K082579 · Ev3, Inc. · DQY · Nov 20, 2008 · Cardiovascular
Device Facts
Record ID
K082579
Device Name
EVERCROSS .035 OTW PTA DILATATION CATHETER
Applicant
Ev3, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Nov 20, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EverCross .035" OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
Device Story
EverCross is an over-the-wire (OTW) 0.035" dual-lumen PTA dilatation catheter; features distally mounted semi-compliant inflatable balloon and atraumatic tapered tip for crossing tight stenoses. Distal catheter proximal to balloon includes hydrophilic coating. Manifold contains two lumens: 'THRU' lumen for guidewire passage (max 0.035" diameter); 'BALLOON' lumen for inflation/deflation using contrast/saline mixture. Two radiopaque markers facilitate balloon positioning relative to stenosis. Used by physicians in clinical settings for peripheral vascular interventions. Device enables mechanical dilation of arterial stenoses and dialysis fistulae, and post-stent expansion; benefits patients by restoring blood flow and maintaining vascular access patency.
Clinical Evidence
Bench testing and biocompatibility testing only. No clinical data provided.
Technological Characteristics
Dual-lumen OTW catheter; 0.035" guidewire compatibility; semi-compliant inflatable balloon; hydrophilic coating on distal shaft; radiopaque markers for positioning. Materials and design specifications are consistent with standard PTA dilatation catheters.
Indications for Use
Indicated for patients requiring dilation of stenoses in iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries; treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae; and stent post-dilatation in peripheral vasculature.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K090509 — FOXCROSS .035 PTA CATHETER 14 MM X 20MM X 80 CM, MODEL 10342-20, FOXCROSS .035 PTA CATHETER · Abbott Vascular-Vascular Solutions · Mar 20, 2009
K220410 — Ebony HP PTA OTW 0.035 Catheter · Natec Medical , Ltd. · Jun 29, 2022
K112697 — CHARGER PTA BALLON DILATATION CATHETER · Boston Scientific · Nov 30, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for EV3, a robotics kit. Above the logo are some handwritten characters that are difficult to decipher. The EV3 logo consists of three black circles arranged in a triangular formation, with the letters "ev3" in a stylized font below the circles. The "3" has a registered trademark symbol next to it.
NOV 20 2008
# 510(k) Summary
## EverCross™ .035" OTW PTA Dilatation Catheter
| 510(k) Summary | This summary of 510(k) safety and effectiveness information is being<br>submitted in accordance with the requirements of 21 C.F.R § 807.92. |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant | ev3 Inc. |
| Submitter | cv3 Inc.<br>9600 54th Ave. N<br>Plymouth, MN 55442<br>Tel: 763-398-7000<br>Fax: 763-398-7200 |
| Contact Person | David Worrell, MS, RAC<br>Director, Regulatory Affairs |
| Date Prepared | September 3rd, 2008 |
| Device Trade Name | EverCrossTM .035" OTW PTA Dilatation Catheter |
| Device Common Name | PTA Dilatation Catheter |
| Classification Name | Catheter, Angioplasty, Peripheral, Transluminal (21 CFR 800.1250,<br>Product Code LIT |
| Classification Panel | Cardiovascular |
| Predicate Devices | Sterling OTW PTA Balloon Dilatation Catheter (K053116), Admiral<br>Xtreme PTA Balloon Catheter OTW .035 (K062809), Savvy Long and<br>Sleek PTA Catheter (K072947), and Dorado PTA Balloon Dilatation<br>Catheter (K072283). |
| Intended use | The EverCross .035" OTW PTA Dilatation Catheter is intended to<br>dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive<br>lesions of native or synthetic arteriovenous dialysis fistulae. This<br>device is also indicated for stent post-dilatation in the peripheral<br>vasculature. |
| Device Description | The EverCross Peripheral Dilatation Catheter is an over the wire<br>(OTW) 0.035" dual lumen catheter with a distally mounted semi-<br>compliant inflatable balloon and an atraumatic, tapered tip to aid in<br>crossing tight stenoses. The distal catheter, proximal to the balloon, is<br>covered with a hydrophilic coating. The catheter manifold includes<br>two lumens. The lumen marked "THRU" is the central lumen of the |
| | catheter which terminates at the distal tip. This lumen is used to pass<br>the catheter over a guidewire with a maximum outer diameter of 0.035<br>inches. The lumen, marked "BALLOON" is used to inflate and<br>deflate the dilatation balloon with a solution of contrast medium and<br>saline. The balloon has two radiopaque markers for positioning the<br>balloon relative to the stenosis. |
| Performance data | Bench testing and biocompatibility testing were performed to support<br>a determination of substantial equivalence. Results from this testing<br>provide assurance that the proposed device has been designed and<br>tested to assure conformance to the requirements for its intended use. |
| Summary of Substantial<br>Equivalence | The EverCross .035" OTW PTA Dilatation Catheter has the following<br>similarities to the predicate devices:<br>• Similar fundamental scientific technology (all predicates)<br>• Similar operating principle (all predicates)<br>• Similar balloon lengths (Savvy Long and Sleek K072947)<br>• Similar rated burst pressures (Dorado K072283). |
| Conclusion | Based on the similar indications for use, technological characteristics<br>and performance testing, ev3 believes the EverCross .035" OTW PTA<br>Dilatation Catheter is substantially equivalent to the Sterling OTW<br>PTA Balloon Dilatation Catheter (K053116), the Admiral Xtreme<br>PTA Balloon Catheter (K062809), the Savvy Long and Sleek PTA<br>Catheters (K072947) and the Dorado PTA Balloon Dilatation Catheter<br>(K072283). |
ev3 Corporate World Headquarters / Peripheral Vascular ● 9600 54th Avenue North ● Plymouth, MN 55442 ● PH +1 763 398 7000 ev3 Europe SAS • ¡06/108 rue La Boetie • 75008 Paris, France • PH +33 156 88 59 10 Page 1 of 2 ev3 Neurovascular • 9775 Toledo Way • Irvine, CA 92618 • PH +1 949 837 3700 FoxHollow Technologies ● 740 Bay Road ● Redwood City, CA 94063 ● PH +1 650 421 8400
{1}------------------------------------------------
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle with its wings spread, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is rendered in black, and the text is also in black against a white background.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### NOV 2 0 2008
ev3 Inc. c/o Mr. David Worrell 4600 Nathan Lane North Plymouth, MN 55442-2920
Re: K082579
EverCross 0.035" OTW PTA Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (2) Product Code: DQY Dated: November 12, 2008 Received: November 13, 2008
Dear Mr. Worrell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to dcvices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
### Pagc 2 - Mr. David Worrell
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Fedcral statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) promarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, plcase contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
una R. Volmer
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page 1 of 1
#### Indications for Use Statement
K082579 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: EverCross™ .035" OTW PTA Dilatation Catheter
Indications for Use:
The EverCross 0.035" OTW PTA Dilatation Catheter is intended to dilate stenoses in the iliac, femoral, ilio-femoral, popliteal, infra-popliteal, and renal arteries, and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This device is also indicated for stent post-dilatation in the peripheral vasculature.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. 2) Thell remmer
(Division Sign-Off) Division of Cardiovascular Devices 510(k) Number Ko82579
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.