0.014 INTERMEDIATE GUIDE WIRE W/PHOSPHORYLCHOLINE POLYMER COATING
Device Facts
| Record ID | K955135 |
|---|---|
| Device Name | 0.014 INTERMEDIATE GUIDE WIRE W/PHOSPHORYLCHOLINE POLYMER COATING |
| Applicant | Biocompatibles , Ltd. |
| Product Code | DQX · Cardiovascular |
| Decision Date | Apr 9, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1330 |
| Device Class | Class 2 |
Indications for Use
The Floppy and Intermediate guide wires are intended for use to facilitate the placement of PTA and/or PTCA balloon dilatation catheters within the peripheral or coronary vasculature. The guide wires are a 0.014 inch size and designed for safe use with appropriately sized balloon catheters. The guide wires are not intended for use in the cerebral vasculature.
Device Story
0.014-inch diameter guide wires; 175 cm length; 30 cm radiopaque spring coil distal tip; stainless steel core wire; hydrophilic phosphorylcholine polymer (PCP) coating. Used by physicians in clinical settings to navigate PTA/PTCA balloon catheters through peripheral or coronary vasculature. PCP coating reduces friction during catheter advancement. Device facilitates balloon placement; aids in therapeutic procedures; improves procedural efficiency. Not for cerebral use.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
0.014-inch diameter; 175 cm length; stainless steel core; radiopaque spring coil; hydrophilic phosphorylcholine polymer (PCP) coating. Sterilization method not specified.
Indications for Use
Indicated for facilitating placement of PTA and/or PTCA balloon dilatation catheters within peripheral or coronary vasculature. Contraindicated for use in cerebral vasculature.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
- Advanced Cardiovascular Systems, Inc. 0.014" Hi-Torque Floppy II® Guide Wire
- Advanced Cardiovascular Systems, Inc. 0.014" Hi-Torque Intermediate® Guide Wire
Related Devices
- K962266 — 0.014 INCH EXTRA LONG FLOPPY AND 0.014 EXTRA LONG INTERMEDIATE GUIDE WIRES WITH PHOSPHORYLCHOLINE (PC) POLYMER COATING · Biocompatibles Cardiovascular, Inc. · Sep 10, 1996
- K032615 — ASAHI PTCA CATHETER GUIDE WIRES, MODELS AGH147000 AND AGH147300 · Asahi Intecc Co., Ltd. · Sep 16, 2003
- K083904 — ASAHI SUOH PTCA GUIDE WIRE · Asahi Intecc Co., Ltd. · Jan 29, 2009
- K041531 — ASAHI PTCA GUIDE WIRE CONFIANZA PRO · Asahi Intecc Co., Ltd. · Aug 3, 2004
- K043422 — ASAHI PTCA GUIDE WIRE, J SHAPE SERIES · Asahi Intecc Co., Ltd. · Jan 12, 2005