← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K960755

# CARDIOTHERM BLOOD CARDIOPLEGIA SYSTEMS (K960755)

_Medtronic Cardiac Surgery, Medtronic, Inc. · DQX · May 23, 1996 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K960755

## Device Facts

- **Applicant:** Medtronic Cardiac Surgery, Medtronic, Inc.
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** May 23, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Medtronic CardioTherm™ Blood Cardioplegia System is intended for mixing, cooling, warming, and delivery of oxygenated blood and/or asanguineous cardioplegia solution.

## Device Story

Single-use disposable device; integrates into extracorporeal circuit. Inputs: arterial blood from oxygenator; asanguineous cardioplegia solution. Mechanism: polycarbonate housing with nonporous polypropylene hollow fibers; blood/cardioplegia flows through fibers; temperature-regulated water flows around fibers for heat exchange. Output: temperature-controlled blood/cardioplegia mixture delivered to patient. Clinician-operated; used in surgical settings. Tubing sets with specific inner diameters determine blood-to-cardioplegia mixing ratios (e.g., 1:1, 2:1, 4:1, 9:1). Device includes bubble chamber, vent port, temperature probe port, and pressure monitoring line. Facilitates myocardial protection during cardiac surgery by providing precise temperature and composition control of cardioplegic solution.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, design analysis, and comparison of technological characteristics and nominal specifications against legally marketed predicate devices.

## Technological Characteristics

Heat exchanger with polycarbonate housing; nonporous polypropylene hollow fibers; polyethylene screen support. Tubing sets: polyvinyl chloride (PVC) with polycarbonate/PVC connectors. Dimensions: tubing inner diameters 3/32" to 17/64". Connectivity: standalone extracorporeal circuit component. Sterilization: not specified. Sensing: temperature probe port, pressure monitoring line. Energy: external water-based cooling/warming source.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems ([K934355](/device/K934355.md))
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems ([K896807](/device/K896807.md)/ [K944219](/device/K944219.md))
- Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems ([K904171](/device/K904171.md))
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System ([K925369](/device/K925369.md))

## Reference Devices

- Medtronic Extracorporeal InterSept™ Custom Tubing Packs ([K800178](/device/K800178.md))
- Medtronic InterSept™ Crystalloid Cardioplegia Delivery System ([K885154](/device/K885154.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

MAY 23 1996
K960755

# SUMMARY OF SAFETY AND EFFECTIVENESS

## COMPANY AND CONTACT PERSON

Medtronic, Inc.
Cardiopulmonary Division
4633 E. La Palma Avenue
Anaheim, CA 92807
714-779-3700

Debra J. Kridner, Manager
Regulatory Affairs

## DEVICE NAME

CardioTherm™ Blood Cardioplegia Delivery Systems
(CTXXX Series and CTXXXBR Series)

## NAME OF PREDICATE OR LEGALLY MARKETED DEVICE

- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/ K944219)
- Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)

## DESCRIPTION OF DEVICE

Each Medtronic CardioTherm™ Blood Cardioplegia System is a single use, disposable device designed to mix arterial blood from an oxygenator with asanguineous cardioplegia solution in specific ratios depending on the tubing set configurations. The blood/cardioplegia solution is then cooled/warmed and delivered to the patient.

The Medtronic CardioTherm™ Cardioplegia Heat Exchanger consists of a polycarbonate housing which incorporates a bubble chamber. A stopcock is attached to the upper luer port of the housing lid allowing for air venting through a purge line during priming and pressure monitoring through a vented or nonvented pressure monitoring line throughout the procedure. A temperature probe port is located adjacent to the blood outlet port at the bottom of the device, allowing for temperature monitoring of the blood/cardioplegia solution prior to patient delivery.

The heat exchanger housing contains nonporous polypropylene hollow fibers, which are supported on their exterior by a polyethylene screen. These polypropylene fibers run longitudinally through the polycarbonate chamber. Blood/cardioplegia solution flows through the polypropylene hollow

510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 144

{1}

fibers. The walls of these fibers provide a barrier between the blood/cardioplegia solution and the cooling/warming water.

The flow of solution through the Medtronic CardioTherm™ Blood Cardioplegia Heat Exchanger is as follows:

## Blood/Cardioplegia

The blood and/or cardioplegia solution enters the bottom of the device, flows up through the polypropylene hollow fibers, flows back down a center collecting tube, and exits the from the bottom of the device.

## Water

The cooling/warming water enters a side port, passes around the outside walls of the polypropylene fibers, then exits through a parallel side port.

Heat exchange occurs as the blood/cardioplegia solution passes through the hollow fibers, while the temperature regulated cooling/warming water passes around the outside walls of the same fibers. The blood/cardioplegia solution is cooled/ warmed as it flows through the heat exchanger.

By connecting tubing of specific inner diameters to the heat exchanger, a predetermined approximate ratio of blood to cardioplegia solution may be delivered to the patient. The tubing sets are manufactured with polyvinyl chloride tubing and polycarbonate/polyvinyl chloride connectors which are currently used as components in the commercially distributed Medtronic Extracorporeal InterSept™ Custom Tubing Packs (K800178) and Medtronic InterSept™ Crystalloid Cardioplegia Delivery System (K885154). Four of the various delivery ratios available are:

1. a tubing set for 1:1 ratio of blood to asanguineous cardioplegia.
2. a tubing set for 2:1 ratio of blood to asanguineous cardioplegia.
3. a tubing set for 4:1 ratio of blood to asanguineous cardioplegia.
4. a tubing set for 9:1 ratio of blood to asanguineous cardioplegia.

These tubing ratios are common to commercially available blood cardioplegia systems; including the Medtronic Electromedics, Inc. Blood Cardioplegia Delivery Systems, the Gish Biomedical, Inc. Single Pass Blood Cardioplegia (SPBC) Systems, the Avecor Cardiovascular MYOtherm™ Cardioplegia Delivery Systems and the Sorin Biomedical BCD ADVANCED Blood Cardioplegia System.

Tubing inner diameters for commercially available tubing sets generally range from 3/32" (2.4 mm) to 17/64" (6.7mm).

510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 145

{2}

# STATEMENT OF INTENDED USE

The Medtronic CardioTherm™ Blood Cardioplegia System is intended for mixing, cooling, warming, and delivery of oxygenated blood and/or asanguineous cardioplegia solution.

# STATEMENT OF INTENDED USE OF PREDICATE/MARKETED DEVICES

The Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems are intended to mix, cool, and deliver oxygenated blood/cardioplegia solution in a predetermined ratio.

The Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems are intended to mix, cool/warm, and deliver oxygenated blood/cardioplegia solution.

The Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems are intended to mix, cool, and deliver oxygenated blood/cardioplegia solution in a predetermined ratio.

The Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System are intended to mix, cool/warm, and deliver oxygenated blood/cardioplegia solution.

# STATEMENT OF TECHNOLOGICAL CHARACTERISTICS COMPARISON

A table comparing the intended use and technological characteristics of the Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems with the four noted substantially equivalent devices is provided in Attachment I.

# DETERMINATION OF SUBSTANTIAL EQUIVALENCE

This premarket notification submission provides substantial equivalence information and rationale which addresses the introduction to commercial distribution of the Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia System.

The Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems are substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:

- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/K944219)
- Avecor Cardiovascular - MYOtherm™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)

The Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems have an intended use which is substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:

510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 146

{3}

- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/K944219)
- Avecor Cardiovascular - MYOther™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)

The Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems have technological characteristics which are substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:

- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K89680/K944219)
- Avecor Cardiovascular - MYOther™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)

The design, construction, materials and nominal specifications of the Medtronic Cardiopulmonary CardioTherm™ Blood Cardioplegia Systems are either identical or substantially equivalent to other Blood Cardioplegia Systems currently in commercial distribution. These predicate/marketed devices include:

- Medtronic Electromedics, Inc. - Blood Cardioplegia Delivery Systems (K934355)
- Gish Biomedical, Inc. - Single Pass Blood Cardioplegia (SPBC) Systems (K896807/K944219)
- Avecor Cardiovascular - MYOther™ Cardioplegia Delivery Systems (K904171)
- Sorin Biomedical - BCD ADVANCED Blood Cardioplegia System (K925369)

510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 147

{4}

DEVICE COMPARISONS -- GENERAL CHARACTERISTICS AND NOMINAL SPECIFICATIONS

|   | Medtronic, Inc.
CardioTherm™
Blood Cardioplegia System | Medtronic, Inc./Electromedics Inc.
Blood Cardioplegia.
Delivery Systems | Gish Biomedical, Inc.
SPBC
Blood Cardioplegia Systems | Avecor Cardiovascular
MYOtherm™
Cardioplegia Delivery Systems | Sorin Biomedical
BCD ADVANCED
Blood Cardioplegia System  |
| --- | --- | --- | --- | --- | --- |
|  510(k) Number: | This submission | K934355 | K896807 | K904171 | K925369  |
|  Catalog Number(s): | CT-XXX Series
CT-XXXBR Series | D1081A, D1080A, D1082A | SPBC-XX Series
SPBC-XXBR Series | MYOtherm™ XX Series
MYOtherm™ XX BR Series | BCD ADV X Series
(includes bridge/shunt system)  |
|  Intended Use: | To mix, cool, warm and deliver
oxygenated blood/cardioplegia
solution. | To mix, cool, and deliver
oxygenated blood/cardioplegia
solution in a predetermined ratio. | To mix, cool/warm, and deliver
oxygenated blood/cardioplegia
solution. | To mix, cool, and deliver
oxygenated blood/cardioplegia
solution in a predetermined ratio. | To mix, cool/warm, and deliver
oxygenated blood/cardioplegia
solution.  |
|  Performance Characteristics:  |   |   |   |   |   |
|  Heat Exchanger  |   |   |   |   |   |
|  Priming Volume (mL) | 46 | 55 | 83 | 68 | 55  |
|  Maximum Flow Rate (LPM) | 1 | 3 | not rated | not rated | .5  |
|  Fluid Path Pressure (maximum) | 500 mmHg | not rated | 300 mmHg | 500 mmHg | not rated  |
|  Water Path Pressure (maximum) | 45 psi | 45 psi | 65psi | 40 psi | 80 psi  |
|  Technological Characteristics:  |   |   |   |   |   |
|  Configuration | Heat Exchanger | Heat Exchanger | Heat Exchanger | Heat Exchanger | Heat Exchanger  |
|  Heat Exchanger  |   |   |   |   |   |
|  Vent Port | Yes | Yes | Yes | Yes | Yes  |
|  Bubble Chamber | Yes | Yes | Yes | Yes | Yes  |
|  Temperature Well | Yes | Yes | Yes | Yes | Yes  |
|  Pressure Monitoring Line | Yes | Yes | Yes | Yes | Yes  |
|  Patient Delivery Line | Yes | Yes | Yes | Yes | Yes  |
|   | Various Tubing Sets | Various Tubing Sets | Various Tubing Sets | Various Tubing Sets | Various Tubing Sets  |

510(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 148

{5}

DEVICE COMPARISONS -- GENERAL CHARACTERISTICS AND NOMINAL SPECIFICATIONS

|   | Medtronic, Inc.
CardioTherm™
Blood Cardioplegia System | Medtronic, Inc./Electromedics Inc.
Blood Cardioplegia.
Delivery Systems | Gish Biomedical, Inc.
SPBC
Blood Cardioplegia Systems | Avecor Cardiovascular
MYOTherm™
Cardioplegia Delivery Systems | Sorin Biomedical
BCD ADVANCED
Blood Cardioplegia System  |
| --- | --- | --- | --- | --- | --- |
|  Materials  |   |   |   |   |   |
|  Heat Exchanger
Housing
Heat Exchanger | Polycarbonate
Polypropylene | Polycarbonate
Stainless steel | Polycarbonate
Aluminum | Polycarbonate
Stainless steel | Polycarbonate
Aluminum  |
|  Tubing Sets | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate | Polyvinyl chloride
Polycarbonate  |
|  Specifications  |   |   |   |   |   |
|  Tubing Set Ratios | Separate sets
(1:1, 2:1, 4:1 and 9:1) | Separate sets
(1:1, 2:1 and 4:1) | Separate sets
(1:1, 2:1, 4:1 and 9:1) | Separate sets
(1:1, 2:1, 4:1 and 9:1) | Separate sets
(1:1, 2:1, 4:1 and 8:1)  |
|  Tubing Inner Diameter | 1/12 inch to 1/4 inch | 1/8 inch to 1/4 inch | 3/32 inch to 17/64 inch | 1/8 inch to 1/4 inch | 3/32 inch to 17/64 inch  |
|  Ratio Capability | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 9:1
and all crystalloid with
various tubing sets) | Variable
(1:1, 2:1 and 4:1) | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 9:1,
and all crystalloid with
various tubing sets) | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 9:1,
and all crystalloid with
various tubing sets) | Variable (w/bridge clamps)
(all blood, 1:1, 2:1, 4:1, 8:1,
and all crystalloid with
various tubing sets)  |
|  Mode of Operation | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need. | This blood/cardioplegia system
is used in the extracorporeal
circuit for mixing, cooling,
warming and delivery of
oxygenated blood and/or
cardioplegia solution. The
multiple configurations allow
the clinician the capability of
providing varying ratios of
blood to cardioplegia solution
based on the patients’ clinical
need.  |

10(k) Notification - CardioTherm™
Medtronic, Inc.
Medtronic Cardiopulmonary
February 22, 1996
Confidential
Page 149

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K960755](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K960755)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
