PTA AND PTV

K022722 · NuMED, Inc. · DQY · Sep 13, 2002 · Cardiovascular

Device Facts

Record IDK022722
Device NamePTA AND PTV
ApplicantNuMED, Inc.
Product CodeDQY · Cardiovascular
Decision DateSep 13, 2002
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Z-MED-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, illiac, and renal arteries. These catheters are not designed to be used in the coronary arteries. Tyshak-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries. Tyshak II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; Z-MED II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; Mullins-X PTA - This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries.

Device Story

Catheter series (Tyshak-X, Tyshak II-X, Z-MED-X, Z-MED II-X, Mullins-X) designed for percutaneous transluminal valvuloplasty (PTV) of pulmonary valve and percutaneous transluminal angioplasty (PTA) of peripheral arteries. Device consists of balloon catheter with modified inner tubing; tubing features multi-layer polyamide (Vestamid PA12) extrusion reinforced with 304 LV stainless steel braid. Radiopaque platinum/iridium bands facilitate visualization. Operated by physicians in clinical settings to dilate stenotic valves or vessels. Catheter tracks over 0.035" guidewire. Braided construction enhances trackability and prevents catheter body stretching during removal. Output is mechanical dilation of target anatomy. Benefits include improved blood flow through pulmonary valve or peripheral arteries.

Clinical Evidence

Bench testing only. Testing included balloon deflatability, inflation/deflation times (<20 seconds), maximum luminal injection pressure (400 psi), bond integrity (min 2 lbs), diameter/profile/introducer fit, and guidewire compatibility.

Technological Characteristics

Catheter with balloon; inner tubing composed of multi-layer polyamide (Vestamid PA12) with 304 LV stainless steel braid; radiopaque platinum/iridium image bands; 20% BaSO4 loading; Pantone 295C coloring. Compatible with 0.035" guidewire. Sterilization method not specified.

Indications for Use

Indicated for patients with isolated pulmonary stenosis or valvular pulmonary stenosis with minor congenital heart disease requiring PTV, and for PTA of femoral, iliac, and renal arteries. Contraindicated for use in coronary arteries.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # SUMMARY OF SAFETY AND EFFECTIVENESS DATA SEP 1 3 2002 August 14, 2002 Submitted By: NuMED, Inc., 2880 Main St., Hopkinton, NY 12965 (Ph) 315-328-4491 (F) 315-328-4941 Contact Person: Nichelle LaFlesh Device Name: Tyshak-X, Tyshak II-X, Z-MED-X, Z-MED II-X, Mullins-X Predicate Devices: Tyshak, Tyshak II, Z-MED, Z-MED II, Mullins The only material that has changed is the material in the inner Biocompatibility Testing: tubing. It has gone from being Pebax to Vestamid with stainless steel braiding. The testing that was done on this new material in enclosed in the biocompatibility section of this submission as well as the data for the other materials that have not changed. | Bench Testing Performed | Acceptance Criteria | |-----------------------------------------------|------------------------------------------------------------------------------------------------------------------| | 2. Balloon Deflatability | Deflation achieved in less than 20 seconds. | | 3. Inflation/ Deflation Times Test | Inflation and deflation achieved in less than 20<br>seconds. | | 5. Maximum Luminal Injection Pressure Test | Must withstand 400 psi. | | 6. Bond Integrity | Minimum of 2 lbs. at all test points. | | 7. Diameter/ Profile Test and Introducer Test | Must fit through rated introducer size. | | 9. Guidewire Compatibility | Minimal resistance while pushing catheter over<br>guidewire. Force may not be great enough to<br>bend guidewire. | Laboratory (Bench) Testing: # Comparison Information: # Change Description: The only change from the current catheter configuration to the new X line is the inner tubing configuration. The current inner tubing on all catheter models is composed of Pebax {1}------------------------------------------------ polyamide material. The tubing is sized to accept a 0.035" guidewire and provide adequate guidewire movement and catheter trackability. This tubing is also fitted with radiopaque image bands of platinum/iridium. This tubing is common to all catheters in NuMED's product lines. The proposed tubing is comprised of a multi layer extrusion of polyamide (Vestamid PA12) that surrounds a braid of 304 LV Stainless Steel. This tubing is loaded with 20% BaSO4 and is Colored blue with Pantone 295C. All materials have been certified as biocompatible. This tubing is designed to enhance guidewire movement and also to avoid any stretching of the catheter body during a tougher than normal catheter removal. ### Important Considerations: The proposed inner tubing is a nylon (polyamide) tubing that is very similar to the current material. There will be no significant material changes in the tubing or the image bands. The only change will be the construction of the inner tubing in changing from a polyamide tubing to a braided polyamide tubing. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle. SEP 1 3 2002 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Numed, Inc. c/o Ms. Nichelle R. LaFlesh Regulatory Affairs Manager P.O. Box 129 Nicholville, NY 12965 Re: K022722 > Trade Name: PTA and PTV Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous catheter Regulatory Class: Class II (two) Product Code: DQY Dated: August 15, 2002 Received: August 16, 2002 Dear Ms. LaFlesh: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 – Ms. Nichelle R. LaFlesh forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. Delguth Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications For Use 510(k) Number (if known): Device Names: Z-MED-X Catheter, Tyshak-X Catheter, Tyshak II-X Catheter, Z-MED II-X Catheter, and Mullins-X PTA Catheter Indications For Use: Z-MED-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis י - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not l require surgical intervention. And; This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, illiac, and renal arteries. These catheters are not designed to be used in the coronary arteries. Tyshak-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis ■ - 트 A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries. Tyshak II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis B - A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; Z-MED II-X Catheter - This catheter is recommended for Percutaneous Transluminal Valvuloplasty (PTV) of the pulmonary valve. - A patient with isolated pulmonary stenosis 트 - I A patient with valvular pulmonary stenosis with other minor congenital heart disease that does not require surgical intervention. And; Mullins-X PTA - This catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, and renal arteries. These catheters are not designed to be used in the coronary arteries. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter-Use<br>(Optional Format 1-2-96) | |------------------------------------------|----|--------------------------------------------------| |------------------------------------------|----|--------------------------------------------------| Division of Cardiovascular & Respiratory Devices | 510(k) Number | K022722 | |---------------|---------| |---------------|---------|
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...