K171452 · Boston Scientific Corporation · DQY · Aug 21, 2017 · Cardiovascular
Device Facts
Record ID
K171452
Device Name
MAMBA and MAMBA FLEX Microcatheters
Applicant
Boston Scientific Corporation
Product Code
DQY · Cardiovascular
Decision Date
Aug 21, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The MAMBA™ and MAMBA™ Flex Microcatheters are intended to provide support to facilitate the placement of guidewires in the coronary vasculature, and can be used to exchange one guidewire for another. The microcatheters are also intended to assist in the delivery of contrast media into the coronary vasculature.
Device Story
MAMBA and MAMBA Flex are percutaneous microcatheters used in coronary vasculature; provide platform for guidewire exchange without losing position; assist in delivery of contrast media and therapeutic agents. Device features inner lumen for 0.014 in. guidewires; coil-reinforced shaft for torque control; hydrophilic coating on distal portion for advancement; radiopaque marker at distal tip for visualization. Used by physicians in clinical settings. Benefits include improved wire support to reach lesions and procedural efficiency during guidewire exchange.
Clinical Evidence
No clinical data. Evidence consists of bench testing (per FDA Class II Special Controls for PTCA catheters) and animal testing to verify mechanical performance, torque response, and biocompatibility.
Technological Characteristics
Percutaneous catheter; coil-reinforced shaft; hydrophilic distal coating; radiopaque distal marker; single luer hub. Compatible with 0.014 in. guidewires. Lengths: 135cm (MAMBA/Flex) and 150cm (Flex). Biocompatibility per ISO 10993.
Indications for Use
Indicated for patients requiring additional wire support to reach lesions in the coronary vasculature and for the delivery of contrast media or therapeutic agents to selected sites.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K161126 — ASAHI Corsair Pro · Asahi Intecc Co., Ltd. · Aug 25, 2016
K210647 — SuperCross microcatheter, SuperCross FT microcatheter, SuperCross AT microcatheter · Vascular Solutions, LLC · Apr 30, 2021
K182360 — Teleport Microcatheter · Orbusneich Medical Trading, Inc. · Nov 9, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
per 21 CFR §807.92
| Sponsor: | Boston Scientific Corporation<br>300 Boston Scientific Way<br>Marlborough, Massachusetts 01752<br>USA |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Sherry Kollmann<br>Three Scimed Place<br>Maple Grove, MN 55311-1566<br>Phone: 763-494-1111<br>Fax: 763-494-2981<br>e-mail: sherry.kollmann@bsci.com |
| Prepared | 16 May 2017 |
| Proprietary Name | MAMBA™ and MAMBA™ Flex Microcatheters |
| Common Name | Percutaneous Catheter |
| Product Code | DQY |
| Classification | Class II, 21 CFR Part 870.1250 |
| Predicate Device | ASAHI Corsair Microcatheter:<br>K151103 (23 June 2015)<br>K083127 (03 February 2009) |
| Reference Device | Renegade Hi-Flo Microcatheter<br>K140329 (15 May, 2014) |
{1}------------------------------------------------
## Device Description
The MAMBA™ and MAMBA™ Flex devices (referred to as the "MAMBA Microcatheters") are designed to provide a platform for the user to exchange guidewires without losing their position in the vasculature.
The MAMBA device is available in a 135cm length, while MAMBA Flex is available in 135cm and 150cm lengths.
The MAMBA Flex models have a smaller outer diameter and a more flexible distal portion than the MAMBA device.
Marks on the proximal portion of the catheter shaft and a radiopaque marker at the distal tip facilitate placement of the device.
The device's inner lumen permits the use of 0.014 in./0.36 mm or smaller quidewires. The shaft consists of a coil which allows the device to be torqued to assist in advancement. The distal portion of the shaft is coated hydrophilic coating to assist in advancement.
The proximal end has a single luer hub for flushing or injecting through the inner lumen and is for guide wire lumen access during a guidewire exchange.
The MAMBA Microcatheters are to be used in patients where vessel conditions require additional wire support to reach the lesion and will deliver radiopaque media and therapeutic agents to selected sites in coronary vasculature.
#### Indications for Use / Intended Use
The MAMBA™ and MAMBA™ Flex Microcatheters are intended to provide support to facilitate the placement of guidewires in the coronary vasculature, and can be used to exchange one guidewire for another.
The microcatheters are also intended to assist in the delivery of contrast media into the coronary vasculature.
## Comparison of Technological Characteristics
The MAMBA Microcatheters incorporate substantially equivalent device materials and design, packaging materials and design, fundamental technology, manufacturing processes, sterilization process and intended use as those featured in the ASAHI Corsair Microcatheter, K151103 and K083127.
{2}------------------------------------------------
#### Performance Data
The MAMBA Microcatheters were subjected to testing according to the requirements of Guidance for Industry and FDA Staff – Class II Special Controls for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters, September 8, 2010 (non-balloon component tests were applicable). Bench, animal and biocompatibility testing in accordance with ISO 10993 were performed to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during the testing; therefore, this device may be considered substantially equivalent to the predicate device.
#### Conclusion
Based on the indications for use, technological characteristics, and safety and performance testing, the MAMBA Microcatheters have been shown to be appropriate for their intended use and are considered to be substantially equivalent to the Asahi Corsair Microcatheter.
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 21, 2017
Boston Scientific Corporation Ms. Sherry Kollmann Principal Regulatory Affairs Specialist Three Scimed Place Maple Grove, Minnesota 55311
Re: K171452
Trade/Device Name: MAMBA and MAMBA Flex Microcatheters Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: July 21, 2017 Received: July 24, 2017
Dear Ms. Kollmann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{4}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.