K070970 · Penumbra, Inc. · DQY · Aug 17, 2007 · Cardiovascular
Device Facts
Record ID
K070970
Device Name
NEURON SUPPORT CATHETER SYSTEM, MODEL 5F/6F
Applicant
Penumbra, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 17, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Neuron Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Device Story
Neuron Intracranial Access System consists of Neuron Delivery Catheter (outer, 6F proximal/5F distal) and Neuron Select Catheter (inner, 3.5F). Used by physicians in clinical settings to facilitate introduction of interventional devices into peripheral, coronary, and neuro vasculature. Delivery catheter functions with guidewire or inner select catheter. System provides access pathway for interventional procedures; aids in navigation to target vasculature. Benefits include improved access to complex vascular anatomy.
Clinical Evidence
Bench, in vitro, and in vivo testing performed on materials, components, subassemblies, and final assemblies. Devices met all specifications. No clinical trial data provided.
Technological Characteristics
Percutaneous catheter system. Materials are medical grade, biocompatible, and industry-standard. System includes outer delivery catheter (6F/5F) and inner select catheter (3.5F).
Indications for Use
Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. No specific patient population, age, or contraindications provided.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K142321 — Benchmark Intracranial Access System · Penumbra, Inc. · Jan 26, 2015
K213390 — Benchmark BMX96 Access System · Penumbra, Inc. · Aug 18, 2022
K083125 — NEURON SELECT CATHETER SIZE 070 · Penumbra, Inc. · Nov 21, 2008
K180201 — Route 92 Medical Access System · Route 92 Medical, Inc. · May 23, 2018
K212838 — Benchmark Intracranial Access System · Penumbra, Inc. · May 27, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
K070970 p 1 of 2
Image /page/0/Picture/1 description: The image shows the logo for Penumbra, Inc., a company that specializes in stroke intervention. The logo features the company name in a simple, sans-serif font, with a curved line above the name. The words "stroke intervention" are written in a smaller font below the company name. The logo is simple and modern, and it effectively communicates the company's focus on stroke intervention.
AUG 17 2007
## 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92(c)
### 510(k) Number
| Date Summary Prepared | April 4, 2007 |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Neuron™ Intracranial Access System |
| Common Name | Percutaneous Catheter |
| Classification Name | Percutaneous Catheter<br>(21 CFR Part 870.1250; Product Code DQY) |
| Submitted By | Penumbra, Inc.<br>2401 Merced Street, Suite 200<br>San Leandro, CA 94577 |
| Contact | Theresa Brander-Allen<br>VP of Regulatory and Quality<br>Tel: 510-618-3223<br>Fax: 510-352-1766<br>tballen@penumbrainc.com |
#### Predicate Devices
- Penumbra Balloon Guide Catheter (K053491), manufactured by Penumbra, Inc.
- Guider Softip Guiding Catheter XF 5F (K010853), manufactured by Boston Scientific Corporation
### Device Description
The Neuron Intracranial Access System has a Neuron Delivery Catheter (outer catheter), which is available with a diameter of 6F proximal and 5F distal, and a Neuron Select Catheter (inner catheter), which measures 3.5F and has multiple tip configurations. The Neuron Delivery Cather may be used with a guidewire or with the Neuron Select Catheters.
Materials used in the Neuron Intracranial Access System devices are manufactured from medical grade materials that are commonly used in the industry, are similar or identical to the predicate devices, and have historically been demonstrated to be both biocompatible and suitable for this use.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the text "K070970" and "picefz". The text appears to be handwritten and is slightly faded. The text is arranged horizontally, with "K070970" on the left and "picefz" on the right.
Image /page/1/Picture/1 description: The image shows the logo for Penumbra, a stroke intervention company. The logo features the word "Penumbra" in a bold, sans-serif font, with the words "stroke intervention" in a smaller font below it. Above the word "Penumbra" is a curved line. The logo is simple and modern, and it effectively communicates the company's focus on stroke intervention.
#### Intended Use
The Neuron Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. The indication statement of the Neuron Intracranial Access System is substantially equivalent to the legally marketed predicate devices.
#### Substantial Equivalence
The intended use, method of operation, methods of construction and materials used, are either identical or substantially equivalent to the existing, legally marketed, predicate devices. Therefore, Penumbra believes the Neuron Intracranial Access System is substantially equivalent to the predicate devices.
#### Testing
Bench testing, in vitro testing, and in vivo testing have been performed on the device materials, components, subassemblies, and final assemblies. The devices tested acceptably met the specifications.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1 7 2007
Penumbra, Inc. c\o Ms. Theresa Brandner-Allen VP of Regulatory and Quality 2401 Merced St, Suite 200 San Leandro, CA 94577
Re: K070970
Trade/Device Name: Neuron™ Intracranial Access System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dated: August 10, 2007 Received: August 13, 2007
Dear Ms. Brandner-Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Brandner-Allen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
QML
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Penumbra, a stroke intervention company. The logo features the word "Penumbra" in a bold, sans-serif font, with the words "stroke intervention" in a smaller font underneath. Above the word "Penumbra" is a curved line. The logo is simple and modern, and it effectively communicates the company's focus on stroke intervention.
# Indications for Use
| 510(k) Number (if known): | K 070970 |
|---------------------------|----------|
|---------------------------|----------|
Device Name:
Neuron™ Intracranial Access System
Indications for Use:
The Neuron™ Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Prescription Use _X (Per 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
QQ Mall
sion Silah-Off ision of Cardiovascular Devices
Page of
K070970 510(k) Number
Neuron™ Intracranial Access System Traditional 510(k) 04Aprr07
CONFIDENTIAL
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.