← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K163000

# AMPLATZER TorqVue 45x45 Delivery Sheath (K163000)

_Aga Medical COrporation (Wholly Owned by St. Jude Medical CO · DQY · Dec 23, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K163000

## Device Facts

- **Applicant:** Aga Medical COrporation (Wholly Owned by St. Jude Medical CO
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Dec 23, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The AMPLATZER™ TorqVue™ 45°x45° Delivery Sheath is intended to provide a pathway through which devices are introduced within the chambers of the heart.

## Device Story

The AMPLATZER TorqVue 45°x45° 80 cm Delivery Sheath is a percutaneous catheter system used to provide a pathway for device introduction into heart chambers. The system consists of a sheath, a dilator, and a flush adapter. The sheath body and dilator are radiopaque for visualization under fluoroscopy. The physician inserts the sheath and dilator into the vasculature; the dilator eases penetration and minimizes vessel trauma. Once positioned, the dilator is removed, and the sheath serves as a conduit for other devices. The device is used in a clinical setting (e.g., cath lab) by a physician. The 80 cm length is a modification of the existing 100 cm product family, intended to provide consistent length options for various cardiac procedures. The device facilitates minimally invasive cardiac interventions by providing a stable, navigable access route.

## Clinical Evidence

Bench testing only. No clinical data provided. Substantial equivalence is supported by design verification and validation testing previously performed for the predicate device family, supplemented by packaging design verification for the subject device.

## Technological Characteristics

Percutaneous delivery sheath and dilator system. Components include a radiopaque sheath, radiopaque dilator, and flush adapter. Available in 12 Fr and 14 Fr sizes with an 80 cm usable length. Operates via manual insertion under fluoroscopic guidance. No energy source or software components.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- AMPLATZER TorqVue LA and AMPLATZER TorqVue 45°x45° (14 Fr) Delivery Sheaths ([K120120](/device/K120120.md))
- AMPLATZER TorqVue 45°x45° Delivery Sheath 9-13 Fr sizes ([K083214](/device/K083214.md))

## Reference Devices

- TorqVue Delivery system (ITV) ([K072313](/device/K072313.md))
- AMPLATZER TorqVue LA (TVLA 1 and TVLA 2) ([K120120](/device/K120120.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of three human profiles facing right, stacked on top of each other to create a sense of depth and unity. The profiles are rendered in a dark color, contrasting with the white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 23, 2016

AGA Medical Corporation (wholly Owned By St. Jude Medical Co) Mary Johnson Regulatory Affairs Specialist I 5050 Nathan Lane Plymouth, Minnesota 55442

Re: K163000

Trade/Device Name: The AMPLATZER TorqVue 45° x 45° Delivery Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 29, 2016 Received: December 1, 2016

Dear Mary Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Brian D. Pullin -S

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K163000

Device Name

AMPLATZER™ TorqVue™ 45° x45° Delivery Sheath

Indications for Use (Describe)

The AMPLATZER™ TorqVue™ 45° x 45° Delivery Sheath is intended to provide a pathway through which devices are introduced within the chambers of the heart.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(K) SUMMARY

| 510(k) Number           | K163000                                                                                                                                                                                                            |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:              | AGA Medical Corporation<br>A wholly owned subsidiary of St. Jude Medical<br>5050 Nathan Lane<br>Plymouth, MN                                                                                                       |
| Contact<br>Person:      | Mary Johnson,<br>Regulatory Affairs Specialist I<br>Phone: 651-756-2213<br>Fax: 651-756-5744<br>E-mail: mjohnson16@sjm.com                                                                                         |
| Date Prepared:          | October 27, 2016                                                                                                                                                                                                   |
| Trade Name:             | AMPLATZERTM TorqVueTM 45°x45° Delivery Sheath                                                                                                                                                                      |
| Common<br>Name:         | Catheter Delivery System                                                                                                                                                                                           |
| Classification:         | Class II, 21 CFR 870.1250<br>Catheter, Percutaneous                                                                                                                                                                |
| Product Code:           | DQY                                                                                                                                                                                                                |
| Predicate<br>Device(s): | Primary Predicate-AMPLATZERTM TorqVueTM LA and AMPLATZER TorqVue<br>45°x45° (14 Fr) Delivery Sheaths (K120120)<br><br>Secondary Predicate-AMPLATZERTM TorqVueTM 45°x45° Delivery Sheath 9-13 Fr<br>sizes (K083214) |

{4}------------------------------------------------

| Device<br>Description:  | The AMPLATZER™ TorqVue™ 45°x45° 80 cm Delivery Sheath (TV45x45 80cm) is<br>an extension of the AMPLATZERTM TorqVue 45x45 Delivery Sheath (TV45x45<br>100cm) product family (K120120 and K083214) and is available in both a 12Fr and<br>14Fr sheath size.<br><br>The TV45x45 80cm is identical to the cleared TV45x45 (100cm) family of sheaths with<br>the exception of having a shorter 80 cm length and the tip taper length of the 12Fr<br>(80cm) dilator modeled on the taper length of the existing 14 Fr TV45x45 (100cm)<br>dilator.<br><br>The 80 cm length is consistent with several TorqVue Delivery System products<br>including the TorqVue™ Delivery system (ITV) cleared under submission K072313 and<br>the TorqVue™ (TVLA) delivery sheaths cleared under submission K120120.<br><br>Similar to the predicate the TorqVue 45°x45° 80 cm Delivery Sheath includes a flush<br>adapter, sheath and dilator. <b>Figure 1</b> illustrates the delivery sheath and identifies the<br>following essential components:<br><b>Sheath (A)</b> -Catheter that enables device delivery by the physician. The body of             the sheath is radiopaque for visibility under fluoroscopy. For added visibility the             sheath also has a radiopaque marker band near the distal tip. <b>Dilator (B)</b> -Used to ease penetration of tissue and minimize vessel trauma during             sheath insertion into and through the vasculature. For added visibility the dilator             is radiopaque for visibility under fluoroscopy. <b>Flush Adaptor</b> -The flush adaptor for both the 12F(C) and 14F (D) sheaths             enables connection with syringes for flushing the sheath lumen.<br>Figure 1. AMPLATZER TV 45x45 (80cm) Components<br>Image: [AMPLATZER TV 45x45 (80cm) Components] |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use: | The AMPLATZER™ TorqVue™ 45°x45° Delivery Sheath is intended to provide a<br>pathway through which devices are introduced within the chambers of the heart.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{5}------------------------------------------------

| Comparison to predicate:       | AMPLATZER             TM TorqVue 45°x45° (80cm) Delivery Sheath is an extension of the<br>AMPLATZER             TM TorqVue 45°x45° product family (K083214, K120120) line.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                        |                              |                           |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|------------------------------|---------------------------|
|                                | The AMPLATZER             TM TorqVue 45°x45° 80cm subject device is available in either 12 or<br>14Fr and shares the same design (with the exception of being 80cm length versus<br>100cm), materials, accessories, and manufacturing processes as the predicate. The<br>dilator used with the 12 Fr subject device will be updated to have the same tip geometry<br>as the predicate TV45x45 100cm 14 Fr device (the same dilator tip taper length and<br>80cm length is also consistent with the marketed AMPLATZER TorqVue LA (TVLA1<br>and TVLA 2) devices, K120120).<br>Table 1 summarizes the changes being made to 12 and 14 Fr versions of the TV45x45             device. |                                        |                              |                           |
|                                | <b>Table 1. TV45x45 Change Summary</b>                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                        |                              |                           |
|                                | TV45x45 80 cm (Subject)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | TV45x45 14Fr 100cm (Predicate-K120120) | TVLA 1&2 (Predicate-K120120) |                           |
|                                | Device Usable Length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 80 cm                                  | 100 cm                       | 80 cm                     |
|                                | Tip Taper Length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Extended tip taper length              | Extended tip taper length    | Extended tip taper length |
|                                | 1The 14Fr size was cleared under K120120; the remaining Fr sizes in the TV45x45             family were cleared earlier under K083214.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                        |                              |                           |
| Functional and Safety Testing: | The intended use, components, materials and fundamental design of the TV45x45 80 cm             are identical to the cleared predicate TV45x45 (100cm) device (TV45x45 device family             cleared in two submissions K083214 and K120120). Design Verification and Design             Validation testing have been previously performed and provided in cleared applications             for the predicate device (TV 45x45 (100 cm, K120120 and K083214), TVLA1 and             TVLA 2 Delivery Sheath (K120120) devices). Separate packaging design verification             has been completed on the subject device.                                                    |                                        |                              |                           |
| Conclusion:                    | AGA/SJM considers the TorqVue 45°x45° 80 cm Delivery Sheath to be substantially             equivalent to the TorqVue 45°x45° 100 cm Delivery Sheath. This conclusion is based             upon the devices' similarities in principles of operation, technology, materials and             indications for use. The devices are identical with the exception of a length and dilator             tip angle change.                                                                                                                                                                                                                                                                |                                        |                              |                           |

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K163000](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K163000)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
