HI-TORQUE FLOPPY GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING, HI-TORQUE STANDARD GUIDE WIRE WITH HYDROCOAT HYDROPHILIC

K974773 · Guidant Corp. · DQX · Mar 13, 1998 · Cardiovascular

Device Facts

Record IDK974773
Device NameHI-TORQUE FLOPPY GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING, HI-TORQUE STANDARD GUIDE WIRE WITH HYDROCOAT HYDROPHILIC
ApplicantGuidant Corp.
Product CodeDQX · Cardiovascular
Decision DateMar 13, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ACS HI-TORQUE EXTRA S'PORT™ Guide Wire with HYDROCOAT™ Hydrophilic Coating and the ACS HI-TORQUE IRON MANTM Guide Wire with HYDROCOAT™ Hydrophilic Coating are also intended to facilitate the placement of equipment such as atherectomy, IVUS and compatible stent devices during other diagnostic and therapeutic intravascular procedures.

Device Story

Steerable guide wires used to facilitate placement of interventional equipment (balloon catheters, atherectomy devices, IVUS, stents) during PTCA, PTA, and other intravascular procedures. Device consists of stainless steel core wire with radiopaque distal tip; distal portion features HYDROCOAT™ hydrophilic coating; proximal section features PTFE coating. Available in .010", .014", and .018" diameters; 190 cm (extendable) or 300 cm (exchange) lengths. Operated by physicians in clinical settings (e.g., cath lab). Provides tactile feedback and trackability to navigate vasculature; hydrophilic coating reduces friction during navigation. Output is physical guidance of interventional tools to target lesion site; aids clinical decision-making by enabling successful delivery of therapeutic devices.

Clinical Evidence

Bench testing only. Tests included accelerated aging, distal tip pull test, distal tip turns-to-failure, rotational accuracy, tip flexibility, and coating adherence/integrity. Results demonstrated that the subject devices met acceptance criteria and performed equivalently to predicate devices.

Technological Characteristics

Stainless steel core wire; radiopaque distal tip (straight or J-shape); distal portion coated with HYDROCOAT™ hydrophilic coating; proximal section coated with PTFE. Diameters: .010", .014", .018". Lengths: 190 cm (extendable) or 300 cm (exchange).

Indications for Use

Indicated for patients undergoing percutaneous transluminal coronary angioplasty (PTCA), percutaneous transluminal angioplasty (PTA), or other diagnostic and therapeutic intravascular procedures requiring the placement of balloon dilatation catheters, atherectomy devices, IVUS, or compatible stents.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 194773 167 # 510(k) SUMMARY This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92. | 1. | Submitter's Name: | Guidant Corporation<br>Advanced Cardiovascular Systems, Inc. | MAR 1 3 1998 | |----|----------------------|--------------------------------------------------------------|--------------| | | Submitter's Address: | 3200 Lakeside Drive<br>Santa Clara, CA 95052 | | | | Telephone: | 408-235-3995 | | | | Fax: | 408-235-3743 | | | | Contact Person: | Margaret Anderson | | | | Date Prepared: | December 19, 1997 | | - 2. Device Trade Name: - HI-TORQUE FLOPPY® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TOROUE STANDARD® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TOROUE INTERMEDIATE® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TORQUE APPROACH® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TORQUE FLOPPY II® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TORQUE TRAVERSE® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - ACS HI-TORQUE EXTRA S'PORT™ Guide Wire with HYDROCOAT™ Hydrophilic . Coating - HI-TORQUE FLOPPY® EXTRA SUPPORT Guide Wire with HYDROCOAT™ . Hydrophilic Coating - HI-TORQUE FLOPPY II ® EXTRA SUPPORT Guide Wire with HYDROCOAT™ . Hydrophilic Coating - ACS HI-TORQUE IRON MAN® Guide Wire with HYDROCOAT™ Hydrophilic Coating . Device Common Name: Guide Wire Device Classification Name: Catheter Guide Wire (74DQX) {1}------------------------------------------------ #### Predicate Device: 3. - HI-TORQUE FLOPPY® Guide Wire with MICROGLIDE® Coating . - HI-TORQUE STANDARD® Guide Wire with MICROGLIDE® Coating . - HI-TORQUE INTERMEDIATE® Guide Wire with MICROGLIDE® Coating . - HI-TORQUE APPROACH® Guide Wire with MICROGLIDE® Coating . - HI-TORQUE FLOPPY II® Guide Wire with MICROGLIDE® Coating . - HI-TORQUE TRAVERSE® Guide Wire with MICROGLIDE® Coating . - ACS HI-TORQUE EXTRA S'PORT™ Guide Wire with MICROGLIDE® Coating ' . - HI-TORQUE FLOPPY® EXTRA SUPPORT Guide Wire with MICROGLIDE® Coating . - HI-TORQUE FLOPPY II® EXTRA SUPPORT Guide Wire with MICROGLIDE® Coating . - ACS HI-TORQUE IRON MAN® Guide Wire with MICROGLIDE® Coating . - ChoICE™ PT Plus Guide Wire . - Device Description: 4. The HI-TORQUE Guide Wires with HYDROCOAT™ Hydrophilic Coating are steerable guide wires with a nominal diameters of .010",014" and .018"and available in two lengths: a 190 cm extendable length and a 300 cm exchange length. The proximal end of the 190 cm models are tapered to fit into the hypotube portion of the ACS DOC Guide Wire Extension. The wires are constructed from a stainless steel core wire. The distal end of this guide wires have radiopaque tips that are available either as a straight, shapeable configuration or as a preshaped J. The hydrophilic coating is applied to the distal portion of the wire guide wire. The proximal section of the guide wire is coated with polytetrafluoroethylene. - క. Intended Use: The HI-TORQUE BALANCE® Guide Wire with HYDROCOAT™ Hydrophilic Coating and the ACS HI-TORQUE BALANCE MIDDLEWEIGHT™ Guide Wire with HYDROCOAT™ Hydrophilic Coating have the following intended uses: - To facilitate the placement of balloon dilatation catheters during percutaneous . transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). - The ACS HI-TORQUE EXTRA S'PORT™ Guide Wire with . HYDROCOAT™ Hydrophilic Coating and the ACS HI-TORQUE IRON MANTM Guide Wire with HYDROCOAT™ Hydrophilic Coating are also intended to facilitate the placement of equipment such as atherectomy, IVUS and compatible stent devices during other diagnostic and therapeutic intravascular procedures. - Technological Characteristics: 6. {2}------------------------------------------------ Comparisons of the new and predicate devices show that technological characteristics such as materials, biocompatibility, performance properties, sterilization, and packaging are identical or substantially equivalent to the currently marketed predicate devices. The design feature that distinguish the new guide wires from that of the predicate wires is the new hydrophilic coating. #### 7. Performance Data: Bench testing was performed to demonstrate that the ACS HI-TORQUE® Guide Wire with HYDROCOAT™ Hydrophilic Coating met the acceptance criteria and performed similar to the predicate HI-TORQUE® Guide Wire with MICROGLIDE® Coating. The following tests were performed: - Accelerated Aging . - Distal Tip Pull Test . - Distal Tip Turns-to-Failure Test . - Rotational Accuracy Test . - Tip Flexibility Test . - . Coating Adherence/Integrity The results from the bench tests showed that the new ACS HI-TORQUE® Guide Wires with HYDROCOAT™ Hydrophilic Coating met the acceptance criteria and performed in a manner equivalent the predicate ACS HI-TORQUE® Guide Wire with MICROGLIDE® Coating and the ChoICE™ PT Plus Guide Wire. No new safety or effectiveness issues were raised during the testing program. #### 8. Conclusions: Since the new guide wires have the same intended use, design and technological characteristics, equivalent performance properties, identical sterilization and packaging, and no new safety or effectiveness issues, the ACS HI-TORQUE® Guide Wires with HYDROCOAT™ Hydrophilic Coating may be considered substantially equivalent to the predicate ACS HI-TORQUE® Guide Wires with MICROGLIDE® Coating. Additionally, the new guide wires performed substantially equivalent with regards to coating adhesion integrity to the ChoICETM PT Plus Guide Wire therefore, we conclude the new guide wire may be considered substatntially equivalent to the predicate devices. {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR | 3 1998 Ms. Margaret Anderson Guidant Corporation 3200 Lakeside Drive Santa Clara, CA 95054 Re: K974773 HI-TORQUE® Guide Wires with HYDROCOAT™ Hydrophilic Coating Regulatory Class: II (two) 74 DQX Product Code: December 19, 1997 Dated: Received: December 22, 1997 Dear Ms. Anderson: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good. manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) ^ inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {4}------------------------------------------------ #### Page 2 - Ms. Margaret Anderson This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use Statement ### 510(k) Number (if known): ### Device Name: - HI-TORQUE FLOPPY® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TORQUE STANDARD® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TORQUE INTERMEDIATE® Guide Wire with HYDROCOAT™ Hydrophilic . Coating - HI-TORQUE APPROACH® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TORQUE FLOPPY II® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - HI-TORQUE TRAVERSE® Guide Wire with HYDROCOAT™ Hydrophilic Coating . - ACS HI-TORQUE EXTRA S'PORT™ Guide Wire with HYDROCOAT™ Hydrophilic . Coating - HI-TORQUE FLOPPY® EXTRA SUPPORT and HI-TORQUE FLOPPY II ® EXTRA . SUPPORT Guide Wire with HYDROCOAT™ Hydrophilic Coating - ACS HI-TORQUE IRON MAN® Guide Wire with HYDROCOAT™ Hydrophilic Coating . #### Indications for Use: - · Facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). - The ACS HI-TORQUE EXTRA S'PORT™ Guide Wire with HYDROCOAT™ Hydrophilic Coating and the ACS HI-TORQUE IRON MAN™ Guide Wire with HYDROCOAT™ Hydrophilic Coating are also intended to facilitate the placement of equipment such as atherectomy and compatible stent devices during other diagnostic and therapeutic procedures. ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter (Optional Format 1-1-96) Tu A. A (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices K974773 510(k) Number __
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...