ILT SUPPORT CATHETER .014, MODEL C114LW1; ILT SUPPORT CATHETER ADVANCING MECHANISM, MODEL A11AM1
K012169 · Intra Luminal Therapeutics, Inc. · DQY · Aug 30, 2001 · Cardiovascular
Device Facts
Record ID
K012169
Device Name
ILT SUPPORT CATHETER .014, MODEL C114LW1; ILT SUPPORT CATHETER ADVANCING MECHANISM, MODEL A11AM1
Applicant
Intra Luminal Therapeutics, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Aug 30, 2001
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The ILT Support Catheter with or without the advancing mechanism is indicated to be used in conjunction with a steerable guidewire in order to access discreet regions of the vasculature. Once the region has been accessed, an exchange of one guidewire for another can occur. The ILT Support Catheter may also be used to provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
Device Story
Single-lumen intravascular catheter; used with 0.014" steerable guidewire to access remote cardiovascular regions; provides conduit for saline or diagnostic contrast agents. Includes optional advancing mechanism (handle) for incremental guidewire advancement. Used in clinical settings by physicians. Device facilitates guidewire exchange and fluid delivery; benefits patient by enabling access to difficult-to-reach vascular sites.
Clinical Evidence
Bench testing only. Biocompatibility testing and in vitro performance studies conducted to evaluate biological and functional characteristics. No clinical data presented.
Technological Characteristics
Single-lumen intravascular catheter; 135 cm effective length; 0.037" outside diameter; accommodates 0.014" guidewire. Optional handle with advancing mechanism. Sterilized by ethylene oxide gas. Single-use.
Indications for Use
Indicated for patients requiring access to discreet regions of the vasculature for guidewire exchange or delivery of saline/contrast agents, using a steerable guidewire.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
K031417 — ILT SAFE CROSS 0.035 SUPPORT CATHETER, MODEL C135SN1 · Intra Luminal Therapeutics, Inc. · Jul 11, 2003
K991059 — SPECTRANETICS 0.014 SUPPORT CATHETER, MODEL 518-016, SPECTRANETICS 0.018 SUPPORT CATHETER, MODEL 518-017 · Spectranetics Corp. · Nov 16, 1999
K214107 — XO CROSS CORONARY Support Catheter · Transit Scientific, LLC · Apr 11, 2022
K170996 — Wingman 14C Crossing Catheter · Reflow Medical, Inc. · Jun 9, 2017
K083833 — ENABLER-P CATHETER, 1000 PART NUMBER:1160 · Endocross, Ltd. · May 22, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
AUG 3 0 2001
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS
# SUBMITTER INFORMATION
| A. | Company Name: | IntraLuminal Therapeutics, Inc. |
|----|--------------------|---------------------------------------------------------------------------------|
| B. | Company Address: | 6354 Corte Del Abeto - Suite A<br>Carlsbad, CA 92009 |
| C. | Company Phone: | (760) 918-1820 |
| D. | Company Facsimile: | (760) 918-1823 |
| E. | Contact Person: | Pamela Misajon<br>Vice President of Regulatory Affairs and Quality<br>Assurance |
#### DEVICE IDENTIFICATION
| A. | Device Trade Name: | ILT Support Catheter and ILT Support Catheter<br>Advancing Mechanism |
|----|----------------------|-----------------------------------------------------------------------------------------|
| B. | Catalog Number: | C114LW1 – ILT Support Catheter<br>A115AM1 – ILT Support Catheter Advancing<br>Mechanism |
| C. | Device Common Name: | Percutaneous Catheter |
| D. | Classification Name: | Percutaneous Catheter |
| E. | Device Class: | Class II (per 21 CFR 870.1250) |
### IDENTIFICATION OF MODIFIED DEVICE
The ILT Support Catheter is a similar in design, materials, mode of operation and intended use to the IntraLuminal Therapeutics, Inc. ILT 0.014" Catheters cleared under 510(k) K001992. The modified device separates the catheter from the advancing mechanism.
#### DEVICE DESCRIPTION
The ILT Support Catheter is a single-lumen intravascular catheter intended for percutaneous use. It is designed to be used in conjunction with a 0.014" steerable guidewire to gain access to locations within the cardiovascular system that are remote from the site of insertion. Once accessed, guidewires may be exchanged within the catheter. The ILT Support Catheter may also be used for The ILT Support Catheter may also be used to provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
> 046 ﺎ ﮨﺎ
{1}------------------------------------------------
The effective length of the ILT Support Catheter is a nominal 135 cm. The nominal The enective longin of the 12 - Saffench (0.037"). The distal region incorporates a maximum outside diamotor to 210 fretton (onder fluoroscopy. The inside diameter of the ILT Support Catheter shaft will accommodate a commercially available 0.014" quidewire.
-The ILT-Support Catheter Advancing Mechanism is a handle with a connecting luer in The ILT Oupport Catholor nated to the ILT Support Catheter. The handle includes and advancing mochanism that allows for incremental advancement of the guidewire. This anovaning moonament the froo movements of the guidewire through the catheter shaft.
The ILT Support Catheter and the ILT-Support-Catheter-Advancing-Mechanism are packaged in a Tyvek® covered tray and heat-sealed to form a sterile barrier. The packaged catheter is sterilized by ethylene oxide gas. The ILT Support Catheter is provided "STERILE" and "Non-pyrogenic", and is intended for single use only.
#### INTENDED USE
The ILT Support Catheter is indicated to be used in conjunction with a steerable guidewire in order to access discreet regions of the vasculature. Once the region has been accessed, an exchange of one guidewire for another can occur. The ILT Support Catheter may also be used to provide a conduit for the delivery of saline solutions or diagnostic contrast agents.
#### TECHNOLOGICAL CHARACTERISTICS
The ILT Support Catheter is similar in basic materials, design, construction and mechanical performance to the previously cleared ILT .014" Catheter. The modification separates the advancing mechanism (handle) from the catheter and allows the ability to use the ILT Support Catheter with or without the advancing mechanism..
## BIOCOMPATIBILITY AND PERFORMANCE DATA
Biocompatibility testing and in vitro bench studies were conducted to evaluate the biological and performance characteristics of the ILT Support Catheter. Biocompatibility test results indicate that the device materials are biocompatible. Performance test results indicate that the device satisfies functional performance requirements when used as indicated.
## CONCLUSIONS DRAWN FROM STUDIES
The results of testing demonstrate that the ILT Support Catheter with out the advancing mechanism is substantially equivalent to the predicate device and is capable of safely and accurately performing the stated intended use.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 3 0 2001
Ms. Pamela Misajon Vice President of Regulatory Affairs and Quality Assurance IntraLuminal Therapeutics, Inc. 6354 Corte Del Abeto, Suite A Carlsbad, CA 92009
K012169 Re:
ILT Support Catheter and ILT Support Catheter Advancing Mechanism Regulation Number: 870.1250 Regulatory Class: II (two) Product Code: DQY Dated: August 10, 2001 Received: August 13, 2001
Dear Ms. Misajon:
We have reviewed your Section 510(k) notification of intent to market the device referenced we have teviewed your becally be is substantially equivalent (for the indications for use above and we nave determined the avveeted predicate devices marketed in interstate commerce Stated in the energiane) to legally manage of the Medical Device Amendments, or to devices that prof to May 20, 1970, the chassistence with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosment Act (11ct). Tournal) ; actives ; ; increases) ; ; } = } = = = include requirements for annual provisions of the rea. "The Prices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 remarket / pprovally, while Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Manancial ing Praval regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspoculity, the Read and may result in regulatory action. In addition, FDA may publish compry with the Original concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Ms. Pamela Misajon
This letter will allow you to begin marketing your device as described in your 510(k) premarket This letter will anow you to ocemanaling , quivalence of your device to a legally marketed noutication. The I Driving of vice in the your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and If you desire specific days of the stic devices), please contact the Office of Compliance at additionally 607.10 for in This chaghestions on the promotion and advertising of your device, (201) 594-4560. Runnenary, 2011 - 12:4639. Also, please note the regulation prease conder the Childer reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small information on your responsion.com Assistance at its toll-free number (800) 638-2041 or 1901) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
James E. Dillard III
Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE
KO12169
510(k) Number: Device Name:
ILT Support Catheter (with or without Advancing Mechanism)
Indications For Use:The ILT Support Catheter with or without the advancing mechanism is indicated to be used in conjunction with a steerable guidewire in indicated to be access discreet regions of the vasculature. Once the order to accessed, an exchange of one guidewire for another can occur. The ILT Support Catheter may also be used another san occanduit for the delivery of saline solutions or diagnostic contrast agents.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
✓
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Per 21 CFR 801.109)
Division of Cardiovascular & Perinatal Services
510(k) Number K021669
016
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.