K142051 · Merit Medical Systems, Inc. · DQX · Dec 17, 2014 · Cardiovascular
Device Facts
Record ID
K142051
Device Name
PRELUDE PLASTIC JACKETED GUIDE WIRE
Applicant
Merit Medical Systems, Inc.
Product Code
DQX · Cardiovascular
Decision Date
Dec 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The Merit Prelude Plastic Jacketed introducer guide wire is intended to facilitate the placement of devices during diagnostic and interventional procedures, specifically sheath introducers. The wire is indicated for the peripheral vasculature only.
Device Story
Metallic core wire with radiopaque polymer jacket; facilitates placement of sheath introducers in peripheral vasculature. Used by clinicians in interventional/diagnostic settings. Device provides mechanical guidance for catheter/sheath insertion; aids in navigating peripheral vessels. Benefits patient by enabling safe, accurate placement of interventional devices.
Clinical Evidence
Bench testing only. No clinical data. Testing included mechanical performance (tensile strength, torque, tip flexibility, fracture/flex tests, corrosion resistance) and biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, genotoxicity, hemolysis, thrombogenicity) per ISO 10993 and other standards.
Technological Characteristics
Metallic core wire with radiopaque polymer jacket. Available in straight and angled versions. Sterilization via Ethylene Oxide (ISO 11135-1). Biocompatibility testing per ISO 10993 series and ASTM F756-08. Mechanical testing per ISO 11070:1998 and FDA Coronary and Cerebrovascular Guide Wire Guidance.
Indications for Use
Indicated for patients requiring peripheral vascular diagnostic or interventional procedures to facilitate the placement of sheath introducers.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three overlapping profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
Merit Medical Systems Inc. Mr. Michael O'Sullivan RA Specialist III Parkmore Business Park West Galway, Ireland
Re: K142051
> Trade/Device Name: Prelude Plastic Jacketed Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: November 14, 2014 Received: November 17, 2014
Dear Mr. Michael O'Sullivan,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K142051
Device Name Prelude Plastic Jacketed guide wire
Indications for Use (Describe)
The Merit plastic jacketed introducer guide wire is intended to facilitate the placement of devices during diagnostic and interventional procedures, specifically sheath introducers. The wire is indicated for the peripheral vasculature only,
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Merit Medical Systems, Inc.
Merit Prelude Plastic Jacketed Introducer Guide Wire
Traditional Premarket Notification 510(k)
| ^^?% &\$) %<br>510(k) Summary | | |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Registration Number:<br>Correspondent Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(+353) 91 703700 (Ext. 3061)<br>(+353) 91 680104<br>Mark Mullaney<br>1721504<br>Merit Medical Ireland Ltd.<br>Parkmore Business Park<br>Parkmore, Galway, Ireland<br>(+353) 91 703700 (Ext. 3223)<br>(+353) 91 680104<br>Michael O'Sullivan<br>17th December 2014<br>9616662 |
| Subject<br>Device | Trade Name:<br>Common/Usual Name:<br>Classification Name: | Prelude Plastic jacketed introducer guide wire<br>Guide Wire<br>Wire, Guide, Catheter |
| Predicate<br>Device | Premarket Notification Predicate Device # 1:<br>Terumo Glidesheath(mini guide wire): K082644<br>Manufacturer: Terumo Medical Corporation | |
| Classification | Class II<br>21 CFR § 870.1330<br>Product code: DQX<br>Division of Cardiovascular Devices | |
| Intended Use | The Merit Prelude Plastic Jacketed introducer guide wire is intended to facilitate<br>the placement of devices during diagnostic and interventional procedures,<br>specifically sheath introducers. The wire is indicated for the peripheral<br>vasculature only. | |
| Device<br>Description | The Merit Prelude Plastic jacketed introducer guide wire consists of a high<br>quality metallic core wire with a radiopaque polymer jacket. The wire will be<br>offered in straight and angled versions, in various lengths. | |
| Comparison to<br>Predicate | The Technological characteristics of the subject Merit Prelude Plastic jacketed<br>introducer guide wire are substantially equivalent to those of the mini guide wire<br>predicate, within the Terumo Glidesheath (mini guide wire) Wire [K082644]. Both<br>wires consist of a metallic core wire coated in a radiopaque polymer jacket. | |
| Safety &<br>Performance<br>Tests | No performance standards have been established under section 514 of the<br>Food, Drug and Cosmetic Act for these devices. A battery of testing was<br>conducted in accordance with protocols based on requirements outlined in<br>guidance's and industry standards and these were shown to meet the<br>acceptance criteria that were determined to demonstrate substantial<br>equivalence.<br><br>Where appropriate, the tests were based on the requirements of the following<br>documents:<br>FDA guidance Coronary and Cerebrovascular Guide Wire Guidance<br>January 1995. ISO 11070: 1998, Sterile Single-Use Intravascular Catheter Introducers. ISO 11135-1: 2007 Sterilization of health care products-Ethylene oxide-<br>Part 1: Requirements for the development, validation and routine control<br>of a sterilization process for medical devices. ASTM F1980-07 Standard Guide for Accelerated Aging of Sterile Barrier<br>Systems for Medical Devices ISO 2233:2000, Packaging - Complete, filled transport packages and<br>unit loads - Conditioning for testing ISO 10993-1: 2009, Biological Evaluation of Medical Devices Part 1:<br>Evaluation and Testing within a risk management process, and the FDA<br>Modified ISO 10993 Test Profile FDA Memo G95-1. ANSI/AAMI/ISO 10993-3:2003, Biological Evaluation of Medical Devices<br>- Part 3: Tests for Genotoxicity, Carcinogenicity and Reproductive<br>Toxicity ISO 10993-4:2002 (Amd.1:2006), Biological evaluation of medical<br>devices – Part 4: Selection of tests for interactions with blood ISO 10993-5:2009, Biological evaluation of medical devices – Part 5:<br>Tests for in vitro cytotoxicity ISO 10993-7:2008, Biological evaluation of medical devices – Part 7:<br>Ethylene oxide sterilization residuals ISO 10993-10:2010, Biological evaluation of medical devices – Part 10:<br>Tests for irritation and skin sensitization ISO 10993-11:2006, Biological evaluation of medical devices – Part 11:<br>Tests for systemic toxicity ASTM F756-08, Standard practice for assessment of hemolytic<br>properties of materials United States Pharmacopeia 36, National Formulary 31, <151> Pyrogen<br>Test. 2013 The following is a list of all significant testing that was successfully completed:<br>Performance Testing-Bench Size Designation | |
{4}------------------------------------------------
Merit Medical Systems, Inc.
Merit Prelude Plastic Jacketed Introducer Guide Wire
Traditional Premarket Notification 510(k)
{5}------------------------------------------------
Merit Medical Systems. Inc. Merit Prelude Plastic Jacketed Introducer Guide Wire Traditional Premarket Notification 510(k)
- Radiodetectability
- Surface ●
- Tensile Strength ●
- Torque Strength
- Tip Flexibility ●
- Fracture test .
- Flex test ●
- . IV Catheter Compatibility
- Dilator Catheter Compatibility ●
- Corrosion Resistance ●
- Tip Shape Testing ●
- ETO Residuals ●
- Bioburden ●
- Pyrogen (LAL) ●
## Biocompatibility
- Cytotoxicity ●
- Sensitization
- Irritation ●
- Acute Systemic Toxicity ●
- Pyrogenicity ●
- Genotoxicity
- Hemolysis ●
- Thrombogenicity ●
- Complement Activation .
- Chemical Characterization .
As all test results were comparable to the predicate device and as the subject Merit Prelude Plastic jacketed introducer guide wire met the predeterminded acceptance criteria applicable to the safety and efficacy of the device, this has demonstrated that the subject device is substantially equivalent to the predicate device.
Summary of Substantial Equivalence Based on the Indications for Use, design, safety and performance testing, the subject Merit Prelude Plastic jacketed introducer guide wire meets the requirements that are considered essential for its intended use and is substantively equivalent to the predicate device, the Terumo Glidesheath(mini guide wire) manufactured by Terumo Medical Corporation, K082644.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.