The BNLE Hydrophilic Jacketed Peripheral Guidewires are intended to direct a catheter to a desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions.
Device Story
Hydrophilic-coated guidewire; nickel-titanium alloy core; tungsten-infused polyurethane jacket. Used to direct catheters to specific peripheral vascular locations during diagnostic/interventional procedures. Operated by physicians in clinical settings. Device provides mechanical support and navigation through vasculature; hydrophilic coating reduces friction. Benefits include improved trackability and navigation. EO sterilized; supplied in dispenser hoop and Tyvek pouch.
Clinical Evidence
Bench testing only. Performance testing included visual inspection, dimensional analysis, device compatibility, tensile strength, torqueability, torque strength, kink resistance, flex resistance, fracture resistance, tip flexibility, radiopacity, particulate evaluation, coating integrity, and lubricity/durability. Biocompatibility testing per ISO 10993-1 included cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemolysis, complement activation, thrombogenicity (canine), and chemical characterization. All acceptance criteria met.
Technological Characteristics
Core: Nickel-titanium alloy. Jacket: Tungsten-infused polyurethane. Coating: Diphenylmethane diisocyanate and Polyvinylpyrrolidone. Dimensions: 150-260 cm length, 0.018-0.035 inch diameter. Sterilization: Ethylene Oxide. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients requiring catheter guidance to anatomical locations within the peripheral vasculature during diagnostic or interventional procedures. Not for neurovascular or coronary use.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
May 14, 2024
Image /page/0/Picture/1 description: The image shows the name Brandie Moore, followed by the title Regulatory Affairs Specialist II. The text is in a simple, sans-serif font. The text is aligned to the left.
Canyon Labs 16217 S Bringhurst Dr., Suite 600 Bluffdale, Utah 84065
Re: K232662
Trade/Device Name: BNLE Hydrophilic Jacketed Peripheral Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: April 30, 2024 Received: April 30, 2024
Dear Brandie Moore:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medical-devices/ device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-andcontinuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Samuel G. Raben -S
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K232662
Device Name
BNLE Hydrophilic Jacketed Peripheral Guidewire
Indications for Use (Describe)
The BNLE Hydrophilic Jacketed Peripheral Guidewires are intended to direct a catheter to a desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Biomerics NLE, along with their address. The logo features a blue molecular-like graphic on the left, followed by the company name in bold, blue letters. Below the name is the address: 10351 Xylon Avenue N., Suite 160, Brooklyn Park, MN 55445.
### 510(k) SUMMARY
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
### Submitter:
Brandie Moore Regulatory Affairs Specialist II 16217 S Bringhurst Dr., Suite 600 Bluffdale, UT 84065 Phone: (801) 243-0534 Email: Brandie.Moore@canyonlabs.com
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
#### Manufacturer:
Biomerics NLE 10351 Xylon Avenue N., Suite 160 Brooklyn Park, MN 55445
#### DATE PREPARED: 31 August 2023
### NAME OF MEDICAL DEVICE:
| Proprietary Name: | BNLE Hydrophilic Peripheral Jacketed Guidewire |
|----------------------|------------------------------------------------|
| Common/Usual Name: | Guidewire |
| Classification Name: | Wire, Guide, Catheter |
#### DEVICE CLASSIFICATION:
| Classification Panel: | Cardiovascular |
|-----------------------|-----------------|
| Regulatory Class: | Class II |
| Product Code: | DQX |
| Regulation Number: | 21 CFR 870.1330 |
### PREDICATE DEVICE:
| Proprietary Name: | Terumo Radifocus Glidewire |
|----------------------|----------------------------|
| Common/Usual Name: | Guidewire |
| Classification Name: | Wire, Guide, Catheter |
| 510(k) Number: | K152740 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Biomerics. The logo consists of a blue abstract design on the left, followed by the word "BIOMERICS" in a bold, dark blue font. Below the word "BIOMERICS" is a horizontal line, followed by the letters "XNLE" in a smaller, lighter blue font.
## DEVICE DESCRIPTION:
The BNLE Hydrophilic Jacketed Peripheral Guidewire is available in 150 cm - 260 cm lengths and in 0.018" - 0.035" diameter ranges. The hydrophilic coated guidewire consists of a nickel-titanium alloy core wire and tungsten infused polyurethane jacket in configurations of wire stiffness and tip shape. The finished guidewire is placed in a dispenser hoop and Tyvek pouch. Five (5) Tyvek pouches are placed into a shelf box with an IFU, and 26 shelf cartons are placed into a shipper box. Labels are placed on the Tyvek pouch and shelf box. The finished product is EO sterilized.
## INTENDED USE/INDICATION FOR USE:
Intended Use: The BNLE Hydrophilic Jacketed Peripheral Guidewires are intended to direct a catheter to a desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures.
Indications for Use: The BNLE Hydrophilic Jacketed Peripheral Guidewires are intended to direct a catheter to a desired anatomical location in the peripheral vasculature during diagnostic or interventional procedures. This device is not intended for neurovascular or coronary interventions.
| Feature | Terumo Radifocus<br>Glidewire K152740<br>(Predicate Device) | BNLE Hydrophilic<br>Jacketed Peripheral<br>Guidewire<br>(Subject Device) | Discussion |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Class | II | II | Identical |
| Product Code | DQX | DQX | Identical |
| Regulation<br>(FDA) | 21 CFR 870.1330 | 21 CFR 870.1330 | Identical |
| Indications<br>for Use | The Glidewire is designed<br>to direct a catheter to the<br>desired anatomical<br>location in the peripheral<br>vasculature during<br>diagnostic or<br>interventional procedures.<br>This device is not intended<br>for neurovascular or<br>coronary interventions. | The BNLE Hydrophilic<br>Jacketed Peripheral<br>Guidewires are intended to<br>direct a catheter to a desired<br>anatomical location in the<br>peripheral vasculature during<br>diagnostic or interventional<br>procedures. This device is not<br>intended for neurovascular or<br>coronary interventions. | Identical |
| Wire<br>Diameter | 0.035" | 0.018" - 0.035" | Substantially<br>Equivalent, within<br>current legally<br>marketed brackets<br>for this type of<br>device. |
| Length of<br>Wire | 260, 300, 350, 400, and<br>450cm | 150, 180, 260cm | Substantially<br>Equivalent, within<br>current legally |
| Feature | Terumo Radifocus<br>Glidewire K152740<br>(Predicate Device) | BNLE Hydrophilic<br>Jacketed Peripheral<br>Guidewire<br>(Subject Device) | Discussion |
| | | | marketed brackets<br>for this type of<br>device. |
| Distal Tip<br>shape | Straight, Angled | Straight, Angled, and J-Tip | Substantially<br>Equivalent, within<br>brackets of<br>predicate device |
| Wire Type | Standard or Stiff | Standard or Stiff | Identical |
| Lengths of<br>the Flexible<br>Tip | 3 and 5cm | 3 cm | Substantially<br>Equivalent, within<br>brackets of<br>predicate device |
| Jacket<br>Material | Tungsten and<br>Polyurethane | Tungsten and Polyurethane | Identical |
| Wire Coating | Hydro gel: Hydrophilic<br>polymer (half-ester methyl<br>vinyl ether-maleic<br>anhydride copolymer)<br>Under coat: Polyvinyl<br>chloride | Hydrophilic polymer:<br>Diphenylmethane diisocyanate<br>and Polyvinylpyrrolidone | Though the<br>materials are<br>different<br>performance<br>testing<br>demonstrates that<br>we meet our<br>design<br>requirements and<br>performance is<br>substantially<br>equivalent to the<br>predicate, Terumo<br>Radifocus<br>Glidewire |
| Core Wire | Nickel-titanium alloy | Nickel-titanium alloy | Identical |
| Guidewire<br>Inserter | Polyethylene | High Density Polyethylene<br>(HDPE) | Substantially<br>Equivalent, both<br>are biocompatible<br>for their intended<br>use per ISO 10993 |
| Packaging | • Individual package<br>• Unit box<br>• Shipping carton | • Individual package<br>• Unit box<br>• Shipping carton | Identical |
| Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide | Identical |
| Shelf Life | 2 years | 1 year | Performance<br>testing<br>demonstrates that<br>we meet our<br>design<br>requirements and |
| Feature | Terumo Radifocus<br>Glidewire K152740<br>(Predicate Device) | BNLE Hydrophilic<br>Jacketed Peripheral<br>Guidewire<br>(Subject Device) | Discussion |
| | | | perform<br>substantially<br>equivalent to the<br>predicate, Terumo<br>Radifocus<br>Glidewire |
#### TECHNOLOGICAL COMPARISON TO PREDICATE DEVICE:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Biomerics NLS. The logo consists of a blue abstract design on the left, followed by the word "BIOMERICS" in bold, dark blue letters. Below "BIOMERICS" is "NLS" in a lighter blue color with a line extending from the left side of the N to the right side of the S.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Biomerics NLE. The logo consists of a blue molecular-like graphic on the left, followed by the word "BIOMERICS" in bold, dark blue letters. Below "BIOMERICS" is "NLE" in a smaller font, underlined by two blue lines.
#### PERFORMANCE TESTING
The BNLE Hydrophilic Peripheral Jacketed Guidewire was thoroughly tested and verifies that it performs as designed and is suitable for its intended use.
Performance Testing included the following:
- Visual Inspection ●
- Dimensional ●
- Device Compatibility ●
- Tensile
- Torqueability ●
- Torque Strength
- . Kink Resistance
- Flex Resistance
- Fracture Resistance
- Tip Flexibility
- Radiopacity
- Particulate Evaluation / Coating Integrity ●
- Lubricity and Durability ●
Biocompatibility per ISO 10993-1 for an external communicating device, limited (<24 hour) blood contacting device.
- · Cytotoxicity L929 MEM Elution
- Maximum Sensitization
- · Irritation Intracutaneous Reactivity
- · Systemic Toxicity
- · Material Mediated Pyrogen
- Hemolysis Extract and Direct Contact
- Complement Activation Assay
- Thrombogenicity in Canine
- · Chemical Characterization
BNLE Hydrophilic Peripheral Jacketed Guidewire met all predetermined acceptance criteria and raised no new concerns regarding safety or efficacy.
BNLE Hydrophilic Peripheral Jacketed Guidewire
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Biomerics. The logo consists of a blue abstract shape on the left, followed by the word "BIOMERICS" in a bold, dark blue font. Below the word "BIOMERICS" is the text "XNLE" in a lighter blue font with a line above it.
# CONCLUSION:
The BNLE Hydrophilic Peripheral Jacketed Guidewire has been demonstrated to be substantially equivalent in design, materials, sterilization, principles of operation, performance, and indications for use to the predicate, Terumo Radifocus Glidewire (K152740).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.