K151981 · Vascular Solution, Inc. · DQY · Aug 13, 2015 · Cardiovascular
Device Facts
Record ID
K151981
Device Name
Turnpike catheter
Applicant
Vascular Solution, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 13, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Turnpike catheters are intended to access discrete regions of the coronary and/or peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and to subselectively infuse/deliver diagnostic and therapeutic agents.
Device Story
Turnpike catheter is a single-lumen device for coronary and peripheral vasculature access. Used by physicians to facilitate guidewire placement/exchange and deliver diagnostic/therapeutic agents. Device features a shaft with two polymer layers encapsulating a braid and dual-layer coil; includes radiopaque tip (polymer or gold-plated) and hydrophilic coating. Compatible with 0.014" guidewires and 5F guide catheters. Available in various configurations and two working lengths.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Single-lumen catheter; shaft constructed of two polymer layers encapsulating a braid and dual-layer coil; radiopaque tip (polymer or gold-plated); hydrophilic coating; compatible with 0.014" guidewires and 5F guide catheters.
Indications for Use
Indicated for accessing discrete regions of coronary and/or peripheral vasculature; facilitating placement and exchange of guidewires; and subselective infusion/delivery of diagnostic and therapeutic agents.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K210647 — SuperCross microcatheter, SuperCross FT microcatheter, SuperCross AT microcatheter · Vascular Solutions, LLC · Apr 30, 2021
K112100 — M2 NAVIGATION CATHETER AND GUIDELINER NAVIGATION CATHETER · Vascular Solutions, Inc. · Mar 20, 2012
K251277 — CoraForce Microcatheter, CoraFlex Microcatheter · Reflow Medical, Inc. · Aug 22, 2025
Submission Summary (Full Text)
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2015
Vascular Solutions, Inc. Beka Vite Regulatory Product Specialist 6464 Sycamore Court North Minneapolis, MN. 55369
Re: K151981
Trade/Device Name: Turnpike Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: II Product Code: DQY Dated: July 16, 2015 Received: July 17, 2015
Dear Ms. Vite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known): _K151981
Device Name: Turnpike catheters
Indications for Use: The Turnpike catheters are intended to access discrete regions of the coronary and/or peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and to subselectively infuse/deliver diagnostic and therapeutic agents.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 2 510(k) Summary
[As required by 21 CFR 807.92]
Date Prepared: July 15, 2015
510(k) Number: _______________________________________________________________________________________________________________________________________________________________
## Submitter's Name / Contact Person
Manufacturer Vascular Solutions, Inc. 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812
Contact Person Beka Vite Regulatory Product Specialist Tel: 763-656-4300 Fax: 763-656-4253
## General Information
| Trade Name | Turnpike catheter |
|---------------------|-------------------------------------------------------|
| Common / Usual Name | Catheter |
| Classification Name | 21 CFR 870.1250, DQY, Percutaneous catheter, Class II |
| Predicate Device | K142065, Turnpike catheter (Vascular Solutions, Inc.) |
#### Device Description
The Turnpike catheters are single lumen catheters designed for use in the coronary and/or peripheral vasculature. The shaft is constructed of two polymer layers that encapsulate a braid and a dual-layer coil. The Turnpike catheters have a radiopaque tip (polymer or gold-plated) and are available in various tip and shaft configurations and one of two working lengths. The Turnpike catheters are hydrophilic coated and are compatible with 0.014" guidewires and 5F guide catheters.
#### Intended Use / Indications
The Turnpike catheters are intended to be used to access discrete regions of the coronary and/or peripheral vasculature. They may be used to facilitate placement and exchange of guidewires and to subselectively infuse/deliver diagnostic and therapeutic agents.
## Technological Characteristics Comparison
There are no technological differences between the subject and predicate devices relevant to the labeling change being effected.
## Substantial Equivalence and Summary of Studies
Since there is no indication or technology change relevant to the labeling change being effected. no tests were necessary to demonstrate substantial equivalence. The addition of the contraindication and warning does not change the indications for use. Therefore, the Turnpike catheters are substantially equivalent to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.