← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K060927

# CERES PTA CATHETER MODELS PTA018,PTA035 VARIOUS SIZES (K060927)

_Ceres Medical Systems, LLC · DQY · Oct 6, 2006 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K060927

## Device Facts

- **Applicant:** Ceres Medical Systems, LLC
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Oct 6, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Ceres PTA Catheter is intended for dilatation of lesions in the femoral, renal, iliac, popliteal, peroneal, and profunda arteries and native or synthetic arteriovenous dialysis fistulae.

## Device Story

Ceres PTA Catheter is a single-use, sterile balloon dilation catheter for angioplasty. Device features double lumen coaxial shaft with inflatable balloon at distal end. Proximal end includes double hub connector; first port attaches to inflation device for balloon inflation/deflation via outer lumen; second port allows guidewire advancement through distally open inner lumen. Platinum radiopaque bands facilitate balloon positioning under fluoroscopy. Used by physicians in clinical settings for percutaneous transluminal angioplasty. Device enables mechanical dilation of arterial lesions and dialysis fistulae to improve blood flow.

## Clinical Evidence

Non-clinical testing only. Testing confirmed safe and effective performance and demonstrated biocompatibility of the device.

## Technological Characteristics

Single-use, sterile balloon dilation catheter. Features double lumen coaxial shaft, inflatable balloon, and double hub connector. Includes platinum radiopaque bands for fluoroscopic visualization. Mechanical device; no software or electronic components.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- JOMED AG Fox Plus PTA Catheter ([K020854](/device/K020854.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## Appendix 2

0C1 - 6 2006

Page 15

#### 5. 510(k) Summary (Revised)

| 510(k) Summary           | This summary of 510(k) safety and effectiveness information is being<br>submitted in accordance with the requirements of 21 C.F.R 807.92.                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter                | Ceres Medical Systems, L.L.C.<br>4665 Sweetwater Blvd., Suite 104<br>Sugar, Land, TX 77479                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Contact Person           | Corrine Bonfiglio, RAC<br>Tel: (858) 481-1638<br>Fax: (858) 481-2363<br>Mobile: (858) 342-0344<br>E-mail: cbonfiglio@meister.net                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Date Prepared            | September 26, 2006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Trade Name               | Ceres PTA Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name              | Percutaneous Transluminal Angioplasty (PTA) Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Name      | Percutaneous Catheter (21 CFR 870.1250, Product Code DQY)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Predicate Device         | The Ceres PTA Catheter manufactured by JOMED AG which was cleared<br>for market entry on April 11, 2002 under 510(k) #K020854.                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Description              | The Ceres PTA catheter is a single-use, sterile balloon dilation catheter for<br>angioplasty. A double lumen coaxial shaft features an inflatable balloon at<br>its distal end. The proximal end provides a double hub connector, whose<br>first port allows the attachment of an inflatable device which is used to<br>inflate and deflate the balloon through the outer lumen. A second hub port<br>allows the advance of a guidewire throughout the entire length of the<br>catheter using the distally open inner lumen. Platinum radiopaque bands<br>are used to locate the balloon under fluoroscopy. |
| Indications for Use      | The Fox Plus PTA Catheter is intended for dilatation of lesions in the<br>femoral, renal, iliac, popliteal, peroneal, and profunda arteries and native<br>or synthetic arteriovenous dialysis fistulae.                                                                                                                                                                                                                                                                                                                                                                                                     |
| Substantial Equivalence  | The Ceres PTA catheter is substantially equivalent to the Fox Plus PTA<br>catheter because the indications for use are identical and the function,<br>design and performance specifications are comparable.                                                                                                                                                                                                                                                                                                                                                                                                 |
| Non-clinical Performance | Non-clinical testing was conducted to confirm the safe and effective<br>performance of the Ceres PTA catheter. Non-clinical testing also<br>demonstrated the biocompatibility of the subject device.                                                                                                                                                                                                                                                                                                                                                                                                        |
| Conclusion               | The Ceres PTA catheter is substantially equivalent to the currently<br>marketed predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |

{1}------------------------------------------------

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

QC7 - 6 2006

Ceres Medical Systems L.L.C. c/o Ms. Corrine M. Bonfiglio 13195 Seagrove Street San Diego, CA 92130

Re: K060927

Ceres PTA Catheters Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (Two) Product Code: DQY Dated: September 26, 2006 Received: September 27, 2006

Dear Ms. Bonfiglio:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

## Page 2 - Ms. Corrine M. Bonfiglio

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

ua R. buchner

Image /page/2/Picture/5 description: The image contains a handwritten symbol or character. It appears to be a stylized or abstract representation, possibly resembling a letter or a signature. The symbol is composed of curved and angular lines, with a bold stroke at the bottom.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Premarket Notification for Ceres PTA Catheter Response to Requested Information September 26, 2006

### Appendix 1

#### Indications for Use Statement (Revised) 4.

510(k) Number (if known): K060927

Device Name: Ceres PTA Catheter

Indications for Use:

The Ceres PTA Catheter is intended for dilatation of lesions in the femoral, renal, iliac, popliteal, peroneal, and profunda arteries and native or synthetic arteriovenous dialysis fistulae.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

and the country of the

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

Duana P. de lunes

ியாவி இடையி Joseph . . Carolovascular Devices

510(k) Number K060927

Page 1 of 1

(Posted November 13, 2003)

Ceres Medical Systems L.L.C. considers this application to be confidential and not subject to disclosure without the express written consent of the applicant.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K060927](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K060927)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
