TRANSEND GUIDEWIRE
Device Facts
| Record ID | K964611 |
|---|---|
| Device Name | TRANSEND GUIDEWIRE |
| Applicant | Boston Scientific Corp |
| Product Code | DQX · Cardiovascular |
| Decision Date | Jan 28, 1997 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1330 |
| Device Class | Class 2 |
Indications for Use
The BSC Transend Steerable Guidewire with hydrophilic coating and accessories are intended for general intravascular use including the peripheral and neurovasculature. The devices will be provided sterile, and are intended for one procedure use only.
Device Story
Steerable guidewire for intravascular navigation; peripheral and neurovascular applications. Device features hydrophilic coating to enhance lubricity. Used by physicians during interventional procedures to facilitate catheter placement. Single-use, sterile device. Performance relies on mechanical properties: tip flexibility, torque strength, and torque response. Clinical benefit derived from navigation through complex vascular anatomy.
Clinical Evidence
Bench testing included tip tensile strength, tip flexibility, torque strength, torque response, coating adherence, coating extraction, and lubricity testing. Biocompatibility and in vivo testing performed. No clinical trial data provided.
Technological Characteristics
Steerable guidewire; hydrophilic coating. Materials evaluated for biocompatibility. Performance specifications: tip tensile, tip flexibility, torque strength, torque response, coating adherence, lubricity. Sterile, single-use.
Indications for Use
Indicated for general intravascular use in peripheral and neurovasculature procedures.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
- BSC Transend Guidewire (K964611)
- Target Therapeutics Dasher-10 Guidewire
- Target Therapeutics Seeker-Lite Guidewire
- Terumo Corporation Glidewire Gold Guidewire
- Micro Interventional Systems Quicksilver Guidewire
Related Devices
- K190843 — Synchro2 Support Guidewire · Stryker · May 1, 2019
- K191687 — pORTAL Steerable Hydrophilic Guidewire and pORTAL EXT Extension Wire · Phenox Limited · Oct 9, 2019
- K252122 — DCwire Micro-guidewire · Shanghai Achieva Medical Suzhou Co., Ltd. · Mar 16, 2026
- K121776 — AGILITY STEERABLE GUIDEWIRE NEUROSCOUT STEERABLE GUDIEWIRE · Codman & Shurtleff, Inc. · Aug 14, 2012
- K971254 — TRANSEND EX PLATINUM GUIDEWIRE · Boston Scientific Corp · Jul 1, 1997