FuStarTM Steerable Introducer is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.
Device Story
FuStar Steerable Introducer functions as a guiding or introducer sheath to facilitate percutaneous access. Device features low-profile design, atraumatic tip, and manual deflection mechanism. Physician operates device to navigate vasculature; visible knob with scale on controller allows estimation of sheath deflecting angle (0 to 180 degrees). Curl options available in 30mm (small) and 50mm (large) radii. Connector options include luer lock or silicone valve. Device provides pathway for various catheters and instruments; assists in precise positioning within cardiac or peripheral vessels. Benefits include reduced risk of trauma during puncture and improved control during interventional procedures.
Clinical Evidence
Bench testing only. Performance testing included primary dimension, X-ray visibility, bending, fatigue (cycle), conical fitting, hub, connection strength, corrosion resistance, and leakage/aspiration tests. Biocompatibility testing included cytotoxicity, acute systemic toxicity, intracutaneous reactivity, hypersensitivity, hemolysis, complement system, thrombosis, pyrogen, and endotoxicity tests. Sterilization and package integrity were also validated.
Technological Characteristics
Steerable introducer sheath; deflectable tip (0-180 degrees); manual deflection via controller knob with scale. Available in 5F-10F diameters and 550mm-900mm lengths. Connector types: luer lock or silicone valve. Materials: unspecified medical-grade polymers/metals. Sterilization: validated. Connectivity: none (mechanical device).
Indications for Use
Indicated for patients requiring a pathway for device introduction into heart chambers, coronary vasculature, or peripheral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{1}------------------------------------------------
Premarket Notification Section 510(k) Submission
FuStar Steerable Introducers
Exhibit #5 510(k) Summary
Ref No.: A2009-018-075
Image /page/0/Picture/1 description: The image shows the logo for SesrCare Lifetech Scientific. The logo is in black and white, with the word "SesrCare" in a bold, sans-serif font. The letters are connected, giving the logo a sense of unity. The letters "TM" are in superscript to the right of the word "SesrCare". Below the word "SesrCare" is the phrase "Lifetech Scientific" in a smaller, sans-serif font.
1
FEB 1 4 2011
## Exhibit #6 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## The 510(k) Number is K100184
| Date of Submission: | 02 SEPT 2010 |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Lifetech Scientific (Shenzhen) Co., Ltd<br>Floor 1-3, Cybio Electronic Building, Langshan 2nd Street, North area of High-tech Park,<br>Nanshan District, Shenzhen, Guangdong, 518057, P. R. China<br>Contact Person:<br>Ms. Xiaoli Shi, International Regulatory / Mr. David Zhang, CTO<br>TEL:86-13418601356 FAX:(86-755)86026251<br>Email:stepbystep0326@hotmail.com |
| Correspondent: | Ms. Diana Hong / Mr. Lee Fu<br>Shanghai Mid-Link Business Consulting Co., Ltd<br>Suite 5D, No.19, Lane 999,Zhongshan Road (S-2),Shanghai, 200030, China<br>TEL: +86-21-22815850 FAX: +86-21-22815850 |
| Proposed Device | FuStar™ Steerable Introducers |
| Classification: | Class II<br>Product Code: DQY<br>Classification Name: catheter, percutaneous<br>Regulation Number: 21 CFR 870.1250<br>Review Panel: Cardiovascular |
| Predicate Device: | Predicate Device 1:<br>510(k) Number: K072313<br>Trade Name: AMPLATZER Torqvue Delivery System<br>Manufacturer: AGA Medical Corporation<br><br>Predicate Device 2:<br>510(k) Number: K061119<br>Trade Name: Enpath Medical Steerable Sheath |
{2}------------------------------------------------
Premarket Notification Section 510(k) Submission FuStar Steerable Introducers Exhibit #5 510(k) Summary Ref No .: A2009-018-075
Image /page/1/Picture/1 description: The image shows the logo for "SesrCare Lifetech Scientific". The word "SesrCare" is written in a bold, sans-serif font, with the letters connected. The letters are black. The words "Lifetech Scientific" are written in a smaller, thinner font below "SesrCare".
Intended Use: FuStarTM Steerable Introducer is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.
Device Description: The FuStarTM Steerable Introducers are designed to perform as a guiding sheath and/or introducer sheath, which provides a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature. Its low-profile design and atraumatic tip reduce the potential risk for trauma during percutaneous punctures while facilitating passage of a wide variety of catheters and devices.
> The proposed introducer can give physicians good control during operation, the curl options are available at 30mm (Small) and 50mm (Large). In each option, there are 2 connector types which are luer lock connector and silicone valve connector, and several specifications combination. Also, the visible knob with scale of controller facilitates the physicians to estimate the deflecting angle of the sheath while intra-operation.
> The products are available in different effective length (550mm, 700mm and 900mm) and diameter (French size, 5F, 6F, 7F, 8F, 9F and 10F). The deflectable angle of sheath is from 0 degree to 180 degree.
Testing Conclusion: Performance testing was conducted to validate and verify that the proposed device met all design specifications including:
> Performance Test: Primary Dimension Test; Exterior Surface Condition X-Ray Visible Testing; Bending Test, Cycle (Fatigue) Test: Conical Fitting Test: Hub Test; Connection Strength Test: Corrosion Resistance test, Leakage and Aspiration Test,
Biocompatibility Test: Cytotoxicity, Acute Systemic Toxicity, Intracutaneous Reactivity Test, Hypersensitivity Test, Hemolysis Test, Complement System Test. Thrombosis Test Sterilization: Sterliization Validation, Package Integrity Test, Pyrogen Test and Endotoxicity Test.
- SE Conclusion: FuStar™ Steerable Introducers, can be claimed to be Substantially Equivalent (SE) to the predicate device, Predicate Device AMPLATZER Torqvue Delivery System K072313 and Enpath Medical Steerable Sheath, K061119.
2
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Lifetech Scientific (Shenzhen) Co., Ltd C/O Ms. Diana Hong Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D. No.19, Lane 999, Zhongshan Road (S-2), Shanghai, 200030, China
FEB 1 4 201
K100184 Trade/Device Name: FuStar™ Steerable Introducers Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DOY Dated: January 18, 2011 Received: January 19, 2011
Dear Ms. Hong:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
1:00 PM IST STATE
{4}------------------------------------------------
Page 2 - Ms. Diana Hong
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?rgblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Dunn P. Viener
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Premarket Notification Section 510(k) Submission FuStar Steerable Introducers Section II Indication for Use Statement · Ref No.: A2009-018-075
Image /page/4/Picture/1 description: The image shows the logo for "SesrCare Lifetech Scientific". The word "SesrCare" is in a bold, sans-serif font, with the "TM" symbol in superscript next to it. Below "SesrCare" is the text "Lifetech Scientific" in a smaller, lighter font. The logo appears to be for a company involved in life sciences or scientific research.
Section II Indication for Use Statement
510(k) Number: K106184
Device Name: FuStar
TM
Steerable Introducers
Indications for Use:
FuStar™ Steerable Introducer is indicated to provide a pathway through which devices are introduced within the chambers and coronary vasculature of the heart or in the peripheral vasculature.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
humph R. Vebus
sion Sign-Off) (Division Sign-Oll)
Division of Cardiovascular Devices
510(k) Number k100184
II-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.