HI-TORQUE BALANCE TREK GUIDE WIRE

K991152 · Guidant Corp. · DQX · Apr 29, 1999 · Cardiovascular

Device Facts

Record IDK991152
Device NameHI-TORQUE BALANCE TREK GUIDE WIRE
ApplicantGuidant Corp.
Product CodeDQX · Cardiovascular
Decision DateApr 29, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The wire is also intended to facilitate the placement of equipment such as atherectomy and compatible stent devices during other diagnostic and therapeutic procedures.

Device Story

Steerable guide wire (0.014" diameter) used to navigate vasculature to facilitate placement of interventional equipment (balloon catheters, stents, atherectomy devices). Features radiopaque distal tip (straight or J-shape) and hydrophilic coating on distal section; proximal section coated with polytetrafluoroethylene. Available in 175 cm, 190 cm, and 300 cm lengths. Used by physicians in clinical settings during PTCA/PTA procedures. Device provides mechanical support and guidance for interventional tools, aiding in successful lesion access and treatment.

Clinical Evidence

Bench testing only. Functional tests included Distal Tip Pull Test, Distal Tip Turns-to-Failure Test, Rotational Accuracy Test, and Tip Flexibility Test. Results demonstrated performance equivalent to the predicate device with no new safety or effectiveness issues.

Technological Characteristics

0.014" diameter steerable guide wire; radiopaque tip; hydrophilic coating (distal); PTFE coating (proximal). Materials, biocompatibility, and performance properties are equivalent to predicate. Design modifications include core and tip coil geometry. Sterilization and packaging identical to predicate.

Indications for Use

Indicated for patients undergoing percutaneous transluminal coronary angioplasty (PTCA), percutaneous transluminal angioplasty (PTA), or other diagnostic and therapeutic procedures requiring the placement of balloon dilatation catheters, atherectomy devices, or compatible stents.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 4/29/99 K991152 31 # ATTACHMENT 4 ### 510(k) Summary The 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92 | 1. | Submitter's Name: | Guidant Corporation<br>Advanced Cardiovascular Systems, Inc. | |-----|-----------------------------|--------------------------------------------------------------------------------------------------| | 2. | Submitter's Address: | 3200 Lakeside Drive<br>Santa Clara, CA 95054 | | 3. | Telephone: | 408-235-3995 | | 4. | Fax: | 408-235-3743 | | 5. | Contact Person: | Margaret Anderson | | 6. | Date Prepared: | April 5, 1999 | | 7. | Device Trade Name: | HI-TORQUE BALANCE TREKTM Guide Wires with<br>HYDROCOATTM Hydrophilic Coating | | 8. | Device Common Name: | Guide Wire | | 9. | Device Classification Name: | Catheter Guide Wire<br>(74DQX) | | 10. | Predicate Device: | ACS HI-TORQUE BALANCE HEAVYWEIGHTTM Guide<br>Wire with HYDROCOATTM Hydrophilic Coating (K982083) | {1}------------------------------------------------ - 11. Device Description: The HI-TORQUE BALANCE TREK™ Guide Wire with HYDROCOAT™ Hydrophilic Coating is a steerable guide wire with a maximum diameter of 0.014″ and available in: 175 cm and 190 cm extendable lengths and a 300 cm exchange length. The proximal end of the 175 cm and 190 cm models are tapered to fit into the hypotube portion of the ACS DOC® Guide Wire Extension. The distal end of the guide wire has a radiopaque tip that are available either as a straight or as a preshaped J. The hydrophilic coating is applied to the distal portion of the wire guide wire. The proximal section of the guide wire is coated with polytetrafluoroethylene. - 12. Intended Use: To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The wire is also intended to facilitate the placement of equipment such as atherectomy and compatible stent devices during other diagnostic and therapeutic procedures. - 13. Technological Characteristics: Comparisons of the new and predicate devices show that technological characteristics such as materials, biocompatibility, performance properties, sterilization, and packaging are identical or substantially equivalent to the currently marketed predicate device. The design modifications of the new guide wire compared to that of the predicate wire are the modified core and tip coils. - 14. Performance Data: In vitro bench testing and in vivo performance evaluations were performed to demonstrate that the HI-TORQUE BALANCE TREK™ Guide Wire with HYDROCOAT™ Hydrophilic Coating met the acceptance criteria and performed similar to the predicate device. The following functional tests were performed: Distal Tip Pull Test, Distal Tip Turns-to-Failure Test, Rotational Accuracy Test and Tip Flexibility Test. The results from the tests demonstrated that the new The HI-TORQUE BALANCE TREK™ Guide Wires met the acceptance criteria and performed in a manner equivalent the predicate ACS HI-TORQUE BALANCE HEAVYWEIGHT™ Guide Wire. No new safety or effectiveness issues were raised during the testing program. {2}------------------------------------------------ #### 15. Conclusions Since the new guide wire has the same intended use, materials, technological characteristics, performance properties, identical sterilization and packaging, and no new safety or effectiveness issues, the HI-TORQUE BALANCE TREK™ Guide Wire with HYDROCOAT™ Hydrophilic Coating may be considered substantially equivalent to the predicate ACS HI-TORQUE BALANCE HEA VYWEIGHT™ Guide Wire with HYDROCOAT™ Hydrophilic Coating. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR 2 9 1999 Ms. Marqaret Anderson Regulatory Affairs Guidant Corporation P.O. Box 58167 Santa Clara, CA 95052 Re: K991152 > HI-TORQUE BALANCE TREK™ Guide Wires with Trade Name: HYDROCOAT™ Hydrophilic Coating Requlatory Class: II Product Code: DOX Dated: April 5, 1999 Received: April 6, 1999 Dear Mr. Anderson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to {4}------------------------------------------------ Page 2 - Ms. Margaret Anderson your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callah Director Division of Cardiovascular, Respiratory and Neuroloqical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## ATTACHMENT 2 ### Indications for Use Statement | 510(k) Number (if known) | | |--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name | HI-TORQUE BALANCE TREK™ Guide Wires with HYDROCOAT™ Hydrophilic Coating | | Indications for Use | The intended use for the HI-TORQUE BALANCE TREK™ Guide Wires with HYDROCOAT™ Hydrophilic Coating is: To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The wire is also intended to facilitate the placement of equipment such as atherectomy and compatible stent devices during other diagnostic and therapeutic procedures. | ### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED #### Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Use OR Over-The-Counter A. Wetteng (Division Sign-Off) Division of Cardiovascular, Respirato and Neurological Device 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...