K193571 · Merit Medical System, Inc. · DQX · Apr 8, 2020 · Cardiovascular
Device Facts
Record ID
K193571
Device Name
Go2Wire Guide Wire
Applicant
Merit Medical System, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Apr 8, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Merit GO2WIRE™ is intended to facilitate the placement and exchange of devices during peripheral diagnostic and interventional procedures. Carefully read all instructions prior to use. Observe all warnings and cautions noted throughout these instructions. Failure to do so may result in complications.
Device Story
The Go2Wire Guide Wire is a sterile, single-use device used by clinicians in peripheral diagnostic and interventional procedures. It facilitates the placement and exchange of catheters and other interventional devices within the vascular anatomy. The device consists of a stainless-steel core wire, a PTFE-coated coil, a white PTFE sleeve, a platinum-tungsten marker coil, and an optional extension system connector. The distal tip is shapeable and radiopaque to allow for visualization under fluoroscopy. A torque device is provided to assist the clinician in steering the wire through the vasculature. The device is available in various lengths and tip configurations (floppy, standard, intermediate/modified J). The clinician manually manipulates the wire to navigate the vascular path, providing a track for subsequent device delivery. The extension system allows for length adjustment to accommodate catheter exchanges. The device benefits the patient by enabling precise, minimally invasive access to peripheral vessels.
Clinical Evidence
Bench testing only. No clinical data. Testing included size designation, radiodetectability, surface, tensile strength, torque strength/ability, tip flexibility, fracture/flex testing, lubricity, corrosion resistance, tip shape, catheter/needle compatibility, extension system validation, and particulate testing. Biocompatibility testing included cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, genotoxicity, hemolysis, thrombogenicity, and complement activation. All results met predetermined acceptance criteria.
Technological Characteristics
Materials: stainless-steel core, PTFE-coated coil, white PTFE sleeve, platinum-tungsten marker coil. Dimensions: 0.035" diameter; lengths 145-300cm. Tip: shapeable, radiopaque, straight or modified J. Sterilization: Ethylene oxide (ISO 11135:2014). Biocompatibility: ISO 10993-1:2018. Standards: ISO 11070:2014, ASTM F1980-07.
Indications for Use
Indicated for patients undergoing peripheral diagnostic and interventional procedures requiring device placement and exchange.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 8, 2020
Merit Medical System, Inc. Michael O'Sullivan Senior Regulatory Affairs Specialist Parkmore Business Park West Galway, Ireland
# Re: K193571
Trade/Device Name: Go2Wire Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: March 5, 2020 Received: March 9, 2020
Dear Michael O'Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Interventional Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K193571
Device Name Go2Wire Guide Wire
Indications for Use (Describe)
The Merit GO2WIRE™ is intended to facilitate the placement and exchange of devices during peripheral diagnostic and interventional procedures.
Carefully read all instructions prior to use. Observe all warnings and cautions noted throughout these instructions. Failure to do so may result in complications.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K193571
Merit Medical Systems, Inc.
Merit Go2Wire Guide Wire
Traditional Premarket Notification 510(k)
#
| | Submitter Name: | Merit Medical Systems, Inc. |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Address: | 1600 West Merit Parkway<br>South Jordan, UT 84095 |
| | Telephone Number: | (+353) 91 703700 (Ext. 3061) |
| | Fax Number: | (+353) 91 680104 |
| | Contact Person: | Mark Mullaney |
| | Registration Number: | 1721504 |
| General<br>Provisions | Correspondent Name: | Merit Medical Ireland Ltd. |
| | Address: | Parkmore Business Park<br>Parkmore, Galway, Ireland |
| | Telephone Number: | (+353) 91 703700 (Ext. 3223) |
| | Fax Number: | (+353) 91 680104 |
| | Contact Person: | Michael O'Sullivan |
| | Date of Preparation: | 20 December 2019 |
| | Registration Number: | 9616662 |
| | Subject<br>Device | Trade Name: |
| | Common/Usual Name: | Guide Wire |
| | Classification Name: | Wire, Guide, Catheter |
| Predicate<br>Device | Trade Name: | Wholey Guide Wire System |
| | Classification Name: | 21 CFR 870.1330 Catheter Guide Wire |
| | Premarket Notification: | K120863 |
| | Manufacturer: | Covidien LLC. |
| Classification | | Class II<br>21 CFR § 870.1330<br>Product code: DQX<br>Division of Cardiovascular Devices |
| Intended Use | | The Merit GO2WIRE™ is intended to facilitate the placement and<br>exchange of devices during peripheral diagnostic and<br>interventional procedures.<br>Carefully read all instructions prior to use. Observe all warnings<br>and cautions noted throughout these instructions. Failure to do so<br>may result in complications. |
| | The Merit Go2Wire guide wire will consist of 0.035" guide wire<br>configurations available in 145, 175, 210, 260, and 300cm. The<br>distal tip flex configurations are Floppy, Standard and Intermediate<br>(Modified J). The distal tip shapes are either straight or modified J. | |
| Device<br>Description | The wire is composed of a stainless-steel core wire, a PTFE coated<br>coil, a white PTFE sleeve, a platinum-tungsten marker coil, and an<br>optional extension system connector. The wire's proximal end is<br>covered by a white PTFE sleeve that terminates at 100cm from the<br>distal tip. The distal 100 cm of the wire is covered with a green PTFE<br>pre-coated coil. The PTFE coated coil and platinum-tungsten coil<br>are welded to the core on the very distal tip. The distal tip is<br>shapeable and radiopaque. A torque device is included to facilitate<br>wire steering within the vascular anatomy. | |
| | An 0.035" extension wire consisting of PTFE coated stainless steel<br>155cm in length to extend the length of the wire to allow initial<br>usage with a short wire and then extendable to allow catheter<br>exchanges when required. | |
| | Comparison<br>to Predicate | The Technological characteristics of the subject Merit Go2Wire<br>Guide Wire are substantially equivalent to those of Predicate Device,<br>the Wholey Guide Wire System[K120863]. The Wholey Guide wire<br>was fully characterized by Merit and an equivalent wire was<br>developed by Merit based upon this characterization. The subject<br>device has the same basic design as the predicate, in that they are<br>composed of a stainless-steel core wire, a PTFE coated coil, a white<br>PTFE sleeve, a platinum-tungsten marker coil, and an optional<br>extension system connector. |
| The fundamental technology and operating principles of the subject<br>and the predicates are the same and while the indications for use<br>wordings are not identical, the intended use is the same, with all<br>wires used for the placement and exchange of devices during<br>peripheral diagnostic and interventional procedures. | | |
| Safety &<br>Performance<br>Tests | No performance standards have been established under section<br>514 of the Food, Drug and Cosmetic Act for these devices. A battery<br>of testing was conducted, on the Merit Go2Wire Guide Wire, in<br>accordance with protocols based on requirements outlined in<br>guidance's and industry standards and these were shown to meet<br>the acceptance criteria that were determined to demonstrate<br>substantial equivalence. | |
{4}------------------------------------------------
{5}------------------------------------------------
Where appropriate, the tests were based on the requirements of the following documents:
- FDA Guidance "Coronary, Peripheral, and Neurovascular ● Guidewires -Performance Tests and Recommended Labelling- Guidance for Industry and Food and Drug Administration Staff" - October 2019
- FDA Guidance "Intravascular Catheters. Wires, and ● Delivery Systems with Lubricious Coatings - Labeling Considerations" - October 2019
- FDA guidance Coronary and Cerebrovascular Guide Wire ● Guidance January 1995.
- ISO 11070:2014, Sterile Single-Use Intravascular Catheter ● Introducers.
- ISO 11135:2014 Sterilization of health care products-● Ethylene oxide-: Requirements for the development, validation and routine control of a sterilization process for medical devices.
- ASTM F1980-07 Standard Guide for Accelerated Aging of . Sterile Barrier Systems for Medical Devices
- ISO 10993-1:2018, Biological Evaluation of Medical Devices ● Part 1: Evaluation and Testing within a risk management process, and the FDA Modified ISO 10993 Test Profile FDA Memo G95-1.
The following is a list of all testing that was successfully completed:
Performance Testing-Bench
- Size Designation ●
- Radiodetectability
- Surface ●
- Tensile Strength ●
- Torque Strength .
- Torqueability .
- Tip Flexibility ●
- Fracture test ●
- Flex test ●
- Torturous Path Lubricity ●
- Corrosion Resistance
- Tip Shape Testing Validation .
- Catheter and Needle Compatibility Validation 0
- Guidewire Clinical Use in Model Validation .
{6}------------------------------------------------
- Extension System Clinical Use in Model Validation ●
- Particulate .
## Biocompatibility
- Cytotoxicity .
- Sensitization .
- Irritation .
- Acute Systemic Toxicity .
- Pyrogenicity .
- Genotoxicity ●
- Hemolysis .
- Thrombogenicity .
- Complement Activation .
All test results were comparable to the predicate devices and the subject Merit Go2Wire Guide Wire met the predeterminded acceptance criteria. This has demonstrated that the subject device is substantially equivalent to the predicate devices.
Based on the comparisons noted, the subject Merit Go2Wire Guide Summary of Wire meets the requirements that are considered essential for its Substantial intended use and is substantively equivalent to the Predicate Equivalence Device, the Wholey Guide Wire System [K120863] manufactured by Covidien LLC
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.