DOCKING-EXTENSION WIRE

K982592 · Schneider (Europe) GmbH · DQX · Oct 21, 1998 · Cardiovascular

Device Facts

Record IDK982592
Device NameDOCKING-EXTENSION WIRE
ApplicantSchneider (Europe) GmbH
Product CodeDQX · Cardiovascular
Decision DateOct 21, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The Docking-Extension Wire is intended for extension of the working length of an already introduced guide wire when exchanging over-the-wire interventional devices during an angioplasty procedure.

Device Story

The SCHNEIDER Docking-Extension Wire is a stainless-steel accessory used to elongate compatible SCHNEIDER extendable guide wires (C-Thru, Hannibal, and Clyde) during angioplasty. The device features a 0.014" outer diameter and a 130 cm length, with a distal preformed superelastic connecting hypotube. A docking aid facilitates the insertion of the guide wire's proximal end into the hypotube, allowing for the transmission of axial forces. The device is used by clinicians in an interventional setting to enable the exchange of over-the-wire interventional devices. By extending the working length of the guide wire, it allows the physician to maintain access while swapping devices, potentially simplifying the procedure and reducing the need for complete guide wire removal and re-insertion.

Clinical Evidence

Bench testing only. Performance testing included docking joint tensile strength, docking junction fatigue, and glued bond break load. Biocompatibility testing included cytotoxicity, coagulation, indirect hemolysis, intracutaneous reactivity, systemic toxicity, and sensitization.

Technological Characteristics

Stainless-steel construction; 0.014" (0.36 mm) outer diameter; 130 cm length; distal superelastic connecting hypotube; docking aid for guide wire introduction. Mechanical coupling principle for axial force transmission. Compatible with SCHNEIDER extendable guide wires (C-Thru, Hannibal, Clyde).

Indications for Use

Indicated for patients undergoing angioplasty procedures requiring the extension of an already introduced guide wire to facilitate the exchange of over-the-wire interventional devices.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## OCT 21 1998 ## 10 510(k) Summary for the SCHNEIDER Docking-Extension Wire This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and CFR 807.92. The assigned 510(k) number is | Date prepared: | July 21, 1998 | |-------------------------|----------------------------------------------------------------------------| | Sponsor: | SCHNEIDER (Europe) GmbH<br>Ackerstrasse 6<br>CH-8180 Bülach<br>Switzerland | | Contact: | Thomas Thaler, Ph. D.<br>Director, Clinical and Regulatory Affairs | | Trade/Proprietary Name: | Docking-Extension Wire | | Classification: | Class II | | Predicate Device: | ACS DOC™ Extension Wire | ## Device Description: The SCHNEIDER Docking-Extension Wire is used to elongate the working length of compatible SCHNEIDER extendable guide wires. The stainless-steel Docking-Extension Wire has an outer diameter of 0.014" (0.36 mm) and a length of 130 cm. Its distal end bears a preformed superelastic connecting hypotube. This superelastic property allows the proximal end of the quide wire to be fixed in the hypotube to such an extent. that axial forces can easily be transmitted. A docking aid enables easy introduction of the extendable end of the guide wire into the hypotube. The device is compatible with the SCHNEIDER extendable guide wires (C-Thru, Hannibal and Clyde). The currently marketed guide wires had been slightly modified at the proximal end to create a tapered end in order to be accommodated into the hypotube of the Docking-Extension Wire. This modification does not in any way affect the safety or performance of the guide wire. ## Technological Characteristics: Equivalence in technological characteristics was substantiated by comparative performance testing including docking joint tensile strength, docking junction fatigue and glued bond break load, and compatibility with interventional devices. Biocompatibility testing of the Docking-Extension Wire was conducted with regard to cytotoxicity, coagulation, indirect hemolysis, intracutaneus reactivity, systemic toxicity and sensitization. The results of all testing indicated that the SCHNEIDER Docking-Extension Wire is equivalent to the ACS DOC™ Extension Wire and is, therefore, safe for the intended use. {1}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 21 1998 Thomas Thaler, Ph.D. SCHNEIDER (Europe) GmbH Ackerstrasse 6 CH-8160 Bülach Switzerland Re: K982592 Docking-Extension Wire Requlatory Class: II (two) Product Code: DQX Dated: July 22, 1998 Received: July 24, 1998 Dear Dr. Thaler: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {2}------------------------------------------------ Page 2 - Thomas Thaler, Ph.D. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Thomas J. Collehan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ | Page | | |------|--| |------|--| 510(k) Number (if known): Device Name: Indications for Use: Docking-Extension Wire The Docking-Extension Wire is intended for extension of the working length of an already introduced guide wire when exchanging over-the-wire interventional devices during an angioplasty procedure. PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) A. Doyle Ratliff Concurrence of CDRH, Office of Device Evaluation (ODE)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...