CORDIS STABILIZER PLUS XS

K962932 · Cordis Corp. · DQX · Oct 25, 1996 · Cardiovascular

Device Facts

Record IDK962932
Device NameCORDIS STABILIZER PLUS XS
ApplicantCordis Corp.
Product CodeDQX · Cardiovascular
Decision DateOct 25, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The Cordis STABILIZER Plus XS Steerable Guidewires are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature.

Device Story

Steerable guidewire; extra support version of WIZDOM guidewire. Input: manual manipulation by physician during angiographic procedures. Function: provides structural support for introduction of interventional devices (e.g., stents) into coronary/peripheral vasculature. Output: physical guidance/pathway for catheters. Used in clinical/interventional settings; operated by physicians. Benefit: facilitates delivery of bulky interventional devices.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Steerable guidewire; utilizes larger diameter corewires and modified tapers compared to conventional WIZDOM guidewire for increased support. Materials identical to predicate WIZDOM guidewire. Class II device (21 CFR 870.1330).

Indications for Use

Indicated for use in angiographic procedures to facilitate introduction and positioning of catheters and interventional devices within coronary and peripheral vasculature.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Submission Summary (Full Text)

{0} OCT 25 1996 K962932 # PREMARKET NOTIFICATION Cordis Corporation STABILIZER™ Plus XS Steerable Guidewire ## SUMMARY OF SAFETY AND EFFECTIVENESS ### I. General Provisions Common or Usual Name: Catheter Guide Wire Proprietary Name: Cordis STABILIZER Plus XS Steerable Guidewire ### II. Name of Predicate Devices Cordis WIZDOM Steerable Guidewire (K953760) Meditech/Boston Scientific Platinum Plus™ Guidewire (K935997) Scimed/Boston Scientific Platinum Plus™ Coronary Guidewire ### III. Classification Catheter Guide Wire, Class II - 21 CFR 870.1330 74DQX - Cardiovascular ### IV. Performance Standards Performance standards have not been established by the FDA under section 514 of the Food, Drug and Cosmetic Act. ### V. Intended Use and Device Description The Cordis STABILIZER Plus XS Steerable Guidewires are intended for use in angiographic procedures to introduce and position catheters and interventional devices within the coronary and peripheral vasculature. The Cordis STABILIZER Plus XS Steerable Guidewire is an extra support version of the Cordis WIZDOM Steerable Guidewire. The extra support versions of the WIZDOM Steerable Guidewire utilize slightly larger diameter corewires and different tapers compared to the conventional WIZDOM guidewire in order to provide the extra support required for bulkier device (such as a stent) introduction. ### VI. Biocompatibility The materials incorporated into the STABILIZER Plus XS are exactly the same as the materials used for the currently marketed WIZDOM Steerable Guidewire.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...