SWITCH-IT CATHETER

K120707 · Vascular Solutions, Inc. · DQY · May 15, 2012 · Cardiovascular

Device Facts

Record IDK120707
Device NameSWITCH-IT CATHETER
ApplicantVascular Solutions, Inc.
Product CodeDQY · Cardiovascular
Decision DateMay 15, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Switch-It catheter is intended to be used in conjunction with guide catheters to facilitate placement and exchange of guidewires in the discrete coronary and peripheral vasculature.

Device Story

The Switch-It is a rapid-exchange catheter designed for use with 6F guide catheters to assist in placing and exchanging 0.014" guidewires within coronary and peripheral vasculature. The device features a 145 cm working length, comprising a 117 cm proximal stainless steel push wire and a 28 cm microcatheter-type shaft, joined by a funnel section for guidewire insertion. The distal 20 cm of the shaft includes a hydrophilic coating and a platinum-iridium marker band for fluoroscopic visualization. The push wire includes positioning marks at 95 cm and 105 cm from the distal tip. Operated by clinicians in interventional settings, the device facilitates guidewire management, improving procedural efficiency during vascular interventions.

Clinical Evidence

Bench testing only. Verification testing included simulated anatomy/concomitant device use, kink, tensile, torque, and dimensional verification. No clinical data provided.

Technological Characteristics

Rapid-exchange catheter; 145 cm total length (117 cm stainless steel push wire, 28 cm microcatheter shaft). Distal 20 cm hydrophilic coating; platinum-iridium marker band. Compatible with 0.014" guidewires and 6F guide catheters. Non-powered, mechanical device.

Indications for Use

Indicated for use in coronary and peripheral vasculature to facilitate the placement and exchange of 0.014" guidewires in conjunction with a 6F guide catheter.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ MAY 1 5 2012 Vascular Solutions, Inc. Traditional 510(k) Premarket Not Switch-It™ Catheter #### 2 510(k) Summary [As required by 21 CFR 807.92] 510(k) Number: K120707 ## Submitter's Information / Contact Person Manufacturer Contact Person Vascular Solutions, Inc. Ellie Gillespie Senior Regulatory Product Specialist 6464 Sycamore Court Minneapolis, MN 55369 USA Tel: 763-656-4230 (direct) Establishment Registration Number: 2134812 Fax: 763-656-4253 ## Date Prepared: March 7, 2012 Email: egillespie@vasc.com #### General Information | Trade Name | Switch-It tm catheter | |---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common / Usual Name | Catheter | | Classification Name | 21 CFR 870.1250, Percutaneous catheter | | Predicate Devices | GuideLiner tm V2 catheter (K112082 - Vascular Solutions, Inc.)<br>SuperCross tm microcatheter (K101659 - Vascular Solutions, Inc.) | ## Device Description The Switch-It catheter (Switch-It) is an addition to the VSI guide extension catheter product family. Switch-It is a rapid-exchange catheter intended for use with a standard 6F guide catheter to place and exchange standard length 0.014" guidewires. The working length of Switch-It is 145 cm, consisting of a 117 cm proximal stainless steel push wire attached to a 28 cm microcatheter-type shaft. The push wire and shaft are joined by a short funnel section that directs a proximally inserted guidewire into the shaft. The inner diameter of the shaft is sized to accommodate an 0.014" guidewire. The distal 20 cm of the shaft has a hydrophilic coating and a platinum-iridium marker band located 2 mm or less from the distal tip, which is visible using standard fluoroscopic methods. The push wire has two positioning marks located at 95 cm (single mark) and 105 cm (double mark) from the distal tip, respectively. #### Intended Use / Indications The Switch-It catheter is intended to be used in conjunction with guide catheters to facilitate placement and exchange of guidewires in the discrete coronary and peripheral vasculature. {1}------------------------------------------------ Vascular Solutions, Inc. Traditional 510(k) Premarket Notification Switch-It™ Catheter #### Substantial Equivalence and Summary of Studies Switch-It is substantially equivalent to the GuideLiner V2 catheter (GuideLiner V2) based on comparisons of intended use, device functionality, and technological characteristics. In addition, the intended use. microcatheter-like design, and materials of Switch-It are similar to the SuperCross microcatheter (SuperCross). Switch-It and the predicate devices are intended to be used in the coronary and peripheral vasculature to facilitate placement and exchange guidewires. The predicate devices are intended to facilitate placement and exchange of guidewires and other interventional devices; however, the smaller ID of Switch-It allows only the passage of guidewires, not interventional devices. The Switch-It intended use is a subset of the GuideLiner V2 and SuperCross indications, as Switch-It is only indicated to facilitate placement and exchange of guidewires and not other interventional devices. The Switch-It design was qualified through the following tests: - . Simulated anatomy/concomitant device use - . Kink - . Tensile - . Torque - . Dimensional verification Results of the verification testing did not raise new safety or performance questions. {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a staff with two snakes entwined around it, representing medicine and health. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 MAY 1 5 2012 Vascular Solutions, Inc. c/o Ms. Ellie Gillespie Senior Regulatory Product Specialist 6464 Sycamore Court Minneapolis. MN 55369 Re: K120707 Switch-It™ Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: March 7, 2012 Received: March 8, 2012 #### Dear Ms. Gillespie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {3}------------------------------------------------ Page 2 - Ms. Ellie Gillespie found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you dobilio povefice as v/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerelv yours. CM of Killebrew Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use K1207 510(k) Number (if known): Device Name: Switch-It" catheter Indications for Use: The Switch-It catheter is intended to be used in conjunction with guide catheters to facilitate placement and exchange of guidewires in the discrete coronary and peripheral vasculature. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) M.F. Kilburn on Sign-Off on of Cardiovascular Devices 1120707 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...