Venture 038 catheter

K171335 · Vascular Solutions, Inc. · DQY · Jun 7, 2017 · Cardiovascular

Device Facts

Record IDK171335
Device NameVenture 038 catheter
ApplicantVascular Solutions, Inc.
Product CodeDQY · Cardiovascular
Decision DateJun 7, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Venture 038 catheter is indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the peripheral vasculature. It may also be used for manual delivery of saline solution or diagnostic contrast agents.

Device Story

Venture 038 catheter is an over-the-wire, single-lumen catheter featuring a radiopaque, deflectable distal tip. Operated by a physician in a clinical setting, the device allows for manual steering and control of a guidewire (0.035" or 0.038") to navigate peripheral vasculature. The distal tip deflection is actuated via a control knob on the proximal handle. The device supports guidewire placement and facilitates manual delivery of saline or diagnostic contrast agents. It is compatible with 6F introducer sheaths and 8F guide catheters. By providing enhanced steerability and support, the device assists clinicians in accessing difficult-to-reach vascular regions, potentially improving procedural efficiency and success in peripheral interventions.

Clinical Evidence

Bench testing only. No clinical data. Verification testing included simulated use, kink radius, tip deflection, tip rotation, tensile strength, liquid leak, injection pressure, aspiration, corrosion, and particulate count. Biocompatibility testing performed per ISO 10993-1 included cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, hemolysis, complement activation, in vitro hemocompatibility, coagulation, and thrombogenicity. All tests met specified acceptance criteria.

Technological Characteristics

Over-the-wire single lumen catheter; radiopaque deflectable distal tip; hydrophilic coating on distal 45 cm; working lengths of 120 cm or 70 cm; compatible with 0.035"/0.038" guidewires, 6F introducer sheaths, and 8F guide catheters. Sterilized via Ethylene Oxide (EO) to SAL 10-6. Biocompatibility per ISO 10993-1.

Indications for Use

Indicated for patients requiring guidewire access to discrete regions of the peripheral vasculature for diagnostic or interventional procedures.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Public Health Service Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles. June 7, 2017 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Vascular Solutions, Inc. Ms. Beka Vite Sr. Regulatory Product Specialist 6464 Sycamore Court North Minneapolis, MN 55369 Re: K171335 Trade/Device Name: Venture 038 catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: May 5, 2017 Received: May 8, 2017 Dear Ms. Vite: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Fernando Aguel-S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K171335 Device Name Venture 038 catheter Indications for Use (Describe) The Venture 038 catheter is indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the peripheral vasculature. It may also be used for manual delivery of saline solution or diagnostic contrast agents. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------------------|--| | <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | | | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary [As required by 21 CFR 807.92] Date Prepared: May 5, 2017 510(k) Number: K171335 ## Submitter's Name / Contact Person | <b>Manufacturer</b> | <b>Contact Person</b> | |--------------------------------------|-----------------------------------| | Vascular Solutions, Inc. | Beka Vite | | 6464 Sycamore Court North | Sr. Regulatory Product Specialist | | Minneapolis, MN<br>55369 USA | Tel: 763-656-4300 | | Establishment Registration # 2134812 | Fax: 763-656-4253 | #### General Information | Trade Name | Venture 038 catheter | |---------------------|----------------------------------------------------------------------------| | Common / Usual Name | Catheter, Percutaneous | | Classification Name | Class II | | Predicate Device | K040922 and K042910 - 0.014" Venture Wire Control Catheter, Velocimed Inc. | | Reference Device | K142065 - Turnpike Catheter, Vascular Solutions, Inc. | ## Device Description The Venture 038 catheter is an over-the-wire single lumen catheter with a radiopaque deflectable distal tip. The deflectable distal tip is actuated by rotating a knob on the control handle of the proximal end of the catheter. The Venture 038 catheter has a working length of 120 cm or 70 cm and has a hydrophilic coating on the distal 45 cm of the catheter. The Venture 038 catheter provides support to 0.035" guidewires and is compatible with 6F introducer sheaths and 8F guide catheters. #### Indications for Use The Venture 038 catheter is indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the peripheral vasculature. It may also be used for manual delivery of saline solution or diagnostic contrast agents. #### Technological Characteristics Comparison A comparison of technological differences between the subject and predicate device is provided below. | Characteristic | Subject Device | Predicate Device | |---------------------|------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------| | Intended Use | Directing, steering, controlling, and supporting a guidewire to access discrete regions of the vasculature | Directing, steering, controlling, and supporting a guidewire to access discrete regions of the vasculature | | Anatomical Location | Peripheral vasculature | Coronary and peripheral vasculature | | | Venture 038 Catheter | Venture Wire Control Catheter | {4}------------------------------------------------ | Characteristic | Subject Device | Predicate Device | |----------------------------|------------------------------------------------------------|-------------------------------| | | Venture 038 Catheter | Venture Wire Control Catheter | | Working Length | 120 cm and 70 cm | 140 cm and 70 cm | | Guidewire<br>Compatibility | 0.035" or 0.038" | 0.014" | | Lubricous Coating | Hydrophilic coating | | | Sterility | Ethylene Oxide (EO) – Sterility Assurance Level (SAL) 10-6 | | ## Substantial Equivalence and Summary of Studies The technological differences between the subject and predicate devices have been evaluated through biocompatibility and bench tests to provide evidence that the Venture 038 catheter is substantially equivalent to the predicate device. The Venture 038 catheter is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design has been verified through the following tests: - Simulated Use - Kink Radius - Visual Inspection ● - . Tip Deflection - Tip Rotation - Tensile Strength ● - . Liquid Leak - Liquid Leak Under Pressure - Injection Pressure - Aspiration - Corrosion - Particulate Count ● Device samples passed the following biocompatibility tests performed in accordance with ISO 10993-1: - . Cytotoxicity - Sensitization - Irritation - Acute Systemic Toxicity - Pyrogenicity ● - Hemolysis - Complement Activation - In Vitro Hemocompatibility - Coagulation ● - Thrombogenicty The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the Venture 038 catheter is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...