To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). This guide wire may also be used with compatible stent devices during therapeutic procedures.
Device Story
Steerable coronary guide wires (0.014" diameter) used to facilitate placement of balloon dilatation catheters during PTCA/PTA procedures; may also support compatible stent delivery. Operated by physicians in clinical settings. Device features include extendable lengths (via DOC Guide Wire Extension), shapeable straight or pre-shaped 'J' distal tips, and various lubricious coatings (hydrophilic/hydrophobic/PTFE/SMOOTHGLIDE) to reduce friction. Modification involves a change in the adhesive formulation used to join core subassemblies. Performance verified via bench testing (tensile strength/hypotube pull tests) and biocompatibility testing of the new adhesive.
Clinical Evidence
Bench testing only. No clinical data presented. Performance evaluated via in vitro bench testing (tensile strength/hypotube pull tests) and biocompatibility testing of the modified adhesive formulation.
Technological Characteristics
Steerable coronary guide wires; 0.014" diameter; 175cm, 190cm, 300cm lengths. Materials include core subassemblies joined by modified adhesive. Coatings: Hydrocoat (hydrophilic), PTFE, hydrophobic, SMOOTHGLIDE. Distal tips: straight shapeable or pre-shaped 'J'. DOC extendable. Sterilization not specified.
Indications for Use
Indicated for patients undergoing percutaneous transluminal coronary angioplasty (PTCA) or percutaneous transluminal angioplasty (PTA) requiring balloon dilatation catheter placement; may be used with compatible stent devices.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 21, 2015
Abbott Vascular Jochen Reich Senior Regulatory Affairs Specialist 26531 Ynez Road Temecula, California 92591
Re: K152709
Trade/Device Name: Hi-Torque Balance, Balance Middleweight, Balance Heavyweight, Balance Trek, Balance Middleweight Universal, Balance Middleweight Universal II, Balance Middleweight Elite Guide wires Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DOX Dated: September 18, 2015 Received: September 21, 2015
Dear Jochen Reich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hillebrand
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): _K152709
Device Name: HI-TORQUE BALANCE Guide Wires and HI-TORQUE BALANCE Guide Wires with Hydrocoat Hydrophilic Coating
Indications for Use:
To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
{3}------------------------------------------------
510(k) Number (if known): _K152709
Device Name: HI-TORQUE BALANCE TREK Guide Wires with Hydrocoat Hydrophilic Coating
Indications for Use:
To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
Prescription Use _____________________________________________________________________________________________________________________________________________________________ x (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
{4}------------------------------------------------
510(k) Number (if known): _K152709
Device Name: HI-TORQUE BALANCE HEAVYWEIGHT Guide Wires with Hydrocoat Hydrophilic Coating
Indications for Use:
To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
{5}------------------------------------------------
510(k) Number (if known): _K152709
Device Name: HI-TORQUE BALANCE MIDDLEWEIGHT Guide Wires and HI-TORQUE BALANCE MIDDLEWEIGHT Guide Wires with Hydrocoat Hydrophilic Coating
Indications for Use:
To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{6}------------------------------------------------
510(k) Number (if known): _K152709
Device Name: HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL Guide Wires with Hydrophilic Coating
Indications for Use:
To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
{7}------------------------------------------------
510(k) Number (if known): _K152709
Device Name: HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II Guide Wires
Indications for Use:
To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). This guide wire may also be used with compatible stent devices during therapeutic procedures.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
{8}------------------------------------------------
510(k) Number (if known): K152709
Device Name: HI-TORQUE BALANCE MIDDLEWEIGHT ELITE Guide Wires
Indications for Use:
To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). This guide wire may also be used with compatible stent devices during therapeutic procedures.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
{9}------------------------------------------------
### 510(k) Summary
# A) HI-TORQUE BALANCE Guide Wires
The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
| 1. Submitter's Name | Abbott Vascular |
|---------------------------------|-------------------------------------------------|
| 2. Submitter's Address | 26531 Ynez Road, Temecula, CA 92591 |
| 3. Telephone | 0049-15165621396 |
| 4. Fax | (951) 914-0339 |
| 5. Contact Person | Jochen Reich |
| 6. Date Prepared | September 28, 2015 |
| 7. Device Trade Name | HI-TORQUE BALANCE |
| 8. Device Common Name | Coronary Guide Wire |
| 9. Device Classification Name | Catheter Guide Wire (DQX) |
| 10. Predicate Device Name | HI-TORQUE BALANCE Guide Wire (K925381. |
| | cleared February 12, 1993), HI-TORQUE |
| | BALANCE Guide Wire with Hydrocoat |
| | Hydrophilic Coating, (K973494, cleared December |
| | 12, 1997) and HI-TORQUE BALANCE Guide |
| | Wire. (K021228, cleared May 15, 2002) |
| | |
#### 11. Device and Change Description
The HI-TORQUE BALANCE Guide Wires are available in a 0.014" outer diameter, in 175 cm and 190 cm extendable lengths, and a 300 cm exchange length. The proximal end of the 175 cm and 190 cm models are tapered to fit into the hypotube portion of the DOC Guide Wire Extension. This enables the physician to extend the working length of this guide wire to facilitate catheter exchanges. The distal tips of the guide wires are available either as a straight tip that is shapeable or as a pre shaped "J". The change being made is in the formulation of an adhesive that is used to adhere two core subassemblies.
#### 12. Indication for Use
The HI-TORQUE BALANCE Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
#### 13. Technological Characteristics
{10}------------------------------------------------
In vitro bench testing performance evaluations demonstrated that the HI-TORQUE BALANCE Guide Wires met the acceptance criteria and performed comparable to the predicate devices. No new safety or effectiveness issues were raised during the testing and therefore, the HI-TORQUE BALANCE Guide Wire Family may be considered substantially equivalent to the predicate devices.
Bench testing and Biocompatibility testing was conducted and the results confirm that the device with the modified adhesive formulation remains appropriate for its intended use.
As for bench testing the tensile strength to join the distal core of the guidewire (where the tip coil is located) with the proximal core of the guidewire was conducted. The hypotube pull test verified that the hypotube joint is sufficiently strong to withstand normal tensile loading during the use of the Guide Wire in a clinical procedure.
The modified adhesive formulation was tested for biocompatibility as per the below listed biocompatibility tests:
- a. Cytotoxicity
- b. Hemolysis, direct
- c. Hemolysis, indirect
- d. Complement Activation
- e. Coagulation (PT and PTT)
- f. Sensitization
- g. Intracutaneous Toxicity (Irritation)
- h. Acute Systemic Toxicity
- Pyrogenicity Material Mediated (Rabbit) i.
{11}------------------------------------------------
# B) HI-TORQUE BALANCE TREK Guide Wires
The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
| 1. Submitter's Name | Abbott Vascular |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submitter's Address | 26531 Ynez Road, Temecula, CA 92591 |
| 3. Telephone | 0049-15165621396 |
| 4. Fax | (951) 914-0339 |
| 5. Contact Person | Jochen Reich |
| 6. Date Prepared | September 28, 2015 |
| 7. Device Trade Name | HI-TORQUE BALANCE TREK |
| 8. Device Common Name | Coronary Guide Wire |
| 9. Device Classification Name | Catheter Guide Wire (DQX) |
| 10. Predicate Device Name | HI-TORQUE BALANCE TREK Guide Wire<br>(K991152, cleared April 29, 1999) and HI-TORQUE BALANCE TREK Guide Wire,<br>(K021228, cleared May 15, 2002) |
# 11. Device Description
The HI-TORQUE BALANCE TREK Guide Wires are available in a 0.014" outer diameter, and in an 190 cm extendable length and a 300 cm exchange length. The proximal end of the 190 cm models are tapered to fit into the hypotube portion of the DOC Guide Wire Extension. This enables the physician to extend the working length of this guide wire to facilitate catheter exchanges. The distal tips of the guide wires are available either as a straight tip that is shapeable or as a pre shaped "J". The change being made is in the formulation of an adhesive that is used to adhere two core subassemblies.
12. Indication for Use
The HI-TORQUE BALANCE TREK Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
#### 13. Technological Characteristics
{12}------------------------------------------------
In vitro bench testing performance evaluations demonstrated that the HI-TORQUE BALANCE TREK Guide Wires met the acceptance criteria and performed comparable to the predicate devices. No new safety or effectiveness issues were raised during the testing and therefore, the HI-TORQUE BALANCE TREK Guide Wire Family may be considered substantially equivalent to the predicate devices.
Bench testing and Biocompatibility testing was conducted and the results confirm that the device with the modified adhesive formulation remains appropriate for its intended use.
As for bench testing the tensile strength to join the distal core of the guidewire (where the tip coil is located) with the proximal core of the guidewire was conducted. The hypotube pull test verified that the hypotube joint is sufficiently strong to withstand normal tensile loading during the use of the Guide Wire in a clinical procedure.
The modified adhesive formulation was tested for biocompatibility as per the below listed biocompatibility tests:
- a. Cytotoxicity
- b. Hemolysis, direct
- c. Hemolysis, indirect
- d. Complement Activation
- e. Coagulation (PT and PTT)
- f. Sensitization
- g. Intracutaneous Toxicity (Irritation)
- h. Acute Systemic Toxicity
- Pyrogenicity Material Mediated (Rabbit) i.
{13}------------------------------------------------
# C) HI-TORQUE BALANCE HEAVYWEIGHT Guide Wires
The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
| 1. Submitter's Name | Abbott Vascular |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submitter's Address | 26531 Ynez Road, Temecula, CA 92591 |
| 3. Telephone | 0049-15165621396 |
| 4. Fax | (951) 914-0339 |
| 5. Contact Person | Jochen Reich |
| 6. Date Prepared | September 28, 2015 |
| 7. Device Trade Name | HI-TORQUE BALANCE HEAVYWEIGHT |
| 8. Device Common Name | Coronary Guide Wire |
| 9. Device Classification Name | Catheter Guide Wire (DQX) |
| 10. Predicate Device Name | HI-TORQUE BALANCE HEAVYWEIGHT<br>Guide Wire with Hydrocoat Hydrophilic Coating<br>(K982083, cleared September 11, 1998) and HI-<br>TORQUE BALANCE HEAVYWEIGHT Guide<br>Wire, (K021228, cleared May 15, 2002) |
## 11. Device Description
The HI-TORQUE BALANCE HEAVYWEIGHT Guide Wires are available in a 0.014" outer diameter, in 175 cm and 190 cm extendable lengths, and a 300 cm exchange length. The proximal end of the 175 cm and 190 cm models are tapered to fit into the hypotube portion of the DOC Guide Wire Extension. This enables the physician to extend the working length of this guide wire to facilitate catheter exchanges. The distal tips of the guide wires are available either as a straight tip that is shapeable or as a pre shaped "J". The change being made is in the formulation of the adhesive that is used to adhere two core subassemblies.
#### 12. Indication for Use
The HI-TORQUE BALANCE HEAVYWEIGHT Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
#### 13. Technological Characteristics
{14}------------------------------------------------
In vitro bench testing performance evaluations demonstrated that the HI-TORQUE BALANCE HEAVYWEIGHT Guide Wires met the acceptance criteria and performed comparable to the predicate devices. No new safety or effectiveness issues were raised during the testing and therefore, the HI-TORQUE BALANCE HEAVYWEIGHT Guide Wire Family may be considered substantially equivalent to the predicate devices.
Bench testing and Biocompatibility testing was conducted and the results confirm that the device with the modified adhesive formulation remains appropriate for its intended use.
As for bench testing the tensile strength to join the distal core of the guidewire (where the tip coil is located) with the proximal core of the guidewire was conducted. The hypotube pull test verified that the hypotube joint is sufficiently strong to withstand normal tensile loading during the use of the Guide Wire in a clinical procedure.
The modified adhesive formulation was tested for biocompatibility as per the below listed biocompatibility tests:
- a. Cytotoxicity
- b. Hemolysis, direct
- c. Hemolysis, indirect
- d. Complement Activation
- e. Coagulation (PT and PTT)
- Sensitization f.
- g. Intracutaneous Toxicity (Irritation)
- h. Acute Systemic Toxicity
- Pyrogenicity Material Mediated (Rabbit) i.
{15}------------------------------------------------
# D) HI-TORQUE BALANCE MIDDLEWEIGHT Guide Wires
The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
| 1. Submitter's Name | Abbott Vascular |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Submitter's Address | 26531 Ynez Road, Temecula, CA 92591 |
| 3. Telephone | 0049-15165621396 |
| 4. Fax | (951) 914-0339 |
| 5. Contact Person | Jochen Reich |
| 6. Date Prepared | September 28, 2015 |
| 7. Device Trade Name | HI-TORQUE BALANCE MIDDLEWEIGHT |
| 8. Device Common Name | Coronary Guide Wire |
| 9. Device Classification Name | Catheter Guide Wire (DQX) |
| 10. Predicate Device Name | HI-TORQUE BALANCE MIDDLEWEIGHT<br>Guide Wire (K971815, cleared July 9, 1997), HI-<br>TORQUE BALANCE MIDDLEWEIGHT Guide<br>Wire with Hydrocoat Hydrophilic Coating<br>(K973494, cleared December 12, 1997) and HI-<br>TORQUE BALANCE MIDDLEWEIGHT Guide<br>Wire, (K021228, cleared May 15, 2002) |
## 11. Device Description
The HI-TORQUE BALANCE MIDDLEWEIGHT Guide Wires are available in a 0.014" outer diameter, 175 cm and 190 cm extendable lengths, and a 300 cm exchange length. The proximal end of the 175 cm and 190 cm models are tapered to fit into the hypotube portion of the DOC Guide Wire Extension. This enables the physician to extend the working length of this guide wire to facilitate catheter exchanges. The distal tips of the guide wires are available either as a straight tip that is shapeable or as a pre shaped "J". The change being made is in the formulation of the adhesive that is used to adhere two core subassemblies.
# 12. Indication for Use
The HI-TORQUE BALANCE MIDDLEWEIGHT Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
#### 13. Technological Characteristics
{16}------------------------------------------------
In vitro bench testing performance evaluations demonstrated that the HI-TORQUE BALANCE MIDDLEWEIGHT Guide Wires met the acceptance criteria and performed comparable to the predicate devices. No new safety or effectiveness issues were raised during the testing and therefore. the HI-TORQUE BALANCE MIDDLEWEIGHT Guide Wire Family may be considered substantially equivalent to the predicate devices.
Bench testing and Biocompatibility testing was conducted and the results confirm that the device with the modified adhesive formulation remains appropriate for its intended use.
As for bench testing the tensile strength to join the distal core of the guidewire (where the tip coil is located) with the proximal core of the guidewire was conducted. The hypotube pull test verified that the hypotube joint is sufficiently strong to withstand normal tensile loading during the use of the Guide Wire in a clinical procedure.
The modified adhesive formulation was tested for biocompatibility as per the below listed biocompatibility tests:
- a. Cytotoxicity
- b. Hemolysis, direct
- c. Hemolysis, indirect
- d. Complement Activation
- e. Coagulation (PT and PTT)
- Sensitization f.
- g. Intracutaneous Toxicity (Irritation)
- h. Acute Systemic Toxicity
- Pyrogenicity Material Mediated (Rabbit) i.
{17}------------------------------------------------
# E) HI-TOROUE BALANCE MIDDLEWEIGHT UNIVERSAL Guide Wires
The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
| 1. | Submitter's Name | Abbott Vascular |
|----|-------------------------------|-----------------------------------------------|
| 2. | Submitter's Address | 26531 Ynez Road, Temecula, CA 92591 |
| | 3. Telephone | 0049-15165621396 |
| | 4. Fax | (951) 914-0339 |
| | 5. Contact Person | Jochen Reich |
| | 6. Date Prepared | September 28, 2015 |
| | 7. Device Trade Name | HI-TORQUE BALANCE MIDDLEWEIGHT |
| | | UNIVERSAL |
| | 8. Device Common Name | Coronary Guide Wire |
| | 9. Device Classification Name | Catheter Guide Wire (DQX) |
| | 10. Predicate Device Name | HI-TORQUE BALANCE MIDDLEWEIGHT |
| | | UNIVERSAL Guide Wire with Hydrocoat |
| | | Hydrophilic Coating (K013833, cleared January |
| | | 16, 2002), HI-TORQUE BALANCE |
| | | MIDDLEWEIGHT UNIVERSAL Guide Wire |
| | | with Hydrocoat Hydrophilic Coating (K031678 |
| | | cleared August 26, 2003) and HI-TORQUE |
| | | BALANCE MIDDLEWEIGHT UNIVERSAL |
| | | Guide Wire, (K101011, cleared May 24, 2010) |
| | | |
# 11. Device Description
The HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL Guide Wire is a steerable guide wire available in a maximum outer diameter of 0.0137" and in lengths of 190 cm and 300 cm. The distal segment of the guide wire, up to the hypotube, is coated with hydrophilic coating to reduce friction for improved guide wire movement within the catheter. The distal tip is offered in a straight shapeable configuration and a pre-shaped "J" configuration. The proximal core has a maximum diameter of 0.0137". The proximal end of the guide wire is coated with PTFE, which reduces friction of the wire within a catheter. The BMW Universal Guide Wire is DOC® extendable in the 190 cm lengths. The change being made is to the formulation of an adhesive that is used to adhere two core subassemblies.
# 12. Indication for Use
The HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA).
{18}------------------------------------------------
## 13. Technological Characteristics
Comparisons of the new and predicate devices show that the technological characteristics such as product performance, materials, design, sterilization, packaging, and intended use are substantially equivalent to the current marketed predicate devices.
## 14. Performance Data
In vitro bench testing performance evaluations demonstrated that the HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL Guide Wires met the acceptance criteria and performed comparable to the predicate devices. No new safety or effectiveness issues were raised during the testing and therefore, the HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL Guide Wire Family may be considered substantially equivalent to the predicate devices.
Bench testing and Biocompatibility testing was conducted and the results confirm that the device with the modified adhesive formulation remains appropriate for its intended use.
As for bench testing the tensile strength to join the distal core of the guidewire (where the tip coil is located) with the proximal core of the guidewire was conducted. The hypotube pull test verified that the hypotube joint is sufficiently strong to withstand normal tensile loading during the use of the Guide Wire in a clinical procedure.
The modified adhesive formulation was tested for biocompatibility as per the below listed biocompatibility tests:
- a. Cytotoxicity
- b. Hemolysis, direct
- c. Hemolysis, indirect
- d. Complement Activation
- e. Coagulation (PT and PTT)
- f. Sensitization
- g. Intracutaneous Toxicity (Irritation)
- h. Acute Systemic Toxicity
- i. Pyrogenicity Material Mediated (Rabbit)
{19}------------------------------------------------
# E) HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II Guide Wires
The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
| 1. Submitter's Name | Abbott Vascular |
|-------------------------------|-------------------------------------------------------------------------------------------|
| 2. Submitter's Address | 26531 Ynez Road, Temecula, CA 92591 |
| 3. Telephone | 0049-15165621396 |
| 4. Fax | (951) 914-0339 |
| 5. Contact Person | Jochen Reich |
| 6. Date Prepared | September 28, 2015 |
| 7. Device Trade Name | HI-TORQUE BALANCE MIDDLEWEIGHT<br>UNIVERSAL II |
| 8. Device Common Name | Coronary Guide Wire |
| 9. Device Classification Name | Catheter Guide Wire (DQX) |
| 10. Predicate Device Name | HI-TORQUE BALANCE MIDDLEWEIGHT<br>UNIVERSAL II Guide (K072460, cleared April 4,<br>2008). |
# 11. Device Description
The HI-TOROUE BALANCE MIDDLEWEIGHT UNIVERSAL II Guide Wire is a steerable guide wire available in a diameter of 0.0140" and in lengths of 190 cm DOC extendable length and a 300 cm exchange length. The distal tip is offered in a straight shapeable configuration and a pre-shaped "J" configuration. The distal segment of the guide wire is coated with a new lubricious coating to improve guide wire movement in the catheter. The proximal end of the guide wire utilizes SMOOTHGLIDE coating technology. The change being made is in the formulation of an adhesive that is used to adhere two core subassemblies.
12. Indication for Use
The HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). This guide wire may also be used with compatible stent devices during therapeutic procedures.
# 13. Technological Characteristics
{20}------------------------------------------------
In vitro bench testing performance evaluations demonstrated that the HI-TORQUE BALANCE MIDDLEWEIGHT UNIVERSAL II Guide Wires met the acceptance criteria and performed comparable to the predicate devices. No new safety or effectiveness issues were raised during the testing and therefore, the HI-TOROUE BALANCE MIDDLEWEIGHT UNIVERSAL II Guide Wire Family may be considered substantially equivalent to the predicate devices.
Bench testing and Biocompatibility testing was conducted and the results confirm that the device with the modified adhesive formulation remains appropriate for its intended use.
As for bench testing the tensile strength to join the distal core of the guidewire (where the tip coil is located) with the proximal core of the guidewire was conducted. The hypotube pull test verified that the hypotube joint is sufficiently strong to withstand normal tensile loading during the use of the Guide Wire in a clinical procedure.
The modified adhesive formulation was tested for biocompatibility as per the below listed biocompatibility tests:
- a. Cytotoxicity
- b. Hemolysis, direct
- c. Hemolysis, indirect
- d. Complement Activation
- e. Coagulation (PT and PTT)
- f. Sensitization
- g. Intracutaneous Toxicity (Irritation)
- h. Acute Systemic Toxicity
- Pyrogenicity Material Mediated (Rabbit) i.
{21}------------------------------------------------
# F) HI-TORQUE BALANCE MIDDLEWEIGHT ELITE Guide Wires
The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.
| 1. Submitter's Name | Abbott Vascular |
|-------------------------------|-------------------------------------------------------------------------------------|
| 2. Submitter's Address | 26531 Ynez Road, Temecula, CA 92591 |
| 3. Telephone | 0049-15165621396 |
| 4. Fax | (951) 914-0339 |
| 5. Contact Person | Jochen Reich |
| 6. Date Prepared | September 28, 2015 |
| 7. Device Trade Name | HI-TORQUE BALANCE MIDDLEWEIGHT<br>ELITE |
| 8. Device Common Name | Coronary Guide Wire |
| 9. Device Classification Name | Catheter Guide Wire (DQX) |
| 10. Predicate Device Name | HI-TORQUE BALANCE MIDDLEWEIGHT<br>ELITE Guide (K103101, cleared February 10, 2011). |
# 11. Device Description
The HI-TORQUE BALANCE MIDDLEWEIGHT ELITE Guide Wire is a steerable guide wire available in a maximum outer diameter of 0.014" and in lengths of 190 cm and 300 cm. The distal segment of the guide wire is coated with hydrophilic coating to reduce friction for improved guide wire movement within the catheter. The distal tip is offered in a straight shapeable configuration and a pre-shaped "J" configuration. The proximal core has a maximum diameter of 0.0145". The proximal end of the guide wire is coated with hydrophobic coating, which reduces friction of the wire within a catheter. The ELITE Guide Wire is DOC® extendable in the 190 cm lengths. The change being made is in the formulation of an adhesive that is used to adhere two core subassemblies.
12. Indication for Use
The HI-TORQUE BALANCE MIDDLEWEIGHT ELITE Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). This guide wire may also be used with compatible stent devices during therapeutic procedures.
# 13. Technological Characteristics
{22}------------------------------------------------
In vitro bench testing performance evaluations demonstrated that the HI-TORQUE BALANCE MIDDLEWEIGHT ELITE Guide Wires met the acceptance criteria and performed comparable to the predicate devices. No new safety or effectiveness issues were raised during the testing and therefore, the HI-TOROUE BALANCE MIDDLEWEIGHT ELITE Guide Wire Family may be considered substantially equivalent to the predicate devices.
Bench testing and Biocompatibility testing was conducted and the results confirm that the device with the modified adhesive formulation remains appropriate for its intended use.
As for bench testing the tensile strength to join the distal core of the guidewire (where the tip coil is located) with the proximal core of the guidewire was conducted. The hypotube pull test verified that the hypotube joint is sufficiently strong to withstand normal tensile loading during the use of the Guide Wire in a clinical procedure.
The modified adhesive formulation was tested for biocompatibility as per the below listed biocompatibility tests:
- a. Cytotoxicity
- b. Hemolysis, direct
- c. Hemolysis, indirect
- d. Complement Activation
- e. Coagulation (PT and PTT)
- f. Sensitization
- g. Intracutaneous Toxicity (Irritation)
- h. Acute Systemic Toxicity
- Pyrogenicity Material Mediated (Rabbit) i.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.