← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K042568

# EXCELSIOR 1018 AND SL-10 MICROCATHETERS (K042568)

_Boston Scientific Corp · DQY · Oct 15, 2004 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K042568

## Device Facts

- **Applicant:** Boston Scientific Corp
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** Oct 15, 2004
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Boston Scienfitic's Excelsior™ SL-10 Pre-Shaped Microcatheter and Excelsion™ 1018™ Pre-Shaped Microcatheter are intended to assist in the delivery of diagnostic agents, such as contrast media, and therapeutic agents, such as occlusion coils to the peripheral, coronary, and neuro vasculature.

## Device Story

Microcatheters designed for distal vasculature access; used with guiding catheters and steerable guidewires. Features graded shaft stiffness (flexible tip to semirigid proximal section) for tracking; hydrophilic coating reduces friction; radiopaque tip enables fluoroscopic visualization. Physician-operated in clinical settings. Pre-shaped tips facilitate navigation; secondary shaping allowed for anatomical adjustment. Output is the delivery of diagnostic/therapeutic agents to target vasculature. Benefits include improved ease of use and navigation compared to straight-tip predecessors.

## Clinical Evidence

Bench testing only. No clinical data provided. In-vitro testing included tip shape retention, appearance, dimensions, static rupture, tensile strength, and compatibility with guide catheters and coils.

## Technological Characteristics

Percutaneous catheter; hydrophilic-coated shaft; radiopaque tip. Graded stiffness profile. Compatible with GDC coils. Pre-shaped tip configuration with secondary shaping capability. Sterilized device.

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Excelsior 1018 Microcatheter ([K994155](/device/K994155.md))
- Excelsior SL-10 Microcatheter ([K013789](/device/K013789.md))

## Reference Devices

- Guglielmi Detachable Coils ([K962503](/device/K962503.md))
- Guglielmi Detachable Coils ([K001083](/device/K001083.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## OCT 1 5 2004

### 510(K) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Date prepared: September 20, 2004 Contact Person: Victor Ham Regulatory Affairs Project Manager Boston Scientific 47900 Bayside Parkway Fremont, CA 94538 Tel: (510) 440-7604

Device Name: Excelsior™ 1018™ Pre-Shaped Microcatheter, Class II
Excelsior™ SL-10 Pre-Shaped Microcatheter, Class II

Common name: Microcatheters Classification name: Percutaneous Catheters

Product Code: DQY

#### Device Description:

The Excelsior " 1018" Pre-Shaped Microcatheter and the Excelsior" SL-10 Pre-Shaped Microcatheter are a I he zateenton - 1010 - 170 bd the physician in accessing the distal vasculature when used with a guiding catheter and steerable guidewire. Graded shaft stiffness ranging from a highly flexible tip to a semirigid proximal section aids the physician in tracking over selectively placed guidewires without displacement of the wire. A luer fitting located on the cather hub is used for the attachment of accessories. A radiopague tip facilitates fluoroscopic visualization. The outer diameter is coated with a hydrophilic surface that reduces tip tachitatis nuoroscopic 75the catheters are packaged individually with a variety of Pre-Shaped tips.

#### Indications for Use:

Like the predicate devices, Boston Scienfitic's Excelsior™ SL-10 Pre-Shaped Microcatheter and Excelsion™ 1018™ Pre-Shaped Microcatheter are intended to assist in the delivery of diagnostic agents, Excelsion - 1010 - 176 bhappeatic agents, such as occlusion coils to the peripheral, coronary, and neuro vasculature.

### Predicate Device(s):

Excelsior " 1018" Microcatheter (510(k) # K994155 cleared August 3, 2000) Excelsior "" SL-10 Microcatheter (510 (k) # K013789 cleared December 6, 2001)

### Testing in Support of Substantial Equivalence Determination

The predicate devices which were originally cleared for straight steam-shapeable tips were modified to include Pre-Shaped tips. The modification is made for ease of use to the physicians. The modification involves a new tip shaping process and the necessary packaging modifications to protect the new tip shapes.

The results of in-vitro testing (tip shape retention, appearance of the tip, tip dimensions, static rupture, tensile I ho reading of the worling test, and coil, guide catheter compatibility) and biological safety testing support the claim of substantial equivalence.

{1}------------------------------------------------

# Technological Characteristics and Product Feature Comparison

Excelsior ™ SL-10 Microcatheter and the Excelsior ™ 1018 ™ Microcatheter are substantially equivalent to the predicate devices in terms of functionality, materials, method of operation, intended use, indications for use, and biological safety.

| Characteristic       | Results                                                                                                                                                                                                                                              |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shaft Materials      | Same as predicate* device                                                                                                                                                                                                                            |
| Shaft Design         | Same as predicate* device                                                                                                                                                                                                                            |
| Distal Shaft Length  | Same as predicate* device                                                                                                                                                                                                                            |
| Proximal ID / OD     | Same as predicate* device                                                                                                                                                                                                                            |
| Distal ID / OD       | Same as predicate* device                                                                                                                                                                                                                            |
| Tip Markers          | Same as predicate* device                                                                                                                                                                                                                            |
| Coating              | Same as predicate* device                                                                                                                                                                                                                            |
| Effective Length     | Same as predicate* device                                                                                                                                                                                                                            |
| **GDC™ Compatibility | Same as predicate* device                                                                                                                                                                                                                            |
| Tip Configuration    | Offered with Pre-Shaped Tips<br>with the option of secondary<br>shaping for proper adjustment to<br>the anatomy prior to use.<br>As compared to the predicate<br>which was cleared with straight<br>tip, steam shapeable by physicia<br>prior to use |

# Product Feature Comparison for the

** Boston Scientific's Guglielmi Detachable Coils, K962503 & K001083

{2}------------------------------------------------

|  | Product Feature Comparison for the |  |
|--|------------------------------------|--|
|  |                                    |  |

### Excelsior ™ SL-10 Pre-Shaped Microcatheter

| Characteristic         | Results                                                                                                                       |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Shaft Materials        | Same as predicate* device                                                                                                     |
| Shaft Design           | Same as predicate* device                                                                                                     |
| Distal Shaft Length    | Same as predicate* device                                                                                                     |
| Proximal ID / OD       | Same as predicate* device                                                                                                     |
| Distal ID / OD         | Same as predicate* device                                                                                                     |
| Tip Markers            | Same as predicate* device                                                                                                     |
| Coating                | Same as predicate* device                                                                                                     |
| Effective Length       | Same as predicate* device                                                                                                     |
| **GDC™ Compatibility** | Same as predicate* device                                                                                                     |
| Tip Configuration      | Offered with Pre-Shaped Tips<br>with the option of secondary<br>shaping for proper adjustment to<br>the anatomy prior to use. |
|                        | As compared to the predicate<br>which was cleared with straight<br>tip, steam shapeable by physician<br>prior to use          |

*Excelsior SL-10 Microcatheter cleared under K 013789 on December 6, 2001

**Boston Scientific's Guglielmi Detachable Coils, K962503 & K001083

### 510(K) Summary of Safety and Effectiveness

Results of the performance and biocompatibility testing, as presented in this Special 510(K), demonstrate that the Pre-Shaped devices are substantially equivalent to the respective predicate unmodified devices.

The subject catheters with Pre-Shaped tips are substantially equivalent to their respective predicate devices with respect to the following:

- functionality ●
- . intended use
- . indications for use
- . materials

{3}------------------------------------------------

- method of operation .
- biological safety .

Based on the above information provided in this submission, Boston Scientific's Excelsior " 1018" Pre-Shaped Microcatheter is substantially equivalent to Boston Scientific's Excelsior "" 1018"" Microcatheter; and Boston Scientific's Excelsior "" SL-10 Pre-Shaped Microcatheter is substantially equivalent to Boston Scientific's Excelsior™ SL-10 Microcatheter.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a bird or a stylized human figure with outstretched arms, composed of three curved lines.

Public Health Service

OCT 1 5 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Boston Scientific c/o Mr. Victor Ham Regulatory Affairs Project Manager 47900 Bayside Parkway Fremont, CA 94538

K042568 Re:

K042506
Trade/Device Name: Excelsior™ 1018™ and SL-10 Pre Shaped Microcatheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dated: September 20, 2004 Received: September 21, 2004

Dear Mr. Ham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{5}------------------------------------------------

Page 2 – Mr. Victor Ham

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic forth in the quality systellis (QS) regulation (21-542 of the Act); 21 CFF 100-1050.
product radiation control provisions (Sections 531-542 of the Action Section S product radiation control provisions (overloris (overlor states in your Section 510(k)
This letter will allow you to begin marketing your device of your device to a legal This letter will anow you to begin makemily of substantial equivalence of your device to a legally premarket notification. The FDA Intellig of substantial equice and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please
s and the may be and cases and as a the color set the reculption continen a If you desire specific advice for your de not and one on one note the regulation entitled, they may obtain contact the Office of Compliation at (217) 276-6801' (21CFR Part 807.97). You may obtain "Misbranding by reference to premance routlited. Set from the Division of Small other general information on your responsion.com its toll-free number (800) 638-2041 or
Manufacturers, International and Consumer Assistance at its toll-bir consin html Manufacturers, International and Ochosa.http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

ua R. Lochner

GA Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ K042568

Device Name:

Excelsior™ 1018" Pre-Shaped Microcatheter
Excelsior™ SL-10 Pre-Shaped Microcatheter

Indications For Use:

Boston Scienfitic's Excelsior™ SL-10 Pre-Shaped Microcatheter and Excelsion™ 1018" Pre-Shaped Boston Scientitle S L.celsion - SE-76 Tre Online Microcanns, such as contrast media, and Microcatifeler are intended to assist in the domery of charger, coronary, and neuro vasculature.

Prescription Use_ × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Duna R. Lochner

(Division Sign-Off) Division of Cardiovascular Devices

510(k) Number Ko42568

Page 1 of __ 1 ___

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K042568](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K042568)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
