← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K173052

# Heraeus TMW Guidewire (K173052)

_Heraeus Medical Components, LLC · DQX · Oct 20, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K173052

## Device Facts

- **Applicant:** Heraeus Medical Components, LLC
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** Oct 20, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Heraeus TMW Guidewire is intended to facilitate the introduction and placement of catheters and interventional devices during Percutaneous Transluminal Angioplasty (PTA) and/or Percutaneous Transluminal Coronary Angioplasty (PTCA).

## Device Story

Heraeus TMW Guidewire is a steerable, disposable, bi-metal core guidewire used to facilitate catheter placement during PTA/PTCA. Device features Triton Nitinol distal core, stainless steel proximal core, hydrophilic coating, and 3 cm radiopaque distal tip. Operated manually by clinicians in clinical settings. Provides mechanical support and guidance for interventional devices; aids in navigating vasculature. Benefits include precise catheter delivery and procedural support. No software or automated processing involved.

## Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-1), particulate testing, corrosion resistance, dimensional inspection, sterilization validation, package integrity, tensile strength, tip stiffness, torque response, and radiopacity. All tests met predetermined acceptance criteria.

## Technological Characteristics

Bi-metal core guidewire; distal core: Triton Nitinol alloy; proximal core: PTFE-coated stainless steel. Features: hydrophilic coating, 3 cm radiopaque distal tip. Dimensions: 0.014" diameter, 180-300 cm length. Sterilization: Ethylene Oxide (ISO 10993-7). Manual operation; no energy source or software.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Hi-Torque Balance Middleweight Guidewire ([K152709](/device/K152709.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 20, 2017

Heraeus Medical Components, LLC Margaret Batchelder Principal Regulatory Specialist 2605 Fernbrook Lane North, Suite J Plymouth, Minnesota 55447

Re: K173052

Trade/Device Name: Heraeus TMW Guidewires Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: September 27, 2017 Received: September 28, 2017

Dear Margaret Batchelder:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K173052

Device Name Heraeus TMW Guidewire

Indications for Use (Describe)

The Heraeus TMW Guidewire is intended to facilitate the introduction and placement of catheters and interventional devices during Percutaneous Transluminal Angioplasty (PTA) and/or Percutaneous Transluminal Coronary Angioplasty (PTCA).

Type of Use (Select one or both, as applicable)

| <label><input checked="true" type="checkbox"/> Research Use (Part 21 CFR 601 Subpart D)</label> | <label><input type="checkbox"/> Compassionate Use (21 CFR 601 Subpart E)</label> |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary -- K173052 6

| Submitter:                                        | Heraeus Medical Components, LLC<br>1805 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447                                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                   | Margaret Batchelder, Principal Regulatory Specialist<br>1805 Fernbrook Lane North, Suite J<br>Plymouth, MN 55447<br>763-225-6675 (voice)<br>763-559-7676 (fax)                                                                                                                                                                                                                                                                                                                |
| Summary Preparation<br>Date                       | September 27, 2017                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Name                                       | Heraeus TMW Guidewire                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device<br>Classification:                         | Common Name: Guidewire<br>Classification Name: Catheter, Guidewire Device Classification: Class II, 21 CFR §870.1330<br>Product Code: DQX                                                                                                                                                                                                                                                                                                                                     |
| Intended Use:                                     | The Heraeus TMW Guidewire is intended to facilitate the introduction<br>and placement of catheters and interventional devices during<br>Percutaneous Transluminal Angioplasty (PTA) and/or Percutaneous<br>Transluminal Coronary Angioplasty (PTCA).                                                                                                                                                                                                                          |
| Contraindications:                                | The guidewire is not intended for use in the cerebral vasculature.                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Description                                | Heraeus TMW Guidewire is a steerable guidewire constructed of a bi-<br>metal core wire.<br>The distal core is made of Triton™ Nitinol alloy and has a coiled wire<br>feature at the distal end. The proximal core is comprised of PTFE<br>coated Stainless Steel material.<br>The wire is available in .014" diameter and lengths ranging from 180<br>to 300 cm. The distal segment is coated with hydrophilic coating and<br>is shapeable with a 3 cm radiopaque distal tip. |
| Predicate Device:                                 | The subject device is substantially equivalent to the Hi-Torque<br>Balance Middleweight Guidewire (K152709, 10/21/2015)<br>manufactured by Abbott Vascular.                                                                                                                                                                                                                                                                                                                   |
| Principle of<br>Operation:                        | The Heraeus TMW Guidewire is operated by a manual process.                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Comparison of<br>Technological<br>Characteristics | The key technological and performance similarities examined between<br>the approved devices and the proposed Heraeus TMW Guidewire<br>device are as follows:                                                                                                                                                                                                                                                                                                                  |
|                                                   | Indications for use - The Indications for use for the proposed device is<br>a subset of indications for use of the predicate device as proposed<br>device,                                                                                                                                                                                                                                                                                                                    |
|                                                   | Fundamental scientific technology, including design are equivalent to<br>the predicate devices                                                                                                                                                                                                                                                                                                                                                                                |
|                                                   | Operating principle - equivalent to the predicate devices                                                                                                                                                                                                                                                                                                                                                                                                                     |
|                                                   | Packaging materials - equivalent to the predicate device                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                                   | Sterility assurance level and method of sterilization - equivalent to<br>the predicate devices                                                                                                                                                                                                                                                                                                                                                                                |
|                                                   | The length and diameter of the device are equivalent to the<br>dimensions of the predicate devices.                                                                                                                                                                                                                                                                                                                                                                           |
|                                                   | The proposed device and that of the predicate device are identical in<br>that they are constructed with equivalent materials to provide<br>equivalent tip stiffness, torque responses, and coating properties.                                                                                                                                                                                                                                                                |
|                                                   | The proposed device has equivalent tip stiffness characteristics to that<br>the predicate device                                                                                                                                                                                                                                                                                                                                                                              |
| Substantial<br>Equivalence                        | The Heraeus TMW Guidewire System is substantially equivalent to<br>the substantially equivalent to the Hi-Torque Balance Middleweight<br>Guidewire (K152709). Substantial equivalence, which is summarized<br>in the following table, is based on indications for use, physical and<br>technological characteristics, and comparative device testing.                                                                                                                         |

{4}------------------------------------------------

{5}------------------------------------------------

|                                | TMW Guidewire<br>(current submission)                                                                                                                                                                                                                                  | Abbott Vascular<br>Hi-Torque Balance<br>Middleweight Universal II<br>Guidewire (K152709)                                                                                                                                                                                                           | Abbott Vascular<br>Hi-Torque Balance<br>Middleweight Elite<br>Guidewire (K152709)                                                                                                                                                                                                                     |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Common/Usual<br>Name | Catheter Guide Wire                                                                                                                                                                                                                                                    | Catheter Guide Wire                                                                                                                                                                                                                                                                                | Catheter Guide Wire                                                                                                                                                                                                                                                                                   |
| Device Class                   | Class II                                                                                                                                                                                                                                                               | Class II                                                                                                                                                                                                                                                                                           | Class II                                                                                                                                                                                                                                                                                              |
| Product Code /<br>Regulation   | DQX / 21 CFR 870.1330                                                                                                                                                                                                                                                  | DQX / 21 CFR 870.1330                                                                                                                                                                                                                                                                              | DQX / 21 CFR 870.1330                                                                                                                                                                                                                                                                                 |
| Regulation<br>Name             | Catheter Guide Wire                                                                                                                                                                                                                                                    | Catheter Guide Wire                                                                                                                                                                                                                                                                                | Catheter Guide Wire                                                                                                                                                                                                                                                                                   |
| Prescription Use               | Rx Only                                                                                                                                                                                                                                                                | Rx Only                                                                                                                                                                                                                                                                                            | Rx Only                                                                                                                                                                                                                                                                                               |
| Indications for<br>Use         | The Heraeus TMW<br>Guidewire is intended to<br>facilitate the introduction and<br>placement of catheters and<br>interventional devices during<br>Percutaneous Transluminal<br>Angioplasty (PTA) and/or<br>Percutaneous Transluminal<br>Coronary Angioplasty<br>(PTCA). | To facilitate the placement of<br>balloon dilatation catheters<br>during percutaneous<br>transluminal coronary<br>angioplasty (PTCA) and<br>percutaneous transluminal<br>angioplasty (PTA). This guide<br>wire may also be used with<br>compatible stent devices<br>during therapeutic procedures. | To facilitate the placement of<br>balloon dilatation catheters<br>during percutaneous<br>transluminal coronary<br>angioplasty (PTCA) and<br>percutaneous transluminal<br>angioplasty (PTA). This<br>guide wire may also be used<br>with compatible stent devices<br>during therapeutic<br>procedures. |
| Guidewire OD                   | 0.014"                                                                                                                                                                                                                                                                 | 0.014"                                                                                                                                                                                                                                                                                             | 0.014"                                                                                                                                                                                                                                                                                                |
| Guidewire<br>Length            | 180 & 300 cm                                                                                                                                                                                                                                                           | 190 & 300 cm                                                                                                                                                                                                                                                                                       | 190 & 300 cm                                                                                                                                                                                                                                                                                          |
| Guidewire<br>Materials         | Nitinol Cobalt, Stainless<br>Steel, Pebax, Platinum<br>Tungsten, Solder, and<br>Adhesive                                                                                                                                                                               | Nitinol, Stainless Steel,<br>Radiopaque Coil, Solder,<br>Polymer Jacket, Adhesive                                                                                                                                                                                                                  | Nitinol, Stainless Steel,<br>Radiopaque Coil, Solder,<br>Adhesive                                                                                                                                                                                                                                     |
| Guidewire<br>Coating           | PTFE, Hydrophilic coating                                                                                                                                                                                                                                              | PTFE, Hydrophilic coating                                                                                                                                                                                                                                                                          | PTFE, Hydrophilic coating                                                                                                                                                                                                                                                                             |
| Tip Shape                      | Straight Tip                                                                                                                                                                                                                                                           | Straight, J-Tip                                                                                                                                                                                                                                                                                    | Straight, J-Tip                                                                                                                                                                                                                                                                                       |
| Sterile Device?                | Yes                                                                                                                                                                                                                                                                    | Yes                                                                                                                                                                                                                                                                                                | Yes                                                                                                                                                                                                                                                                                                   |
| Sterilization<br>Type          | Ethylene Oxide                                                                                                                                                                                                                                                         | Irradiation                                                                                                                                                                                                                                                                                        | Irradiation                                                                                                                                                                                                                                                                                           |
| EO Sterilization<br>Residuals  | Per ISO 10993-7                                                                                                                                                                                                                                                        | N/A                                                                                                                                                                                                                                                                                                | N/A                                                                                                                                                                                                                                                                                                   |
| Disposable /<br>Reusable       | Disposable                                                                                                                                                                                                                                                             | Disposable                                                                                                                                                                                                                                                                                         | Disposable                                                                                                                                                                                                                                                                                            |

{6}------------------------------------------------

- Performance Testing: In vitro bench tests were utilized to demonstrate equivalence with reference to the FDAs guidance document "Coronary and Cerebrovascular Guidewire Guidance, Jan 1995".
The performance testing assessment supports that the biocompatibility, shelf life, and functional specifications of the proposed Guidewire device were met.

The Heraeus TMW Guidewire device test data supports the claims of substantial equivalence to the predicate devices. Biological Safety of the predicate device has been established through biocompatibility testing carried out in compliance with ISO10993-1:2009 and G95-1, FDA General Program Memorandum: Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.

The following bench tests were conducted or evaluated to support the proposed device:

- . biocompatibility testing (including: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Pyrogen Testing, Hemolysis, Complement Activation, In-Vivo Thrombogenicity)
- particulate testing ●
- corrosion resistance
- dimensional inspection ●
- sterilization ●
- sterile package integrity testing
- tensile strength ●
- tip stiffness ●
- torque response ●
- corrosion resistance ●
- . radiopacity
- . tip stiffness

The Heraeus TMW Guidewire System met all predetermined acceptance criteria and compared favorably with the predicate device.

#### Conclusion: Heraeus considers the Heraeus TMW Guidewire to be equivalent to the predicate device. This conclusion is based upon the fact that device has an equivalent intended use, and there are no differences that raise new types of questions of safety and effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K173052](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K173052)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
