ILT SAFE-CROSS DEFLECTING CATHETER, MODEL C114ND1

K032784 · Intra Luminal Therapeutics, Inc. · DQY · Nov 18, 2003 · Cardiovascular

Device Facts

Record IDK032784
Device NameILT SAFE-CROSS DEFLECTING CATHETER, MODEL C114ND1
ApplicantIntra Luminal Therapeutics, Inc.
Product CodeDQY · Cardiovascular
Decision DateNov 18, 2003
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The Safe-Cross Deflecting Catheter is indicated to be used with a steerable guidewire in order to access discrete regions of the vasculature. Once the region has been accessed, an exchange of one guidewire for another can occur. The Safe-Cross Deflecting Catheter may also be used to provide a conduit for the delivery of saline solutions or diagnostic contrast agents.

Device Story

Coaxial lumen intravascular catheter; hydrophilic coating (Biocoat Hydak). Used with steerable guidewires to access remote vascular regions; facilitates guidewire exchange; provides conduit for saline or contrast agents. Operated by clinicians in vascular procedures. Device provides mechanical access and fluid delivery path; enhances procedural efficiency during guidewire placement. Single-use; sterile.

Clinical Evidence

Bench testing only. Biocompatibility testing confirmed material safety; performance testing confirmed functional requirements for vascular access and fluid delivery.

Technological Characteristics

Coaxial lumen intravascular catheter; Biocoat Hydak hydrophilic coating. Length 135-140 cm; shaft 2.5 F; deflecting segment 3.5 F at 10 atm. Accommodates 0.014" guidewires. Sterilized with ethylene oxide gas. Single-use.

Indications for Use

Indicated for use with a steerable guidewire to access discrete regions of the vasculature, facilitate guidewire exchange, and serve as a conduit for saline or diagnostic contrast agents.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K032784 NOV 1 8 2003 # 510(k) SUMMARY ### SUBMITTER INFORMATION | A. | Company Name: | IntraLuminal Therapeutics, Inc. | |----|---------------|---------------------------------| |----|---------------|---------------------------------| B. Company Address: C. Company Phone: Company Facsimile: D. - ய் Contact Person: nc 6354 Corte del Abeto Suite A Carlsbad, CA 92009 (760) 918-1820 (760) 918-1823 Pamela Misajon Vice President of Requlatory Affairs and Quality Assurance ## DEVICE IDENTIFICATION - A. Trade Name: B. Catalog Number: C114ND1 C. Support Catheter Common Name: D. Classification Name: Percutaneous Catheter - Device Class: E. ILT Safe-Cross® Deflecting Catheter Class II (per 21 CFR 870.1250) # IDENTIFICATION OF MODIFIED DEVICE The ILT Safe-Cross® Deflecting Catheter is identical in design, materials and mode of operation to the IntraLuminal Therapeutics, Inc. ILT Safe-Cross Deflecting Catheter cleared under 510(k) K031692. ## DEVICE DESCRIPTION The Safe-Cross Deflecting Catheter is a coaxial lumen intravascular catheter coated with a Biocoat Hydak® hydrophilic coating intended for percutaneous use. It is designed to be used in conjunction with guidewires to gain access to and facilitate placement of the wires in locations within the vascular system that are remote from the site of insertion. Once accessed, guidewires may be exchanged within the catheter. The Safe-Cross Deflecting Catheter may also be used to provide a conduit for the delivery of saline solutions or diagnostic contrast agents. {1}------------------------------------------------ The effective length of the Safe-Cross Deflecting Catheter is 135-140 cm. The shaft nominal outside diameter is 2.5 F (0.033") and the maximum outside diameter of the deflecting segment is 3.5 F (0.045") at a use pressure of 10 atm. The I.D. at the tip of the Safe-Cross Deflecting Catheter is a nominal 0.016" and the inside diameter will accommodate commercially available 0.014" guidewires. The Safe-Cross Deflecting Catheter is packaged in a Tyvek® covered custom PETG tray that is heat-sealed to form a sterile barrier. The packaged units are sterilized with ethylene oxide gas. The device is provided "STERILE". "Non-Pyrogenic" and is intended for single use only. #### INTENDED USE The Safe-Cross Deflecting Catheter is indicated to be used with a steerable guidewire in order to access discrete regions of the vasculature. Once the region has been accessed, an exchange of one guidewire for another can occur. The Safe-Cross Deflecting Catheter may also be used to provide a conduit for the delivery of saline solutions or diagnostic contrast agents. #### TECHNOLOGICAL CHARACTERISTICS The Safe-Cross Deflecting Catheter is identical in basic materials, design, construction and mechanical performance to the predicate device. #### BIOCOMPATIBILITY AND PERFORMANCE DATA Biocompatibility testing and bench studies were conducted to evaluate the biological and performance characteristics of the Safe-Cross Deflecting Catheter. Biocompatibility test results indicate that the device materials are biocompatible. Performance test results indicate that the device satisfies functional performance requirements when used as indicated. #### CONCLUSIONS DRAWN FROM STUDIES The results of testing demonstrate that the Safe-Cross Deflecting Catheter is substantially equivalent to the predicate devices and is capable of safely and accurately performing the stated intended use. {2}------------------------------------------------ Image /page/2/Picture/11 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling an emblem. The emblem is a stylized representation of a bird or eagle, with three curved lines forming its body and wings, symbolizing health and human services. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # NOV 1 8 2003 IntraLuminal Therapeutics, Inc. c/o Pamela Misajon Vice President of Regulatory Affairs and Quality Assurance 6354 Corte del Abeto, Suite A Carlsbad, Ca 92009 Re: K032784 > ILT Safe-Cross® Deflecting Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II Product Code: DQY Dated: October 17, 2003 Received: October 20, 2003 #### Dear Ms. Misajon: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 -- Ms. Pamala Misajon Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices ,Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### INDICATIONS FOR USE 510(k) Number: K032784 ILT Safe-Cross® Deflecting Catheter Device Name: The Safe-Cross® Deflecting Catheter is indicated to Indications for Use: be used in conjunction with a steerable guidewire in order to access discrete regions of the vasculature. Once the region has been accessed, an exchange of one guidewire for another can occur. The Safe-Cross® Deflecting Catheter may also be used to provide a conduit for the delivery of saline solutions or diagnostic contrast agents. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use OR Over-The-Counter Use (Per 21 CFR 801.109) Division of Cardiovascular & Respiratory Devices 510(k) Number K032784
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...