← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K221846

# Artix Ballon Guiding Sheath (K221846)

_Inari Medical · DQY · Jul 11, 2022 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K221846

## Device Facts

- **Applicant:** Inari Medical
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Jul 11, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Artix Balloon Guiding Sheath is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Artix Balloon Guiding Sheath is also indicated for use as a conduit for retrieval devices. The Artix Balloon Guiding Sheath is intended for use in the peripheral vasculature.

## Device Story

Artix Balloon Guiding Sheath is a single-use, over-the-wire peripheral vascular access device. It consists of a sheath with a distal compliant balloon, integrated hemostasis valve, and inflation/flush ports. Operated by physicians in clinical settings, the device is inserted over a guidewire into a peripheral vessel. The balloon is inflated via a 1 mL syringe to provide temporary flow occlusion during angiographic procedures. The sheath serves as a conduit for inserting other intravascular catheters or retrieval devices. Radiopaque marker bands at the distal tip aid in fluoroscopic visualization. The device facilitates procedural access and vessel management, potentially improving clinical outcomes during peripheral interventions.

## Clinical Evidence

No clinical data or animal testing were required for this 510(k) submission. Substantial equivalence was established through extensive non-clinical bench testing, including biocompatibility (ISO 10993-1), sterilization validation (ISO 11135), and mechanical performance testing (e.g., balloon fatigue, burst volume, kink radius, tensile strength, and simulated use).

## Technological Characteristics

Single-use, over-the-wire sheath; 8 Fr ID, 11 Fr OD; 65 cm or 105 cm lengths. Materials: Pebax 4033 shaft, Chronoprene 40A balloon, etched PTFE liner, 304V stainless steel braid/coil reinforcement, platinum-iridium marker bands. Hydrophilic coating on distal shaft. Sterilization: EtO (ISO 11135). Connectivity: None (mechanical device).

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Concentric Medical, 8F FlowGate2 Balloon Guide Catheter ([K153729](/device/K153729.md))

## Reference Devices

- Medtronic Vascular, Inc., Sentrant Introducer Sheath with Hydrophilic Coating ([K171866](/device/K171866.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The text logo has the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 11, 2022

Inari Medical % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k Saint Paul, Minnesota 55114

Re: K221846

Trade/Device Name: Artix Ballon Guiding Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DOY Dated: June 23, 2022 Received: June 24, 2022

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For Misti Malone Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K221846

Device Name Artix Balloon Guiding Sheath

#### Indications for Use (Describe)

The Artix Balloon Guiding Sheath is indicated for use in facilitating the insertion and guidance of an intravascular catheter into a selected blood vessel. The balloon provides temporary vascular occlusion during these and other angiographic procedures. The Artix Balloon Guiding Sheath is also indicated for use as a conduit for retrieval devices.

The Artix Balloon Guiding Sheath is intended for use in the peripheral vasculature.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | ☐                                       |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## PAGE 1 OF 6

# 510(K) SUMMARY

| Date prepared         | June 9, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name                  | Inari Medical, Inc.<br>6001 Oak Canyon, Suite 100<br>Irvine, CA 92618<br>877.923.4747                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Contact person        | Ellen Nguyen<br>Regulatory Affairs Specialist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Trade name            | Artix Balloon Guiding Sheath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Common name           | Balloon guiding sheath                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Regulation name       | Percutaneous catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Classification number | 21 CFR 870.1250                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Product code          | DQY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulatory class      | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Predicate device      | Concentric Medical, 8F FlowGate2 Balloon Guide Catheter (K153729)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Reference device      | Medtronic Vascular, Inc., Sentrant Introducer Sheath with Hydrophilic Coating (K171866)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Description           | The Artix Balloon Guiding Sheath is a single-use, over-the-wire system designed to<br>facilitate the insertion and guidance of an intravascular catheter into a selected<br>peripheral blood vessel and act as a conduit for retrieval devices. A compliant<br>balloon mounted at the sheath's distal tip provides temporary vascular occlusion<br>during angiographic procedures. The Artix Balloon Guiding Sheath is packaged with<br>the following components:<br>Artix Balloon Guiding Sheath (8 Fr, 65 cm or 105 cm)Two 8 Fr Introducer Dilators (0.014" and 0.035" guidewire compatibility)Balloon Inflation Syringe, 1 mLLarge Bore Syringe, 30 mL3-way Stopcock |
| Indications for Use   | The Artix Balloon Guiding Sheath is indicated for use in facilitating the insertion<br>and guidance of an intravascular catheter into a selected blood vessel. The balloon<br>provides temporary vascular occlusion during these and other angiographic<br>procedures. The Artix Balloon Guiding Sheath is also indicated for use as a conduit<br>for retrieval devices.<br>The Artix Balloon Guiding Sheath is intended for use in the peripheral vasculature.                                                                                                                                                                                                         |

{4}------------------------------------------------

Summary of substantial equivalence

The Artix Balloon Guiding Sheath and predicate device have the same indications for use statement: both are indicated for vascular access, facilitating the insertion and guidance of intravascular devices into selected blood vessels in the peripheral vasculature, providing temporary vessel occlusion during angiographic procedures, and to act as a conduit for retrieval devices. Unlike the predicate device, the Artix Balloon Guiding Sheath is not indicated for use in the neurovasculature.

A tabular comparison of specific technological characteristics between the predicate and subject device is provided below:

| Feature                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 8F FlowGate2 Balloon<br>Guide Catheter - Predicate<br>(K153729)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                           | Inari Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Concentric Medical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Product code                           | DQY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | DQY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Intended<br>use/Indications for<br>use | The Artix Balloon Guiding<br>Sheath is indicated for use in<br>facilitating the insertion and<br>guidance of an intravascular<br>catheter into a selected blood<br>vessel. The balloon provides<br>temporary vascular occlusion<br>during these and other<br>angiographic procedures. The<br>Artix Balloon Guiding Sheath is<br>also indicated for use as a<br>conduit for retrieval devices.<br>The Artix Balloon Guiding<br>Sheath is intended for use in the<br>peripheral vasculature.                                                                                                                                                                                                                                     | FlowGate2 Balloon Guide<br>Catheters are indicated for<br>use in facilitating the<br>insertion and guidance of an<br>intravascular device into a<br>selected blood vessel in the<br>peripheral and neurovascular<br>systems. The balloon<br>provides temporary vascular<br>occlusion during these and<br>other angiographic<br>procedures. The Balloon<br>Guide Catheter is also<br>indicated for use as a conduit<br>for retrieval devices.                                                                                                                                                                                                            |
| Device description                     | The Artix Balloon Guiding<br>Sheath is comprised of a single<br>through lumen shaft with four<br>(4) embedded lumens for<br>balloon inflation and deflation, a<br>proximal integrated hemostasis<br>valve to minimize blood loss, a<br>stopcock with flush port, and a<br>balloon port. To assist with<br>insertion of the sheath into the<br>vasculature, a hydrophilic<br>coating covers a portion of the<br>distal sheath shaft. The distal tip<br>of the sheath shaft contains two<br>(2) radiopaque marker bands to<br>aid with angiographic<br>visualization: The distal marker<br>indicates the location of the<br>sheath's tip, while the proximal<br>marker band marks the location<br>of a compliant balloon mounted | The FlowGate2 Balloon<br>Guide Catheters are coaxial-lumen, braid-reinforced,<br>variable stiffness catheters<br>designed for use in<br>facilitating the insertion and<br>guidance of an intravascular<br>catheter into a selected blood<br>vessel in the peripheral and<br>neurovascular systems. A<br>radiopaque marker is<br>included on the distal end for<br>angiographic visualization. A<br>compliant balloon is<br>mounted on the distal end to<br>provide temporary vascular<br>occlusion during<br>angiographic procedures. A<br>bifurcated luer hub on the<br>proximal end allows<br>attachments for flushing,<br>inflation, and aspiration. |
| Feature                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 8F FlowGate2 Balloon Guide Catheter - Predicate (K153729)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                        | at the distal end. The balloon provides temporary vascular occlusion during angiographic procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Balloon Guide Catheter dimensions and maximum recommended balloon inflation volume are indicated on product label.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Principles of operation                | After the target vessel is accessed and dilated, the Artix Balloon Guiding Sheath is inserted into the vessel over a guidewire. Once the Artix Balloon Guiding Sheath is in position, the dilator is removed, and all compatible catheter devices can be inserted through the Artix Balloon Guiding Sheath for access into the peripheral vasculature. The sheath's balloon can also be inflated using a 1 mL syringe to provide flow occlusion of the vessel during the procedure.                                                                                                                                                                                                                                            | After the target vessel is accessed and dilated, the 8F FlowGate2 Balloon Guide Catheter is inserted into the vessel. The dilator can then be removed to allow compatible working devices to be inserted through the catheter. The balloon can also be inflated to provide flow occlusion of the vessel.                                                                                                                                                                                                                                                                                                                                                |
| Intended patient population            | Adults undergoing interventional procedures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Adults undergoing interventional procedures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Target vessel                          | Peripheral vasculature                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Peripheral and neurovasculature                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Placement duration                     | < 24 hours                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | < 24 hours                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Guidewire compatibility                | 0.014" or 0.035"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 0.035"                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Sterilization                          | EtO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | EtO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Single-use                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Dimensions                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Sheath dimensions                      | OD/ID: 11 Fr/8 Fr Effective Lengths: 105 cm and 65 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | OD/ID: 8 Fr/6.4 Fr Effective Lengths: 100 cm and 95 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Balloon dimensions                     | Ø 11 mm maximum OD Length: 10 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Ø 10 mm maximum OD Length: 10 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Materials                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Shaft material                         | Pebax 4033                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Proximal: Pebax 7233                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Distal: Pebax 6333                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Tip: Pebax 2533, Pebax 3533, BaSO4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Coating                                | Hydrophilic, biocompatible                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Balloon                                | Chronoprene 40A thermoplastic elastomer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | NuSil MED 4025 silicone elastomer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Feature                                | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 8F FlowGate2 Balloon<br>Guide Catheter - Predicate<br>(K153729)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Liner                                  | Etched PTFE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Etched PTFE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Shaft support                          | 304V Stainless Steel braid and<br>coil                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 304V Stainless Steel braid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Strain relief                          | Pellethane 2363-80A<br>Pantone 2104C PMS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Pebax 4033, White<br>Polyolefin, White                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Marker band                            | Platinum-iridium                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Platinum-iridium                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Accessories                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Accessories<br>provided                | Dilators (2), balloon inflation<br>syringe, large bore syringe, 3-<br>way stopcock, hemostasis valve<br>and side port lumen (integrated<br>into sheath shaft)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Dilator, rotating hemostasis<br>valve, Tuohy-Borst valve<br>with sideport, Peel-Away<br>Sheaths, Luer-activated<br>valve, Extension Tubing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Dilator length                         | 115 cm, 72 cm (working length)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 123 cm (working length)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Dilator material                       | Shaft: LDPE/HDPE<br>Hub: HDPE<br>Cap: ABS, 4% Cool Gray                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Shaft: Pebax, BaSO4<br>Hub: White polycarbonate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Dilator OD                             | 0.110 in.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Distal: 0.072 in.<br>Proximal: 0.078 in.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{5}------------------------------------------------

{6}------------------------------------------------

## Biocompatibility

The following biocompatibility tests were completed for the subject device:

- . Cytotoxicity
- . Sensitization
- . Intracutaneous Reactivity
- Acute Systemic Toxicity
- Material-Mediated Pyrogenicity ●
- Hemocompatibility (Hemolysis, Complement Activation, ● Thromboresistance, Platelet and Leukocyte Count, and Partial Thromboplastin Time)

The passing results demonstrate that the subject device and accessories meet biological safety requirements per ISO 10993-1.

.

### Sterilization

The subject device, including its accessories, is sterilized using EtO to achieve a sterility assurance level (SAL) of 106. The subject device has been adopted into a validated sterilization process in accordance with the principles of ISO 11135:2014/Amd 1:2018 (Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine control of a sterilization process for medical devices – Amendment 1: Revision of Annex E, Single batch release) and AAMI TIR 28:2016 (Product adoption and process equivalence for ethylene oxide sterilization) without deviations.

{7}------------------------------------------------

## Non-Clinical Testing

In accordance with the Design Failure Modes and Effects Analysis, verification and validation testing were identified to support the substantial equivalence of the Artix Balloon Guiding Sheath to the FlowGate2 Balloon Guide Catheter. These tests included:

- Pouch Seal Peel Test
- Pouch Seal Inspection and Pouch Dye Penetration ●
- Packaging Retention ●
- Visual and Dimensional Inspections Sheath ●
- Dimensional Inspections Sheath Inner Diameter ●
- Visual and Dimensional Inspections - Dilator
- Dimensional Inspections Dilator Outer Diameter ●
- Guidewire Compatibility, Dilator ●
- Visual Sheath and Dilator ●
- Dimensional Inspections Sheath Working Length ●
- Simulated Use Cycling
- Cap Locking ●
- Device Retrieval Testing ●
- Device Priming ●
- Air Leak Testing, Syringe Pullback ●
- Fluid Leak Testing, Sheath ●
- Fluid Leak Testing, Dilator
- Air Leak Testing, Dilator Removal
- 30 mL Large Bore Syringe Vacuum ●
- Balloon Fatigue Test ●
- Sheath and Dilator Kink Radius ●
- Dye Staining, Post-Use ●
- Vacuum Testing ●
- Dilator Hub to Cap Tensile ●
- 30 mL Large Bore Syringe, Tip Adapter to Barrel Tensile ●
- Unlocking Cap, Torque ●
- Insertion Force, Dilator Through Sheath ●
- Flushing Side Port to Hemostasis Valve Assembly and Inflation Hub . Tensile
- Hemostasis Valve Assembly to Sheath Shaft and Dilator Tensile ●
- Engaged Housing, Cap Axial Detachment
- Balloon Diameter, Inflated
- Barbed Luer to Inflation Hub Torque ●
- Balloon Burst Volume ●
- Retraction Force, Dilator from Sheath ●
- Lubricity, Post-Use ●
- Particulate Matter Determination .
- Torque. Sheath and Dilator ●
- Balloon Inflation Volume to Balloon Diameter ●
- Balloon Deflation Time .

{8}------------------------------------------------

## PAGE 6 OF 6

- Sheath Compatibility with Intravascular Catheter and Diagnostic Catheter in Model
- Sheath Kink to Failure
- Sheath Torque to Failure ●
- Sheath Tip Tensile
- Sheath Burst ●
- Manual Syringe Injection Flow Rate and Max Pressure ●

Test results demonstrated that all acceptance criteria were met; therefore, the device conforms to established product specifications.

Neither animal testing nor clinical testing were required for the determination of substantial equivalence.

## Conclusion

The nonclinical tests demonstrated that the device is substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K221846](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K221846)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
