SCOUTPRO HEMOSTATIC VALVE

K101282 · Biotronik, Inc. · DQY · Jul 21, 2010 · Cardiovascular

Device Facts

Record IDK101282
Device NameSCOUTPRO HEMOSTATIC VALVE
ApplicantBiotronik, Inc.
Product CodeDQY · Cardiovascular
Decision DateJul 21, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The ScoutPro CS Lead Introducer System facilitates lead implantation in the heart via the coronary sinus. The ScoutPro accessories are used in conjunction with the ScoutPro CS Lead Introducer System to facilitate lead implantation in the left side of the heart via the coronary sinus.

Device Story

ScoutPro Hemostatic Valve is an accessory component of the ScoutPro 7F Lead Introducer System. Device facilitates lead implantation in the left side of the heart via the coronary sinus. Modification involves replacing a 3-way spigot with a 2-way spigot from a new supplier. Device is used by clinicians in a clinical setting during lead implantation procedures. No changes to functionality or intended use. Biocompatibility testing performed on modified valve. Output is a mechanical seal for the introducer system to maintain hemostasis during lead placement.

Clinical Evidence

Bench testing only; biocompatibility testing performed on modified hemostatic valves.

Technological Characteristics

Hemostatic valve component of a lead introducer system. Modification: 2-way spigot replaces 3-way spigot. Patient-contacting materials are identical to previously cleared devices. No electronic or software components.

Indications for Use

Indicated for patients requiring coronary sinus lead implantation in the left side of the heart.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ BIOTRONIK, Inc., ScoutPro Hemostatic Valve, Additional Information for Special 510(k) #K 101282 - June 18, 2010 # Modifications to theScoutPro Hemostatic Valve Special 510(k) Premarket Notification ## 1. 510(K) SUMMARY Name and Address of Applicant: BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 JUL 2 1 2010 Establishment Registration Number: 1028232 Device Name: Proprietary Name: Classification: Classification Name: Product Code: June 18, 2010 ScoutPro 7F Class II (21 CFR 870.1250; 870.1310; 870.1330) Wire, Guide, Catheters, Percutaneous DQY, DRE, DQX #### Date Prepared: ### General Description: The ScoutPro family of introducer systems and accessories is specifically used for the placement of coronary sinus leads. It is designed to assist with introducing leads into the vessels of the left side of the heart via the coronary sinus. The system also facilitates access to the coronary sinus venous system as well as probing the coronary sinus. This submission is in regard to a modification in a hemostatic valve, which is an accessory component of the ScoutPro family. The following ScoutPro 7F kit contains the hemostatic valve which is the subject of this Special 510(k): The ScoulPro 7F kit (with accessories) were cleared under 510(k) #K060807 on April 24, 2006: The basic set ScoutPro 7F contains the following components: - . 1 hemostatic valve (which is also separately available) - 2 guiding catheters "MPEP" and "BIO2" . - 1 dilator for the guiding catheter . - 1 peel-away sheath 10F with dilator . - . 1 quide wire - 1 needle . - . 1 syringe - 2 slitter tools 4.9 F and 6.3 F for different lead sizes . #### Device Modification: The hemostatic valve is included with the ScoutPro 7F kit and is also separately available. The hemostatic valve will be modified to include a 2-way spigot instead of a 3-way spigot, from a new supplier, Qosina. All patient contacting materials have already been used in other US legally marketed devices and the modified hemostatic valves have successfully undergone a full battery of biocompatibility testing. As a result of these modifications, updates will be made to the labeling. No changes have been made to the functionality of the hemostatic valve or any other accessories included within the kit. The change is being made to ensure continued supply of the hemostatic valves. {1}------------------------------------------------ #### BIOTRONIK, Inc., ScoutPro Hemostatic Valve, Additional Information for Special 510(k) #K101282 June 18, 2010 The usage of the ScoutPro hemostatic valve and ScoutPro 7F kit remains unchanged and the product characteristics such as indications for use, contraindications, and functions are identical to the previously cleared ScoutPro accessories in submission #K060807, cleared on April 24, 2006. Therefore, the previously cleared versions will serve as predicate devices for the modified products included in this Special 510(k). #### Predicate Device: BIOTRONIK's ScoutPro 7F (#K060807, 24-Apr-2006) #### Indications for Use: The ScoutPro CS Lead Introducer System facilitates lead implantation in the left of the heart via the coronary sinus. The ScoutPro accessories are used in conjunction with the ScoutPro CS Lead Introducer System to facilitate lead implantation in the left of the heart via the coronary sinus. #### Name and Address of Manufacturing Site: BIOTRONIK SE & Co. KG (reg. no. 9610139) Woermannkehre 1, 12359 Berlin, Germany 011-49-30-689-05-1210 #### Name and Address of Manufacturing Site: BIOTRONIK AG (reg. no. 8043892) Ackerstrasse 6 8180 Bülach, Switzerland 011-41-44-864-5169 #### 510(k) Contact Person and Phone Number: Jon Brumbaugh Vice President, Regulatory Affairs and Compliance BIOTRONIK, Inc. 6024 Jean Road Lake Oswego, OR 97035 Phone (888) 345-0374 Fax (503) 635-9936 jon.brumbauqh@biotronik.com . 12 {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of human services. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002 Biotronik, Inc. Mr. Jon Brumbaugh Vice President, Regulatory Affairs and Compliance 6024 Jean Road Lake Oswego, OR 97035 JUL 2 1 2010 Re: K101282 Trade/Device Name: ScoutPro Hemostatic Valve Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DQY Dated: June 18, 2010 Received: June 21, 2010 Dear Mr. Brumbaugh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. {3}------------------------------------------------ Page 2 - Mr. Jon Brumbaugh If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): TBD Device Name: ScoutPro Hemostatic Valve (Part of the ScoutPro 7F Kit) Indications for Use: The ScoutPro CS Lead Introducer System facilitates lead implantation in the heart via the coronary sinus. The ScoutPro accessories are used in conjunction with the ScoutPro CS Lead Introducer System to facilitate lead implantation in the left side of the heart via the coronary sinus. Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...