← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K091345

# GOPHER GOLD CATHETER, MODEL 5620 (K091345)

_Vascular Solutions, Inc. · DQY · Jun 12, 2009 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K091345

## Device Facts

- **Applicant:** Vascular Solutions, Inc.
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Jun 12, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Gopher Gold catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents.

## Device Story

Gopher Gold catheter is a 3F single-lumen percutaneous catheter; 135 cm working length; features positioning marks at 95 cm and 105 cm. Compatible with 0.014" guidewires and 6F guide catheters. Used by physicians in clinical settings to access coronary/peripheral vasculature; facilitates guidewire/device placement and exchange; enables subselective delivery of diagnostic/therapeutic agents. Device is single-use and sterile.

## Clinical Evidence

Bench testing only; no clinical data.

## Technological Characteristics

3F single-lumen catheter; 135 cm length; 0.014" guidewire compatibility; 6F guide catheter compatibility; ethylene oxide sterilization; single-use.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Gopher catheter ([K070372](/device/K070372.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Vascular Solutions, Inc.

Special 510(k) Premarket Notification Gopher Gold Catheter

K691345

#### 510(k) Summary 2

Date Prepared: April 15, 2009

### Submitter's Name / Contact Person

#### Manufacturer

Vascular Solutions, Inc. 6464 Sycamore Court Minneapolis, MN 55369 USA Establishment Registration # 2134812

### Contact Person

Loucinda Bjorklund Senior Regulatory Affairs Associate Tcl: 763.656.4208 (direct); Fax: 763.656.4253 Email: 1bjorklund@vascularsolutions.com

### General Information

Gopher™ Gold catheter Trade Name Common / Usual Name Classification Name Predicate Device

Percutaneous catheter 870.1250 Catheter, percutaneous K070372 Gopher catheter (Vascular Solutions, Inc.)

#### Device Description

The Gopher Gold catheter is a 3F catheter that is compatible with standard 0.014" guidewires and 6F guide catheters. The Gopher Gold is a single-lumen catheter with a working length of 135 cm with printed positioning marks at 95 cm and 105 cm. The Gopher Gold catheter is packaged in a dispenser coil and is sealed within a single sterile barrier pouch. The Gopher Gold catheter is sterilized by ethylene oxide and intended for single use.

#### Intended Use / Indications

The Gopher Gold catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents

#### Substantial Equivalence and Summary of Studies

The Gopher gold catheter is substantially equivalent in intended use and indications to the predicate device. Technological differences in design and materials have been qualified through biomaterial assessments and other design verification testing, the results of which did not raise any new safety or performance questions.

JUN 1 2 2009

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image contains the text "Public Health Service". The text is written in a simple, sans-serif font and is horizontally aligned. The words are stacked on a single line and are easily readable.

Image /page/1/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design resembling an eagle or bird in flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.

2 2009 JUL

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Vascular Solutions, Inc. c/o Ms. Loucinda Bjorklund Senior Regulatory Affairs Associate 6464 Sycamore Court Minneapolis. MN 55369

Re: K091345

> Trade/Device Name: Gopher™ Gold Catheter Common Name: Catheter, Percutaneous Regulation Number: 21 CFR 870.1250 Regulatory Class: II Product Code: DOY Dated: May 5, 2009 Received: May 13, 2009

Dear Ms. Bjorklund:

This letter corrects our substantially equivalent letter of June 12, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{2}------------------------------------------------

## Page 2 - Ms. Loucinda Bjorklund

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Duna R. Lechner

Image /page/2/Picture/6 description: The image shows the name and title of Bram D. Zuckerman, M.D. He is the Director of the Division of Cardiovascular Devices. The division is part of the Office of Device Evaluation. The office is part of the Center for Devices and Radiological Health.

Enclosure

{3}------------------------------------------------

# Indications for Use

K091345 510(k) Number (if known): Device Name: Gopher™ Gold catheter Indications for Use: The Gopher Gold catheter is intended to be used in conjunction with steerable guidewires to access discreet regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents. Prescription Use __ Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 (Posted November 13, 2003)

iovascular Devices

510(k) Number K0911345

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K091345](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K091345)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
