K202610 · Merit Medical Systems, Inc. · DQY · Dec 14, 2020 · Cardiovascular
Device Facts
Record ID
K202610
Device Name
Ventrax Delivery System
Applicant
Merit Medical Systems, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Dec 14, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Ventrax Delivery System is intended to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart
Device Story
Ventrax Delivery System provides conduit for diagnostic/therapeutic catheter delivery to heart chambers/coronary vasculature; supports positioning/maintenance of catheters. System components: 8.5F guiding sheath introducer (integrated valve, sideport), mating pigtail dilator (tapered tip, pigtail for aortic valve crossing, integrated valve, sideport), optional mating straight dilator, and J-tipped InQwire Amplatz guidewire. Used in clinical settings by physicians; provides pathway for device advancement. Benefits patient by facilitating minimally invasive cardiac access. Healthcare provider uses system to navigate vasculature; output is the established pathway for subsequent therapeutic/diagnostic interventions.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity) and mechanical bench testing (tensile strength, leak testing, simulated use, curve retention, radiopacity, and seal strength) per ISO and ASTM standards.
Technological Characteristics
Percutaneous catheter delivery system. Components: 8.5F guiding sheath, pigtail dilator, straight dilator, J-tipped guidewire. Materials evaluated per ISO 10993-1. Sterilization: Ethylene Oxide (ISO 11135). Connectivity: None. Energy source: None (manual). Standards: ISO 11070, ISO 594-1/2, ISO 10993 series, ASTM F2096, ASTM F1929, ASTM F88, ASTM F1980.
Indications for Use
Indicated for providing a pathway for the introduction of diagnostic and therapeutic catheters into the heart chambers and coronary vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2020
Merit Medical Systems, Inc. Elizabeth Lazaro Sr. Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, UT 84095
Re: K202610
Trade/Device Name: Ventrax™ Delivery System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: November 10, 2020 Received: November 12, 2020
Dear Elizabeth Lazaro:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Rachel Neubrander Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K202610
Device Name Ventrax Delivery System
Indications for Use (Describe)
The Ventrax Delivery System is intended to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Merit Medical Systems, Inc.
K202610 Ventrax™ Delivery System Premarket Notification 510(k)
## 510(k) Summary K202610
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Contact Person:<br>Date Prepared: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(610) 651 -5093<br>Elizabeth Lazaro<br>September 04, 2020 |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Ventrax™ Delivery System<br>Delivery System<br>Percutaneous Catheter<br>II<br>DQY<br>870.1250<br>Cardiovascular |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | AMPLATZER® TorqVue® Delivery System<br>Percutaneous Catheter<br>K072313<br>AGA Medical Corporation |
| Reference<br>Device | | No reference devices were used in this submission. |
| Device<br>Description | The Ventrax™ Delivery System is designed to provide a conduit to deliver<br>diagnostic and therapeutic catheters to specific heart chambers and<br>locations. It provides support for positioning and maintaining the position<br>of catheters at specific locations in the heart. The guiding sheath<br>introducer may be used for percutaneous entry. The System consists of<br>four components: a sheath, a pigtail dilator, a straight dilator and a J-<br>tipped guidewire.<br>Ventrax™ Delivery System Components<br>A. 8.5F Guiding Sheath Introducer: provide a conduit to deliver<br>diagnostic and therapeutic catheters within the chambers and<br>coronary vasculature of the heart. The sheath has an integrated<br>valve to restrict blood loss, and a sideport for flushing and<br>withdrawing blood.<br>B. Mating Pigtail Dilator: designed to conform to the sheath<br>introducer inner diameter, has a tapered tip, has a pigtail at the | |
| | distal end to assist aortic valve crossing, has an integrated valve<br>to restrict blood loss, and has a sideport for flushing.<br>C. Mating Straight Dilator: designed to conform to the sheath<br>introducer inner diameter and has a tapered tip. Usage of this<br>straight dilator is optional. This straight dilator is intended to be<br>used only when access is unsuccessful after using the mating<br>pigtail dilator.<br>D. J- tipped InQwire®Amplatz Guidewire 0.035" X 180cm provides<br>pathway for sheath and dilator advancement. | |
| Indications<br>for Use | The Ventrax™ Delivery System is intended to provide a pathway through<br>which devices are introduced within the chambers and coronary<br>vasculature of the heart. | |
| | Summary of the technological characteristics of the Ventrax™ Delivery<br>System and the AMPLATZER® TorqVue® Delivery System are based on<br>the following: | |
| Comparison<br>to Predicate<br>Device | Similar Indications for Use Similar materials Similar design Same sterilization method Same fundamental technology / principle of operation | |
| | The Ventrax™ Delivery System is substantially equivalent to the<br>predicate device, Amplatzer ®TorqVue ®Delivery System cleared by<br>K072313, both delivery systems are designed to facilitate access within<br>the chambers and coronary vasculature of the heart. | |
| | Performance testing of the Ventrax™ Delivery System was conducted<br>based on the risk analysis and when applicable on the requirements of<br>the following standards: | |
| Performance<br>Data | ISO 11135:2014<br>Sterilization of Health-care Products - Ethylene Oxide - Requirements for<br>The Development, Validation and Routine Control of a Sterilization<br>Process for Medical Devices<br>ASTM D1469-2016<br>Standard Practice for Performance Testing of Shipping Containers and<br>Systems<br>ISO 11070:2014<br>Sterile Single-Use Intravascular Introducers, Dilators and Guidewires | |
| | ISO 594-1:1986 | |
| Performance<br>Data cont. | Conical Fittings with a 6% (Luer) Taper for Syringes, Needles and Certain<br>of Other Medical Equipment - Part 1: General Requirements | |
| | ISO 594-2:1988<br>Conical Fittings with a 6% (Luer) Taper for Syringes, Needles, and<br>Certain Other Medical Equipment - Part 2: Lock Fittings | |
| | ISO 10993-1:2018 | |
| | Biological Evaluation of Medical Devices -- Part 1: Evaluation and Testing<br>Within A Risk Management Process | |
| | ISO 10993-7:2008 (R2016) | |
| | Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide<br>Sterilization Residuals | |
| | ISO 2233:2000 (R2016) | |
| | Packaging - Complete, Filled Transport Packages and Unit Loads –<br>Conditioning for Testing | |
| | ASTM D999-08 (R2015) | |
| | Standard Test Methods for Vibration Testing of Shipping Containers | |
| | ASTM F 2096-11 (R2019) | |
| | Standard Test Method for Detecting Gross Leaks in Packaging by<br>Internal Pressurization (bubble Test) | |
| | AAMI TIR 28:2016 | |
| | Product Adoption and Process Equivalence for Ethylene Oxide<br>Sterilization | |
| | IEC 62366-1:2016 | |
| | Medical Devices - Part 1: Application of Usability Engineering to Medical<br>Devices | |
| | AAMI ISO 11607:2019 | |
| | Packaging for Terminally Sterilized Medical Devices- Part 1:<br>Requirements for Materials, Sterile Barrier Systems and Packaging<br>Systems | |
| | ASTM F1929:2015 | |
| | Standard Test Method for Detecting Seal Leaks in Porous Medical<br>Packaging by Dye Penetration | |
| | ASTM F 88/F88M 2015 | |
| | Standard Test Method for Seal Strength of Flexible Barrier Materials | |
| | ASTM F 1980 2016<br>Standard Guide for Accelerated Aging of Sterile Barrier Systems for<br>Medical Devices | |
| | ASTM D 5265 2009 R2016 | |
| | Standard Test Method for Bridge Impact Testing | |
| | ISO 10993-3:2014 | |
| | Biological Evaluation of Medical Devices - Part 3: Tests for Genotoxicity,<br>Carcinogenicity and Reproductive Toxicity | |
| | ISO 10993-4 :2017<br>Biological Evaluation of Medical Devices – Part 4: Selection of Tests for<br>Interactions with Blood | |
| | ISO 10993-5:2009-06 (R2017) | |
| Performance<br>Data cont | Biological Evaluation of Medical Devices - Part 5: Teste for In Vitro<br>Cytotoxicity<br>ISO10993-10:2010 (R2016) | |
| | Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and<br>Skin Sensitization<br>ISO 10993-11:2017 | |
| | Biological Evaluation of Medical Devices - Part 11: Tests for Systemic<br>Toxicity<br>ISO 10993-12:2012 | |
| | Biological Evaluation of Medical Devices - Part 12: Sample Preparation<br>and Reference Materials<br>ISO 14971:2019 | |
| | Application of Risk Management to Medical Devices<br>AAMI ST72 :2019 | |
| | Bacterial Endotoxins-test Methods, Routine Monitoring, And Alternatives<br>to Batch Testing<br>AAMI TIR28:2016 | |
| | Product Adoption and Process Equivalence for Ethylene Oxide<br>Sterilization | |
| | The following performance data was provided in support of the substantial<br>equivalence determination. | |
| | Biocompatibility testing<br>The biocompatibility evaluation for the Ventrax™ Delivery System was<br>conducted in accordance with the FDA Blue Book Memorandum #G95-1<br>"Use of International Standard ISO-10993, ‘Biological Evaluation of<br>Medical Devices Part 1: Evaluation and Testing,’" May 1, 1995, and<br>International Standard ISO 10993-1 “Biological Evaluation of Medical<br>Devices - Part 1: Evaluation and Testing Within a Risk Management<br>Process," as recognized by FDA. The battery of testing included the<br>following tests: | |
| | Cytotoxicity Sensitization Irritation Acute Systemic Toxicity | |
| | Performance Testing-Bench | |
| Performance<br>Data cont. | • Label Adherence and Condition<br>• Tube OD<br>• Sheath Tip to Dilator Gap<br>• Straight Dilator Radiopacity<br>• Sheath Curve<br>• Pigtail Dilator Curve, Pre-Use<br>• Sheath and Pigtail Dilator Curve Orientation<br>• Dilator Tip ID<br>• Simulated Use – Insertion through Simulated Tissue<br>• Simulated Use Testing- Device Exchange Through Sheath<br>• Dilator and Sheath Assembly<br>• Dilator Protrusion Length | |
| Performance<br>Data cont | • Pigtail Dilator Curve, Post-Use<br>• Guidewire Backloading through Pigtail Dilator Valve<br>• HVA Attachment<br>• Sheath Valve Damage<br>• High Pressure Water Leak Testing<br>• Dilator Joint Strengths<br>• Sheath Free Length<br>• Sheath Marker Band Location<br>• Hemostasis Valve Liquid Leak Test (Short and Long Term)<br>• Sheath Joint Strengths<br>• Sheath Hub Thru-Hole ID<br>• Sheath Tip Radiopacity Test<br>• Sheath Tip ID<br>• Sheath Tip Pull Test<br>• Sheath Shaft Tensile Strength | |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, design, safety and performance testing,<br>the Ventrax™ Delivery System meets the requirements that are<br>considered essential for its intended use and is substantially equivalent to<br>the predicate device the Amplatzer® TorgVue® Delivery System | |
{4}------------------------------------------------
{5}------------------------------------------------
{6}------------------------------------------------
Merit Medical Systems, Inc.
- Pyrogenicity
- Hemocompatibility Hemolysis Complete ●
- Hemocompatibility Complement Activation .
- Hemocompatibility Thrombogenicity ●
The Ventrax™ Delivery System is considered an externally communicating device with circulating blood contact for a limited (< 24 hours) duration
{7}------------------------------------------------
K202610 Ventrax™ Delivery System Premarket Notification 510(k)
K072313 manufactured by AGA Medical Corporation.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.