← Product Code [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY) · K021587

# ATTAIN LDS 6216A LEFT-HEART DELIVERY SYSTEM, MODEL 6216A (K021587)

_Medtronic Vascular · DQY · May 30, 2002 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K021587

## Device Facts

- **Applicant:** Medtronic Vascular
- **Product Code:** [DQY](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY.md)
- **Decision Date:** May 30, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1250
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Attain™ LDS 6216A Left-heart delivery system is intended for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

## Device Story

The Attain LDS 6216A is a percutaneous delivery system used by clinicians to facilitate the placement of balloon catheters or cardiac leads into the coronary sinus and left-heart vessels. The system comprises guide wires, adjustable hemostasis valves to minimize blood loss, guide catheters, a guide catheter dilator, guide catheter slitters, and a guide wire clip. The distal tip of the guide catheter incorporates Tungsten Carbide-filled polyether block amide. The device is used in a clinical setting during implant procedures. It functions as a mechanical delivery tool to navigate the vasculature and provide access for therapeutic devices. The system is provided sterile for single use.

## Clinical Evidence

Bench testing only. In vitro performance testing and biocompatibility testing (per ISO 10993-1) were conducted to support substantial equivalence. No clinical data was required or provided.

## Technological Characteristics

Materials: Tungsten Carbide-filled polyether block amide (distal tip). Components: Guide wires, adjustable hemostasis valves, guide catheters, dilator, slitters, guide wire clip. Sterilization: 100% Ethylene Oxide (ETO). Connectivity: None (mechanical device).

## Regulatory Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

## Predicate Devices

- Medtronic Attain LDS 6216 Left-heart Delivery System ([K012130](/device/K012130.md))
- Medtronic Attain Access 6218 Left-heart Delivery System ([K012083](/device/K012083.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MAY 3 0 2002

CONFIDENTIAL May not be reproduced without written permission from Medtronic, Inc.

# 510(K) Summary of Substantial Equivalence

| Date prepared             | May 13, 2002                                                                                            |
|---------------------------|---------------------------------------------------------------------------------------------------------|
| Submitter:                | Medtronic, Inc.<br>Parkmore Business Park West<br>Galway, Ireland.                                      |
| Submission Correspondent: | Karen Reidt, RAC<br>Sr. Product Regulation Manager<br>7000 Central Avenue N.E.<br>Minneapolis, MN 55432 |
| Telephone:                | (763) 514-3914                                                                                          |
| Fax:                      | (763) 514-6424                                                                                          |
| E-Mail:                   | karen.reidt@medtronic.com                                                                               |
| Proprietary Name:         | Attain™ LDS 6216A Left-heart delivery<br>system                                                         |
| Common Name:              | Catheter, Percutaneous                                                                                  |
| Device Classification:    | Class II, 21 CFR § 870.1250                                                                             |
| Product Code:             | 74 DQY                                                                                                  |

## Performance Standard

Performance standards do not currently exist for these devices. None established under Section 514.

#### Device Description

The left-heart delivery system features two guide wires to facilitate venous access, adjustable hemostasis valves to reduce blood loss during the implant procedure, two guide catheters for passing venogram balloon catheters or appropriate leads, a guide catheter dilator to facilitate guide catheter passage, guide catheter slitters for removing guide catheters, and a guide wire clip to help contain the guide wire in the sterile field.

The Attain LDS 6216A Left-heart delivery system combines devices that are either cleared for market distribution via 510(k) or are exempt from premarket notification because of Class I designation.

The key changes being made to the system are the addition of Tungsten Carbidefilled polyether block amide material to the guide catheter distal tip segment and the addition of extra slitters and adjustable hemostasis valves in a revised system configuration.

{1}------------------------------------------------

#### CONFIDENTIAL

#### May not be reproduced without written permission from Medtronic, Inc.

#### Indications for Use

The intended use of the Medtronic Attain LDS 6216A Left-heart delivery system is for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus

### Substantially Equivalent Devices

| Attain LDS 6216A<br>Left-heart<br>Delivery System<br>Device | Predicate Device                                              | Predicate Device Manufacturer            | Predicate 510(k) |
|-------------------------------------------------------------|---------------------------------------------------------------|------------------------------------------|------------------|
| 40 cm and 45 cm<br>Guide Catheters                          | Medtronic Attain LDS 6216<br>Left-heart Delivery System       | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012130          |
| Guide Catheter<br>Dilator                                   | Medtronic Attain LDS 6216<br>Left-heart Delivery System       | Medtronic, Inc.<br>Minneapolis. MN 55432 | K012130          |
| Adjustable<br>Hemostasis Valve                              | Medtronic Attain LDS 6216<br>Left-heart Delivery System       | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012130          |
| 120 cm Guide<br>Wires                                       | Medtronic Attain Access<br>6218 Left-heart Delivery<br>System | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012083          |
| 4 Fr and 6 Fr<br>Guide Catheter<br>Slitters                 | Medtronic Attain LDS 6216<br>Left-heart Delivery System       | Medtronic, Inc.<br>Minneapolis, MN 55432 | K012130          |
| Guide Wire Clip                                             | Class I device, exempt from premarket notification            |                                          |                  |

Attain LDS 6216A Left-heart Delivery System Predicate Devices

### Summary of Studies

In vitro testing and biocompatibility/sterilization testing was performed to support substantial equivalence to the predicate device. The results of this testing indicates that the Attain LDS 6216A meets all of its design and performance requirements.

### Biocompatibility Information

Biocompatibility testing was performed on representative samples of Attain product. Complete testing according to ISO 10993-1 was conducted and all materials were found to be biocompatible.

## Sterilization Validation

The Attain LDS 6216A Left-heart delivery system is sterilized using a 100% Ethylene Oxide (ETO) sterilization process. Appropriate processes for sterilizing the devices were validated.

{2}------------------------------------------------

#### CONFIDENTIAL

May not be reproduced without written permission from Medtronic, Inc.

# Conclusion (Statement of Equivalence)

Through the data and information presented, as well as similarities to a legally Through the data and thronic, Inc. considers the Attain LDS 6216A Left-heart marketed dotho, medically equivalent to the previously discussed legally marketed predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a symbol consisting of three stylized human figures, depicted in a flowing, abstract manner. The figures are arranged in a row, with each figure slightly overlapping the previous one.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 3 0 2002

Ms. Karen Reidt Principal Regulatory Affairs Specialist Cardiac Rhythm Management Medtronic, Inc. 7000 Central Avenue N.E. Minneapolis, MN 55432

Re: K021587

Trade Name: Attain™ LDS 6216A Left-Heart Delivery System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DQY Dated: May 13, 2002 Received: May 15, 2002

#### Dear Ms. Reidt:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your betermined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Ms. Karen Reidt

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Da Carter

Donna-Bea Tillman, Ph.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### CONFIDENTIAL May not be reproduced without written permission from Medtronic, Inc.

### Indications for Use Statement

| 510(k) Number (if known): | N/A |
|---------------------------|-----|
|---------------------------|-----|

Device Name:

Attain™ LDS 6216A Left-heart delivery system

Indications For Use:

The Attain™ LDS 6216A Left-heart delivery system is intended for introducing balloon catheters into the coronary sinus or leads into vessels of the left heart via the coronary sinus.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use _ (Per 21 CFR 801.109)

OR

Over-The-Counter Use_

Division of Cardiovascular & Respiratory Devices
510(k) Number K021587

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K021587](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQY/K021587)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
