K032869 · Cook, Inc. · DQY · Nov 19, 2003 · Cardiovascular
Device Facts
Record ID
K032869
Device Name
COOK CODA BALLOON CATHETER
Applicant
Cook, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Nov 19, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The COOK CODA™ BALLOON CATHETER is intended for temporary occlusion of large vessels, or to expand vascular prostheses.
Device Story
The Cook Coda Balloon Catheter is a 2-lumen percutaneous catheter used for temporary large vessel occlusion or vascular prosthesis expansion. The device features a distal lumen for tracking over a .035-inch guidewire and a secondary lumen for balloon inflation/deflation. Radiopaque bands at the proximal and distal ends of the balloon facilitate positioning under fluoroscopic guidance. The device is operated by a physician in a clinical setting. By inflating the balloon, the clinician achieves vessel occlusion or prosthesis expansion, potentially benefiting patients by facilitating vascular interventions. The device is supplied sterile in double pouches.
Clinical Evidence
Bench testing only. Testing assessed biocompatibility, mechanical properties, performance characteristics, and safety. The device met all pre-defined test criteria.
{0}------------------------------------------------
| 510(k) Summary | NOV 1 9 2003 |
|------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By: | Theodore Heise, PhD, RAC<br>Director of Regulatory Scientific Affairs<br>Cook Incorporated<br>750 N. Daniels Way<br>Bloomington, IN 47404<br>(812) 339-2235<br>September 12, 2003 |
| Names of Device: | |
COOK CODA™ BALLOON CATHETER Trade Name: Common/Usual Name: Balloon catheter Classification Name: Percutaneous catheter
# Intended Use:
The COOK CODA™ BALLOON CATHETER is intended for temporary occlusion of large vessels, or to expand vascular prostheses.
## Predicate Devices:
Predicate devices are the LDOB BALLOON CATHETER (K002286) manufactured by Cook Incorporated (vessel occlusion), and the ILIAC BALLOON CATHETER (K003495) manufactured by Guidant Corporation (prosthesis expansion).
### Device Description:
The COOK CODA™ BALLOON CATHETER is a 2-lumen catheter. The distal lumen extends the length of the catheter for use over a ,035 wire quide. The other lumen communicates with the balloon and is used to expand and deflate it. Radiopaque bands are placed at both the distal and proximal aspects of the balloon to assist with positioning of the device under fluoroscopy. The balloon will be provided in diameters ranging from 30 to 40 mm, and in lengths ranging from 100 to 120 cm. The device is packaged in sterile, sealed double pouches.
#### Substantial Equivalence:
The COOK CODA™ BALLOON CATHETER is comparable with respect to technical characteristics and intended use to predicate devices in terms of 510(k) substantial equivalence.
#### Discussion of Tests and Test Results:
The COOK CODA™ BALLOON CATHETER underwent testing to assess biocompatibility, mechanical properties, performance characteristics, and safety, The device met the test criteria, providing reasonable assurance of device performance for its intended use and supporting substantial equivalence.
#### Conclusions Drawn from Tests:
The COOK CODA™ BALLOON CATHETER met the test criteria, providing reasonable assurance of device performance for its intended use and supporting substantial equivalence.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, rendered in a simple, black line drawing.
NOV 1 9 2003
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Dr. Theodore Heise Director of Regulatory Scientific Affairs c/o Cook Incorporated P.O. Box 489 Bloomington, IN 47402-0489
Re: K032869 CODA Balloon Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: September 12, 2003 Received: September 15, 2003
Dear Dr. Heise:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 – Dr. Theodore Heise
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Arlile B. Boam
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known) __ K03 2869
COOK CODA BALLOON CATHETER
Indications for Use:
Device Name:
The COOK CODA BALLOON CATHETER is intended for temporary occlusion of large vessels, or to expand vascular prostheses.
ll
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ashley B. Boane
Division Sign-Off
Division of Cardiovascular Devices
510(k) Number k032869
Image /page/3/Picture/10 description: The image shows the text "Prescription Use" followed by the text "(Per 21 CFR 801.109)". There is a check mark above the line after the word "Use". The text is black and the background is white.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.