The S-Wire Guidewire System is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used in trans-catheter aortic valve implantation (TAVI) procedures.
Device Story
S-Wire Guidewire System is a sterile, single-use 0.035" diameter, 260 cm long guidewire. It features a nickel-titanium alloy core with PTFE coating and a 304V stainless steel distal coil with silicone coating. The distal end has a pre-formed double curve (360° bend, ~3 cm diameter) with a small S-curve at the tip to provide an atraumatic shape. Used in clinical settings (e.g., cath labs) by physicians to facilitate the introduction and placement of interventional devices within the aorta and heart chambers during TAVI procedures. Includes a 9 Fr disposable plastic introducer to aid loading into intravascular catheters. The device provides a stable track for interventional tools, potentially improving procedural success and patient outcomes during cardiac interventions.
Clinical Evidence
Bench testing only. No clinical data. Testing included dimensional inspection, tensile strength, distal loop compression force, distal tip shape retention, tip stiffness, coating integrity, particulate evaluation, catheter compatibility, flex testing, fracture testing, and corrosion resistance per FDA guidance and ISO 11070. Biocompatibility testing (cytotoxicity, irritation, sensitization, acute systemic toxicity, hemocompatibility, pyrogen) and sterile package integrity testing were performed per ISO 10993. Results met all predetermined acceptance criteria.
Technological Characteristics
Nickel-titanium superelastic alloy core; 304V stainless steel distal coil; PTFE shaft coating; silicone distal coil coating. 0.035" diameter; 260 cm length. Pre-formed double-curve distal tip. Ethylene Oxide sterilization (ISO 10993-7). Includes 9 Fr plastic introducer. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients undergoing interventional procedures within the heart chambers, including trans-catheter aortic valve implantation (TAVI).
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Lake Region Pre-Formed Guidewire (Boston Scientific Safari Pre-Shaped TAVI Guidewire) (K130798, K151244)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/13 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three figures in profile facing right.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 18, 2016
Concert Medical, LLC % Pamela Papineau President Delphi Medical Device Consulting, Inc. 5 Whitcomb Ave Ayer, Massachusetts 01432
Re: K153397
Trade/Device Name: S-Wire Guidewire System Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DOX Dated: November 17, 2015 Received: November 24, 2015
Dear Ms. Papineau:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
M. A. Hilleman
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153397
Device Name S-Wire Guidewire System
Indications for Use (Describe)
The S-Wire Guidewire System is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used in trans-catheter aortic valve implantation (TAVI) procedures.
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - Revised (Clean Copy)
| Date Prepared: | 15 July 2016 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------|
| General Information | |
| Owner's Name: | Concert Medical, LLC |
| Address: | 77 Accord Park Drive<br>Norwell, MA 02061 |
| Telephone: | 781-871-7882 |
| Fax: | 781-724-4555 |
| Contact Person: | Pamela Papineau, RAC (US, EU, CAN)<br>Delphi Medical Device Consulting, Inc.<br>5 Whitcomb Avenue<br>Ayer, MA 01432 |
| Telephone: | (978) 772-3552 |
| Fax: | (978) 796-5460 |
| Subject Device: | |
| Trade Name: | S-Wire Guidewire System |
| Common Name: | Catheter Guide Wire |
| Product Code: | DQX |
| FDA Regulation: | 21 CFR 870.1330 - Catheter Guide Wire |
| Device Classification: | Class II |
| Predicate Device: | |
| Product Name: | Lake Region Pre-Formed Guidewire (Boston Scientific Safari<br>Pre-Shaped TAVI Guidewire) |
| Common Name: | Catheter Guide Wire |
| Product Code: | DQX |
| FDA Regulation: | 21 CFR 870.1330 - Catheter Guide Wire |
| Device Classification: | Class II |
| Premarket Notification: | K130798, K151244 |
| Indications for Use | |
The S-Wire Guidewire System is intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used in trans-catheter aortic valve implantation (TAVI) procedures
### Device Description:
The S-Wire Guidewire System is a sterile, single-use guidewire used for the placement of interventional devices within the aorta and the chambers of the heart. The S-Wire has a 0.035" diameter, and an overall length of 260 cm. The S-Wire construction consists of a nickel-titanium alloy core wire that is PTFE coated, ground and thermally formed. The S-Wire distal coil is
{4}------------------------------------------------
made of 304V stainless steel: the distal and proximal ends of the stainless steel coil is bonded to the core wire. The distal portion of the S-Wire has silicone coating. The distal end of the S-Wire has a pre-formed double curve shape consisting of an approximately 360° bend with an outer diameter of approximately 3 cm, and a small S-curve at the distal tip of the wire that is fully contained within the 3 cm curved area. The S-Wire Guidewire System includes a disposable plastic 9 Fr introducer (6 inches long) that is used as an aid in loading the S-Wire into an intravascular catheter.
## Substantial Equivalence
The Concert Medical S-Wire Guidewire System is substantially equivalent to the Lake Region Pre-Formed Guidewire (K151244 & K130798), which is marketed by Boston Scientific Corp. as the Safari Pre-Shaped TAVI guidewire. Substantial equivalence, which is summarized below, is based on indications for use, physical and technological characteristics, and comparative device testing.
| | Concert Medical S-Wire Guidewire<br>System (current submission) | Lake Region Pre-Formed Guidewire<br>(K151244 & K130798) |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Common/Usual Name | Catheter Guide Wire | Catheter Guide Wire |
| Device Class | Class II | Class II |
| Product Code / Regulation | DQX / 21 CFR 870.1300 | DQX / 21 CFR 870.1300 |
| Regulation Name | Catheter Guide Wire | Catheter Guide Wire |
| Prescription Use | Rx Only | Rx Only |
| Indications for Use | To facilitate the introduction and<br>placement of interventional devices<br>within the chambers of the heart,<br>including those used in trans-catheter<br>aortic valve implantation (TAVI)<br>procedures | To facilitate the introduction and<br>placement of interventional devices<br>within the chambers of the heart,<br>including those used within trans-<br>catheter aortic valve procedures |
| Guidewire OD | 0.035" | 0.035" |
| Guidewire Length | 260 cm | 260 cm – 300 cm |
| Guidewire Materials | Nickel-titanium superelastic alloy,<br>304V stainless steel, adhesive | A313 stainless steel |
| Guidewire Coating | PTFE (shaft); Silicone (distal coil) | PTFE |
| Tip Shape | Atraumatic double curve | Atraumatic double curve |
| Sterile Device? | Yes | Yes |
| Sterilization Type | Ethylene Oxide | Ethylene Oxide |
| EO Sterilization Residuals | Per ISO 10993-7 | Per ISO 10993-7 |
| Disposable / Reusable | Disposable | Disposable |
## Performance Testing:
Performance testing for the Concert Medical S-Wire Guidewire System consists of the measurement of physical characteristics and device integrity in accordance with FDA's Coronary and Cerebrovascular Guidewire Guidance and ISO 11070 (Sterile, single-use intravascular catheter introducers); testing included dimensional inspection, tensile strength (guidewire and introducer), distal loop compression force, distal tip shape retention, tip stiffness, coating integrity, particulate evaluation, catheter compatibility, guidewire flex testing, guidewire fracture testing and corrosion resistance. Where applicable, comparative test results are provided
{5}------------------------------------------------
for the S-Wire and the predicate device. Additional testing included in this Premarket Notification includes ISO 10993 biocompatibility testing (cytotoxicity, irritation, sensitization, acute systemic toxicity, hemocompatibility and pyrogen) and sterile package integrity testing (pouch seal tensile strength and dye penetration). The Concert Medical S-Wire Guidewire System met all predetermined acceptance criteria and compared favorably with the predicate device.
## Conclusion:
The Concert Medical S-Wire Guidewire System has been demonstrated to be substantially equivalent to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.