K014212 · C.R. Bard, Inc. · DQY · Jan 17, 2002 · Cardiovascular
Device Facts
Record ID
K014212
Device Name
BARD CONQUEST PTA BALLOON DILATATION CATHETER
Applicant
C.R. Bard, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Jan 17, 2002
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bard® Conquest™ PTA Balloon Dilatation Catheter is recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.
Device Story
The Bard® Conquest™ PTA Balloon Dilatation Catheter is a coaxial lumen catheter featuring a balloon at its distal tip. It is designed for use by physicians in clinical settings to perform percutaneous transluminal angioplasty. The device includes two radiopaque marker bands beneath the balloon to facilitate accurate positioning within the vasculature under imaging. One lumen is used for guidewire insertion, while the second lumen serves as a channel for balloon inflation and deflation. By inflating the balloon at the site of an obstructive lesion, the device dilates the vessel or fistula, thereby improving blood flow. The device is intended for peripheral vascular applications and is explicitly contraindicated for coronary artery use.
Clinical Evidence
Bench testing only.
Technological Characteristics
Coaxial lumen catheter; balloon mounted on distal tip; two radiopaque marker bands; shaft diameters of 6, 7, or 8 Fr; shaft lengths of 50, 75, or 100 cm. Mechanical device; no energy source or software.
Indications for Use
Indicated for patients requiring percutaneous transluminal angioplasty of femoral, iliac, or renal arteries, and treatment of obstructive lesions in native or synthetic arteriovenous dialysis fistulae. Contraindicated for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Opti-Plast Centurion™ 5.5 Fr PTA Catheter
Related Devices
K040243 — BARD ATLAS PTA BALLOON DILATATION CATHETER · C.R. Bard, Inc. · Feb 13, 2004
K073472 — VACCESS PTA BALLOON DILATATION CATHETER · C.R. Bard, Inc. · Jan 9, 2008
K113468 — ELM PTA BALLON DILATATION CATHETER · Creagh Medical, Ltd. · Dec 20, 2011
K955886 — MEGA PTA BALLOON CATHETER · Cordis Corp. · Mar 11, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Bard Peripheral Technologies C.R. Bard, Inc. 13183 Harland Dr., N.E. Covington, GA 30014
K014212
JAN 1 7 2002
# BARD
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENES INFORMATION
- A.. Submitter Information:
| Submitter's Name: | C.R. Bard, Inc., Peripheral Technologies Division |
|------------------------------------|---------------------------------------------------|
| Submitter's Address: | 13183 Harland Drive, Covington, GA 30014 |
| Contact Person: | Carol Vierling |
| Contact Person's Telephone Number: | (770) 385-2347 |
| Contact Person's FAX Number: | (770) 385-2340 |
| Date of Preparation: | December 14, 2001 |
- B. Device Name:
Bard® Conquest™ PTA Balloon Dilatation Catheter
- C. Predicate Device:
Opti-Plast Centurion™ 5.5 Fir PTA Catheter
- D. Device Description: ·
The Bard® Conquest™ PTA Balloon Dilatation Catheter is a coaxial lumen catheter with a balloon mounted on its distal tip. One lumen accommodates the insertion guidewire and the second provides a channel for inflation/deflation of the balloon. There are two radiopaque marker bands placed beneath the balloon to indicate its position within the vasculature.
{1}------------------------------------------------
#### E. Intended Use:
The Bard® Conquest™ PTA Balloon Dilatation Catheter is recommended for use in Percutaneous Transluminal Angioplasty of the femoral, iliac and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae. This catheter is not for use in coronary arteries.
- F. Technological Characteristics Summary:
The Bard® Conquest™ PTA Balloon Dilatation Catheter is offered in 6, 7 or 8 Fr shaft diameters and shaft lengths of 50, 75 or 100cm, depending on balloon size. Various balloon diameters and lengths are available.
- G. Performance Data:
Bench testing shows that the modified catheter is substantially equivalent to the predicate device, the Opti-Plast Centurion™ 5.5 Fr PTA Catheter.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three wavy lines representing the flow of life. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JAN 1 7 2002
Ms. Carol Vierling Director, Regulatory Affairs C.R. Bard, Inc. Bard Peripheral Technologies 13183 Harland Drive, N.E. Covington, GA 30014
Re: K014212
Trade Name: Bard® Conquest™ PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II (two) Product Code: DQY Dated: December 14, 2001 Received: December 21, 2001
Dear Ms. Vierling:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
#### Page 2 - Ms. Carol Vierling
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours.
Dallas Miller
ram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Device Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## ATTACHMENT 5
# Indications for Use Statement
| Device<br>Name | Bard® ConquestTM PTA Balloon Dilatation Catheter |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | |
| Indications<br>for Use | The Bard® ConquestTM PTA Balloon Dilatation Catheter<br>is recommended for use in Percutaneous Transluminal Angioplasty of the<br>femoral, iliac and renal arteries and for the treatment of obstructive lesions<br>of native or synthetic arteriovenous dialysis fistulae. This catheter is not<br>for use in coronary arteries. |
#### PLEASE DO NOT WRITE BELOW THIS LINE--- CONTINUE ON ANOTHER PAGE IF NEEDED. .
Concurrence of CDRH, Office of Device Evaluation (ODE)
N
Division of Cardiovascular & Respiratory Devices
510(k) Number 2617212
Prescription Use
(Per 21 CFR 801.109
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.