: hub, body shaft, proximal section,<br>distal section, and a radiopaque tip. The distal most 60cm of<br>the outer surface is coated with hydrophilic polymer to      |
|                                   | increase lubricity and the lumen of the catheter is lined with fluoropolymer to facilitate movement of the guidewire. The catheter is compatible with a standard 0.014-inch (0.36mm) guidewire.                                                                                                                                                                                                                                                                                                                         |
| Indications For Use:              | The Teleport XT microcatheters are indicated for:<br>supporting and facilitating the placement of guidewires in the coronary and peripheral vasculature. exchanging guidewires in the coronary and peripheral vasculature. the delivery of contrast media into the coronary, peripheral, and abdominal vasculature.                                                                                                                                                                                                     |
| Technological<br>Characteristics: | The subject device has the following similarities to the predicate devices:<br>Same indications for use Same catheter design Similar materials of construction Same hydrophilic coating 0.014" guidewire compatibility 4F guiding catheter compatibility Maximum allowable pressure of 300 psi Same method of EO sterilization              The following technological differences exist between the subject and predicate device:  specific materials selected exact dimensions of components catheter working length |
| Performance Data:                 | The following performance data were provided in support of the substantial equivalence determination.<br>Sterilization Shelf-Life Performance Testing  Particulate Evaluation Visual Inspection Dimension Inspection Media Flow Rate Simulated Use Vacuum Leakage                                                                                                                                                                                                                                                       |

{4}------------------------------------------------

- > Shaft Burst Pressure

{5}------------------------------------------------

- Guidewire Compatibility >
- Coating Integrity >
- Flexibility and Kinking >
- > Corrosion Resistance
- > Torque Strength
- > Tensile
- > Radiopacity
- Pouch Integrity >
- > Pouch Burst
- > Seal Strength

The Teleport XT microcatheter test results met all acceptance criteria and were similar to the predicate and reference devices.

Conclusion: This information supports a determination of substantial equivalence between the Teleport XT microcatheter and the predicate device described above.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K231608](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K231608)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
