K162467 · Vascular Solutions, Inc. · DQY · Feb 1, 2017 · Cardiovascular
Device Facts
Record ID
K162467
Device Name
Twin-Pass Torque
Applicant
Vascular Solutions, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Feb 1, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Twin-Pass Torque catheter is intended to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and to subselectively infuse/deliver diagnostic and therapeutic agents.
Device Story
Twin-Pass Torque is a dual-lumen percutaneous catheter; features over-the-wire (OTW) lumen and rapid exchange (RX) delivery lumen. Used by physicians in coronary/peripheral vascular procedures to access specific regions, facilitate guidewire exchange, and deliver diagnostic/therapeutic agents. Device includes radiopaque platinum-iridium marker bands for visualization and positioning marks on shaft. Operates via manual manipulation by clinician under fluoroscopic guidance. Benefits include improved access and guidewire management in complex vascular anatomy.
Clinical Evidence
Bench testing only. Verification testing included kink resistance, guidewire insertion, tip/shaft flexibility, torque control/strength/transmission, tensile strength, hub luer tests, and coating lubricity/particulate. Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, and hemocompatibility. All tests met acceptance criteria.
Technological Characteristics
Dual-lumen percutaneous catheter; OTW and RX lumens. Materials include platinum-iridium marker bands. Features hydrophilic coating. Manual operation; no energy source. Sterilization method not specified. Design includes radiopaque markers and positioning marks.
Indications for Use
Indicated for accessing discrete regions of coronary and/or peripheral vasculature to facilitate guidewire placement/exchange and subselective infusion/delivery of diagnostic and therapeutic agents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K991601 — DUAL LUMEN CATHETER, MODEL DL-35-90 · Endologix, Inc. · Oct 28, 1999
K062877 — TWIN-PASS .023 DUAL ACCESS CATHETER, MODEL 5230 · Vascular Solutions, Inc. · Jan 18, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 1, 2017
Vascular Solutions, Inc. Ms. Lisa Gallatin Director of Regulatory 6464 Sycamore Court North Minneapolis, Minnesota 55369
Re: K162467
Trade/Device Name: Twin-Pass Torque Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: December 27, 2016 Received: December 28, 2016
Dear Ms. Gallatin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando
Aguel-S
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K162467
Device Name Twin-Pass Torque
Indications for Use (Describe)
The Twin-Pass catheter is intended to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and to subselectively infuse/deliver diagnostic agents.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
[As required by 21 CFR 807.92]
Date Prepared: December 21, 2016
510(k) Number: __K162467
# Submitter's Name / Contact Person
Manufacturer Vascular Solutions, Inc. 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812 Contact Person Lisa Gallatin, RAC Director of Regulatory Tel: 763-656-4300 Fax: 763-656-4253
# General Information
| Trade Name | Twin-Pass Torque |
|---------------------|-------------------------------------------------------------------|
| Common / Usual Name | Dual access catheter |
| Classification Name | Catheter, percutaneous |
| Predicate Device | K083784, Twin-Pass dual access catheter, Vascular Solutions, Inc. |
| Reference Device | K142065 and K151981, Turnpike catheter |
### Device Description
The Twin-Pass Torque is a dual lumen catheter consisting of an over-the-wire (OTW) lumen that runs the full length of the catheter, exiting just proximal to the distal tip, and a rapid exchange (RX) delivery lumen on the distal segment. The two exit ports are each marked by a radiopaque platinum-iridium marker band and the device shaft has two positioning marks located at 95cm (single mark) and 105cm (double marks) from the distal tip, which are not visible under fluoroscopy.
# Intended Use
The Twin-Pass Torque catheter is intended to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and to subselectively infuse/deliver diagnostic and therapeutic agents.
# Technological Characteristics Comparison
Twin-Pass Torque is similar in design to the predicate device and both are dual lumen, percutaneous catheters intended to access discrete regions of the coronary and peripheral vasculature, facilitate placement and exchange of guidewires and subselectively infuse agents. With the exception of dimensional, material and package configuration differences, the Twin-Pass Torque is similar in design and technological characteristics to the predicate device. The dimensional, material and package configuration differences were successfully evaluated in performance tests.
{4}------------------------------------------------
# Substantial Equivalence and Summary of Studies
The technological differences between the subject and predicate devices have been evaluated through performance and biocompatibility tests and results did not raise new questions of safety or effectiveness. The Twin-Pass dual access catheter is substantially equivalent to the specified predicate device based on comparisons of the device functionality, technological characteristics, and indications for use. The device design has been verified through the following tests:
- . Kink resistance
- . Guidewire insertion
- . Tip flexibility
- . Distal shaft flexibility
- Proximal shaft support ●
- Coating lubricity/Durability ●
- Radiopacity
- Guidewire deflection ●
- Distal tip length ●
- . Torque control
- Torque transmission ●
- Tensile strength .
- . Torque strength
- Torque robustness
- Hub luer tests (air leak, burst, compatibility)
- Hydrophilic coating particulate
- . Package integrity
Device samples passed the following biocompatibility tests performed in accordance with ISO 10993-1:
- . Cytotoxicity
- Sensitization
- . Irritation
- Systemic toxicity ●
- . Pyrogenicity
- . Hemocompatibility
The results of the verification tests met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, the Twin-Pass Torque dual access catheter is substantially equivalent to the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.