← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K152249

# ASAHI Corsair Armet (K152249)

_Asahi Intecc Co., Ltd. · DQY · Mar 4, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K152249

## Device Facts

- **Applicant:** Asahi Intecc Co., Ltd.
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Mar 4, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another. The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature. This device should not be used in coronary vasculature or neuro vasculature.

## Device Story

ASAHI Corsair Armet is a percutaneous catheter designed for peripheral vasculature use. Device consists of distal tip, shaft tube, protector, and connector. Distal tip is tapered; shaft and tip feature hydrophilic coating for smooth navigation; inner lumen lined with PTFE to facilitate guidewire exchange. Operated by clinicians in clinical settings to provide support for guidewire placement and assist in contrast media delivery. Device functions as a mechanical support tool; no electronic or software components. Benefits include improved guidewire tracking and contrast delivery in peripheral vessels.

## Clinical Evidence

Bench testing only. Performed tests: appearance/dimensions, corrosion resistance, force at break, liquid/air leakage, radio-detectability, torque transmission/durability, slide durability, and kink resistance. All tests met acceptance criteria.

## Technological Characteristics

Percutaneous catheter; polyamide-elastomer shaft; hydrophilic coating on distal tip and shaft; PTFE inner lumen; tapered distal tip. Sterilized via Ethylene Oxide (SAL 10^-6). Single-use. Dimensions: 600-1500 mm length; 0.75 mm distal OD; 0.83 mm proximal OD.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- ASAHI Corsair Microcatheter ([K151103](/device/K151103.md))

## Reference Devices

- ASAHI Fubuki 043 ([K141981](/device/K141981.md))
- ASAHI Tornus Support Catheter ([K051772](/device/K051772.md))
- ASAHI PTCA Guide Wire Marker Wire ([K022762](/device/K022762.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 4, 2016

ASAHI Intecc Co., Ltd. % Ms. Candace Cederman Senior Regulatory Affairs Consultant CardioMed Device Consultants, LLC 5523 Research Park Drive, Suite 205 Baltimore, MD 21228

Re: K152249

Trade/Device Name: ASAHI Corsair Armet Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: February 9, 2016 Received: February 11, 2016

Dear Ms. Cederman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K152249

Device Name ASAHI Corsair Armet

#### Indications for Use (Describe)

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another.

The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature.

This device should not be used in coronary vasculature or neuro vasculature.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for ASAHI INTECC CO., LTD. The logo features a stylized red "A" symbol on the left, followed by the company name in dark green, sans-serif font. The company name is written in all capital letters and includes a comma between "CO" and "LTD".

1703 Wakita-cho, Moriyama-ku, Nagoya, Aichi 463-0024 Tel. +81-52-768-1211 Fax. +81-52-768-1221 Branch offices: Tokyo, Nagoya, Osaka, Hong Kong, Amsterdam, Singapore, Beijing Research Facilities and Factories: Osaka, Seto, Thailand, Hanoi

#### 510(k) Summary [as required by 21 CFR 807.92(c)]

# ASAHI Corsair Armet

### 510(k) 152249

| DATE PREPARED:         | February 25, 2016                                                                                                                                                                                              |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT              | ASAHI Intecc Co., Ltd.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya, Aichi 463-0024, Japan                                                                                                                        |
| OFFICIAL CORRESPONDENT | Yoshi Terai<br>President, CEO<br>ASAHI Intecc USA, Inc.<br>2500 Red Hill Avenue, Suite 210<br>Santa Ana, CA 92705<br>Tel: (949) 756-8252<br>FAX: (949) 756-8165<br>e-mail: ASAHI.ra-fda@ASAHI-intecc.com       |
| TRADE NAME:            | ASAHI Corsair Armet                                                                                                                                                                                            |
| DEVICE CLASSIFICATION: | Class 2 per 21 CFR §870.1250                                                                                                                                                                                   |
| CLASSIFICATION NAME:   | Percutaneous Catheter                                                                                                                                                                                          |
| PRODUCT CODE           | DQY- Catheter, Percutaneous                                                                                                                                                                                    |
| PREDICATE DEVICES:     | Primary Predicates:<br>ASAHI Corsair Microcatheter (K151103)<br><br>Reference Devices:<br>ASAHI Fubuki 043 (K141981)<br>ASAHI Tornus Support Catheter (K051772)<br>ASAHI PTCA Guide Wire Marker Wire (K022762) |

# Intended Use/Indications for Use

The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another.

The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature.

This device should not be used in coronary vasculature or neuro vasculature.

{5}------------------------------------------------

# DESCRIPTION:

The ASAHI Corsair Armet consists of a distal tip, a shaft tube that is inserted into vasculature, a protector, and a connector for catheter control and infusion of contrast media. The device has a hydrophilic coating on the outer surface of distal tip and the shaft tube to provide a smooth transition in blood vessels. The distal tip of the Corsair Armet has a tapered shape. PTFE is applied to the inner lumen of the catheter for the purposes of a smooth transition and exchange of guidewires.

# Comparison with predicate devices:

Comparisons of the ASAHI Corsair Armet and predicate devices show that the technological characteristics of the ASAHI Corsair Armet such as the components, design, materials, sterilization method, shelf life and operating principle are identical or similar to the currently marketed predicate devices.

The intended use/indications of the Subject Device are a subset of the primary predicates. There are specific design features of the Subject device that are similar to the primary predicate but not identical. Additional reference devices have been used to demonstrate equivalence for these similar features.

| Name of Device               | ASAHI Corsair Armet                                                                                                                                                                                                                                                                                                                                                                                       | ASAHI Corsair Microcatheter                                                                                                                                                                                                                                                                                                                                            |
|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                       | Current Application                                                                                                                                                                                                                                                                                                                                                                                       | K151103                                                                                                                                                                                                                                                                                                                                                                |
| Intended Use and Indications | The ASAHI Corsair Armet is intended to provide support to facilitate the placement of guide wires in the peripheral vasculature, and can be used to exchange one guide wire for another.<br><br>The ASAHI Corsair Armet is also intended to assist in the delivery of contrast media into the peripheral vasculature.<br><br>This device should not be used in coronary vasculature or neuro vasculature. | The ASAHI Corsair Microcatheter is intended to provide support to facilitate the placement of Guide wires in the coronary and peripheral vasculatures and can be used to exchange one guide wire for another.<br><br>The Corsair Microcatheter is also intended to assist in the delivery of contrast media into the coronary, peripheral, and abdominal vasculatures. |
| Sterilization                | Provided sterile via Ethylene Oxide to SAL10-6                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                        |
| Target Body Location         | Peripheral                                                                                                                                                                                                                                                                                                                                                                                                | Peripheral, Coronary                                                                                                                                                                                                                                                                                                                                                   |
| Hydrophilic coating          | Yes                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                        |
| Effective Length             | 600-1500 mm                                                                                                                                                                                                                                                                                                                                                                                               | 900 - 1500mm                                                                                                                                                                                                                                                                                                                                                           |
| Nominal OD                   | Distal : 0.75mm<br>Proximal : 0.83 mm                                                                                                                                                                                                                                                                                                                                                                     | Distal : 0.75mm<br>Proximal : 1.0mm                                                                                                                                                                                                                                                                                                                                    |
| Catheter Shaft Material      | Polyamide-elastomer                                                                                                                                                                                                                                                                                                                                                                                       | Polyamide-elastomer                                                                                                                                                                                                                                                                                                                                                    |
| Single Use                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                        |

{6}------------------------------------------------

# Non Clinical testing / Performance Data:

Non clinical laboratory testing was performed on the ASAHI Corsair Armet to determine substantial equivalence. The following testing/assessments were performed:

- Appearance/Dimensions ●
- Corrosion Resistance ●
- Force at Break ●
- Liquid Leakage under Pressure ●
- Air Leakage ●
- Leak and Damage under High Static Pressure ●
- Radio-Detectability ●
- Torque Transmission .
- Slide Durability ●
- Kink Resistance ●
- . Torque Durability

The in vitro bench tests demonstrated that the ASAHI Corsair Armet met all acceptance criteria and performed similarly to the predicate devices. Performance data demonstrate that the device functions as intended, and is substantially equivalent to the predicate devices.

# BIOCOMPATIBILITY:

The ASAHI Corsair Armet was compared to the predicate devices. Based on similarities of the materials used in the subject device to its predicates and reference devices, the biocompatibility of the ASAHI Corsair Armet was verified to be the same as those of the predicates and reference devices.

# CONCLUSION:

The ASAHI Corsair Armet has identical intended use, a subset of the indications, and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate devices. Performance data demonstrates that the device functions as intended.

Therefore, the ASAHI Corsair Armet is substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K152249](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K152249)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
