← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K241510

# CROSSLEAD Tracker (K241510)

_Asahi Intecc Co., Ltd. · DQX · Aug 14, 2024 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K241510

## Device Facts

- **Applicant:** Asahi Intecc Co., Ltd.
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** Aug 14, 2024
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

This product is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. Do not use the guide wire in neurovascular.

## Device Story

CROSSLEAD Tracker is a steerable guide wire used to direct catheters to specific anatomical locations within the peripheral vasculature. Operated by physicians during diagnostic or interventional procedures, the device acts as a trackable pathway for catheter delivery. It features a hybrid Ni-Ti and stainless-steel core wire, a Pt-Ni inner coil, and a Pt-Ni/stainless-steel outer coil, with a hydrophilic distal coating and hydrophobic proximal coating. The device is provided sterile via ethylene oxide. By providing radiopacity for visualization and a flexible tip for navigation, it facilitates precise catheter placement, potentially improving procedural success in peripheral vascular interventions.

## Clinical Evidence

Bench testing only. Performance data included tensile strength, torque strength, torqueability, tip flexibility, coating integrity, catheter compatibility, kink resistance, radio-detectability, and dimensional verification. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous irritation, systemic toxicity, USP rabbit pyrogen, hemolysis, partial thromboplastin time, in vivo thromboresistance, and Sc5b-9 complement activation. All tests met acceptance criteria.

## Technological Characteristics

Guide wire, 0.014 inch (0.36mm) diameter, 100-300cm lengths. Materials: Ni-Ti and stainless-steel core, Pt-Ni/stainless-steel coils, Ag-Sn solder. Hydrophilic distal/hydrophobic proximal coatings. Sterilization: Ethylene Oxide. Mechanical navigation via steerable tip.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Regalia XS 1.0 ([K083146](/device/K083146.md)/[K163426](/device/K163426.md))

## Reference Devices

- MINAMO ([K190176](/device/K190176.md))
- ASAHI Gaia Next ([K192599](/device/K192599.md))
- ASAHI Gladius ([K150445](/device/K150445.md)/[K163426](/device/K163426.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

August 14, 2024

Asahi Intecc Co., Ltd. Cynthia Valenzuela Director, Quality Systems, Regulatory Affairs/Compliance 3002 Dow Avenue Suite 212 Tustin, California 92780

Re: K241510

Trade/Device Name: CROSSLEAD Tracker Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQX Dated: July 15, 2024 Received: July 15, 2024

Dear Cynthia Valenzuela:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/7 description: The image shows the name Lydia S. Glaw -S. Next to the name is the text 'Digitally signed by Lydia S. Glaw -S Date: 2024.08.14 16:22:50 -04'00'.' The text indicates that the name is a digital signature. The date of the signature is August 14, 2024.

Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices 

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known) K241510

Device Name CROSSLEAD Tracker

Indications for Use (Describe)

This product is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. Do not use the guide wire in neurovascular.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K241510

# 510(k) Summary (as required by 21 CFR § 807.92) ЛЯЛНІ ІNTECC CO.,LTD.

# Global Headquarters and R&D Center

3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan

TEL:+81-561-48-5551 FAX:+81-561-48-5552

http://www.asahi-intecc.co.jp/

### CROSSLEAD Tracker

#### 510(k)_K241510

| Date Prepared:         | 27 MAY 2024                                                                                                                                                                                                   |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant:             | ASAHI INTECC CO., LTD.<br>3-100 Akatsuki-cho,<br>Seto, Aichi 489-0071 Japan                                                                                                                                   |
| Contact:               | C. Cynthia Valenzuela<br>Director, Quality and Regulatory Affairs<br>ASAHI INTECC USA, Inc.<br>3002 Dow Avenue, Suite 212<br>Tustin, CA 92780<br>Tel: (949) 745-1617<br>e-mail: cynthiav@asahi-intecc-us-.com |
| Trade Name:            | CROSSLEAD Tracker                                                                                                                                                                                             |
| Device Classification: | Class 2 per 21 CFR § 870.1330                                                                                                                                                                                 |
| Classification Name:   | Catheter, Guide, Wire                                                                                                                                                                                         |
| Product Code:          | DQX - Catheter Guide Wire                                                                                                                                                                                     |
| Predicate Device:      | Regalia XS 1.0 K083146/K163426                                                                                                                                                                                |
| Reference Devices:     | MINAMO K190176<br>ASAHI Gaia Next K192599<br>ASAHI Gladius K150445/K163426                                                                                                                                    |

### INTENDED USE/INDICATIONS FOR USE:

This product is designed to direct a catheter to the desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. Do not use the guide wire in neurovascular.

{5}------------------------------------------------

## DEVICE DESCRIPTION:

The CROSSLEAD Tracker is steerable quide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 100cm, 200cm, 235cm and 300cm length. This Guide Wire consists of a hybrid Ni-Ti and stainless-steel core wire with a Pt-Ni inner coil and Pt-Ni and stainless-steel outer coil is soldered to the core wire with Ag-Sn solder. The coil has radiopacity to achieve visibility and can be made to bend easily with the vessel curve. A hydrophilic coating is applied to the distal portion of the guide wire. A hydrophobic coating is applied to proximal portion. The basic structure, construction, and material of the CROSSLEAD Tracker are similar to that previously described in the predicate Regalia XS 1.0 (K083146/K163426) and reference devices MINAMO (K190176), ASAHI Gaia Next(K192599) and ASAHI Gladius (K150445/K163426).

{6}------------------------------------------------

## COMPARISON WITH PREDICATE AND REFERENCE DEVICES:

Comparisons of the CROSSLEAD Tracker and predicate / reference devices show that the technological characteristics of the subject device such as the components, design, materials, sterilization method, shelf life and operating principle are similar to the currently marketed predicate and reference devices. The intended use of the subject device and its predicates are the same.

| Name of Devices                 | CROSSLEAD Tracker                                                                                                                                                                                                               | Regalia XS 1.0                                                                                                                                                                                                                           | MINAMO                                                                                                                                                                                                                                                                                     | Gaia Next                                                                                                                                                                                                                                                                                                                                                                                                                      | ASAHI Gladius                                                                                                                                                                                                                               |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                 | Subject                                                                                                                                                                                                                         | Predicate                                                                                                                                                                                                                                | Reference                                                                                                                                                                                                                                                                                  | Reference                                                                                                                                                                                                                                                                                                                                                                                                                      | Reference                                                                                                                                                                                                                                   |
| 510(k)                          | TBD                                                                                                                                                                                                                             | K083146,K163426                                                                                                                                                                                                                          | K190176                                                                                                                                                                                                                                                                                    | K192599                                                                                                                                                                                                                                                                                                                                                                                                                        | K150445,K163426                                                                                                                                                                                                                             |
| Intended Use and<br>Indications | This product is designed to<br>direct a catheter to the<br>desired anatomical location<br>in the peripheral vasculature<br>during diagnostic or<br>interventional procedures.<br>Do not use the guide wire in<br>neurovascular. | ASAHI Peripheral Guide<br>Wires are intended to<br>facilitate the placement and<br>exchange of diagnostic and<br>therapeutic devices during<br>intravascular procedures.<br>This device is intended for<br>peripheral vascular use only. | PCI Guide Wires are intended<br>to facilitate the placement of<br>balloon dilatation catheters<br>during percutaneous coronary<br>intervention (PCI) and<br>percutaneous transluminal<br>angioplasty (PTA).<br>The PCI Guide Wires are not<br>intended for use in the<br>neurovasculature. | ASAHI PTCA Guide Wires<br>are intended to facilitate the<br>placement of balloon dilatation<br>catheters during percutaneous<br>transluminal coronary<br>angioplasty (PTCA) and<br>percutaneous transluminal<br>angioplasty (PTA), including<br>use in crossing or assisting in<br>crossing de novo coronary<br>chronic total occlusions (CTO)<br>The ASAHI PTCA Guide<br>Wires are not to be used in<br>the neurovasculature. | ASAHI Peripheral Guide<br>Wires are intended to<br>facilitate the placement and<br>exchange of diagnostic and<br>therapeutic devices during<br>intravascular procedures.<br>This device is intended for<br>peripheral vascular use<br>only. |
| Nominal OD                      | 0.36mm (0.014inch)                                                                                                                                                                                                              | 0.36mm (0.014inch)                                                                                                                                                                                                                       | 0.36mm (0.014inch)                                                                                                                                                                                                                                                                         | 0.36mm (0.014inch)                                                                                                                                                                                                                                                                                                                                                                                                             | 0.46mm(0.018inch)<br>0.36mm(0.014inch)                                                                                                                                                                                                      |
| Overall Length                  | 100, 200, 235,300 cm                                                                                                                                                                                                            | 180, 300cm                                                                                                                                                                                                                               | 190, 300cm                                                                                                                                                                                                                                                                                 | 190 cm                                                                                                                                                                                                                                                                                                                                                                                                                         | 200cm, 235cm,300cm                                                                                                                                                                                                                          |
| Outer Coil                      | Platinum-Nickel and<br>Stainless Steel                                                                                                                                                                                          | Platinum-Nickel and<br>Stainless Steel                                                                                                                                                                                                   | Platinum-Nickel and<br>Stainless Steel                                                                                                                                                                                                                                                     | Platinum-Nickel and<br>Stainless Steel                                                                                                                                                                                                                                                                                                                                                                                         | Platinum-Nickel and<br>Stainless Steel                                                                                                                                                                                                      |

{7}------------------------------------------------

| Name of Devices      | CROSSLEAD Tracker                      | Regalia XS 1.0                         | MINAMO                                 | Gaia Next                              | ASAHI Gladius                          |
|----------------------|----------------------------------------|----------------------------------------|----------------------------------------|----------------------------------------|----------------------------------------|
| Tapered Core<br>Wire | Nitinol and<br>Stainless Steel         | Stainless Steel                        | Nitinol and<br>Stainless Steel         | Stainless Steel                        | Stainless Steel                        |
| Inner Coil           | Platinum-Nickel and<br>Stainless Steel |                                        | Stainless Steel                        | Platinum-Nickel and<br>Stainless Steel | Stainless Steel                        |
| Tip Shape            | Straight,<br>Pre-shape                 | Straight<br>J-tip                      | Straight<br>Pre-shape<br>J-tip         | Pre-shape                              | Straight<br>Pre-shape                  |
| Coating              | Hydrophilic<br>Hydrophobic             | Hydrophilic<br>Hydrophobic             | Hydrophilic<br>Hydrophobic             | Hydrophilic<br>Hydrophobic             | Hydrophilic<br>Hydrophobic             |
| Sterilization        | Provided sterile via<br>Ethylene Oxide | Provided sterile via<br>Ethylene Oxide | Provided sterile via<br>Ethylene Oxide | Provided sterile via<br>Ethylene Oxide | Provided sterile via<br>Ethylene Oxide |

The following technological differences exist between the subject and predicate devices:

- Coil structure
- Core shaft

{8}------------------------------------------------

# NON-CLINICAL TESTING/PERFORMANCE DATA:

Non-clinical laboratory testing was performed on the CROSSLEAD Tracker to determine substantial equivalence. The following testing/assessments were performed:

- Tensile Strength ●
- Torque Strength ●
- Torqueability ●
- Tip Flexibility ●
- Coating Integrity ●
- Catheter Compatibility ●
- Visual Inspection ●
- Corrosion Resistance
- Kink Resistance ●
- Radio - Detectability
- Dimensional Verification
- Coating Integrity / Acute Particulate Characterization ●

The in vitro bench tests demonstrated the CROSSLEAD Tracker met all acceptance criteria and performed similarly to the predicate and reference devices Performance data demonstrates the device functions as intended and has a safety and effectiveness profile that is similar to the predicate and reference devices.

## BIOCOMPATIBILITY:

Testing was performed to assess biocompatibility of the modified coating material. The following tests were performed:

- Cytotoxicity .
- Sensitization
- Intracutaneous Irritation ●
- Systemic Toxicity ●
- USP Rabbit Pyrogen, Material Mediated ●
- Hemolysis ●
- Partial Thromboplastin Time .
- In Vivo Thromboresistance .
- . Sc5b-9 Complement Activation

{9}------------------------------------------------

# CONCLUSION:

The CROSSLEAD Tracker has the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principles as the predicate and reference devices. Performance data demonstrates that the device functions as intended.

Therefore, the CROSSLEAD Tracker is substantially equivalent to the predicate and reference devices.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K241510](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K241510)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
