PRE-FORMED GUIDEWIRES

K130798 · Lake Region Medical, Inc. · DQX · Aug 23, 2013 · Cardiovascular

Device Facts

Record IDK130798
Device NamePRE-FORMED GUIDEWIRES
ApplicantLake Region Medical, Inc.
Product CodeDQX · Cardiovascular
Decision DateAug 23, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The Pre-Formed guidewires are intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used within transcatheter aortic valve procedures. This wire is not intended for use in the cerebrovasculature or coronary arteries.

Device Story

Pre-Formed Guidewire is a stainless steel wire (0.035" diameter, 260-300cm length) consisting of a core and a PTFE-coated coil. Distal end features a double-curve tip. Used by physicians in clinical settings to facilitate placement of interventional devices within heart chambers during procedures like transcatheter aortic valve implantation. Device provides structural support and guidance for catheters. Benefits include improved navigation and positioning of interventional tools within cardiac anatomy.

Clinical Evidence

Bench testing and GLP animal study performed. Bench testing included dimensional, tensile strength, tip flexibility, coating adherence, catheter compatibility, and ISO-standard mechanical/corrosion testing. Biocompatibility testing per ISO 10993 (cytotoxicity, irritation, sensitization, hemolysis, thrombogenicity, etc.) confirmed safety. GLP animal study evaluated performance and safety compared to predicate; results supported substantial equivalence.

Technological Characteristics

Materials: Stainless steel (ASTM A313) core and coil; coil coated in Green PTFE. Dimensions: 0.035" diameter, 260-300cm length. Design: Double-curve distal tip; stainless steel weld at proximal/distal ends. Sterilization: Not specified. Connectivity: None (mechanical device).

Indications for Use

Indicated for facilitating introduction and placement of interventional devices within heart chambers, including transcatheter aortic valve procedures. Contraindicated for use in cerebrovasculature or coronary arteries.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image contains the logo for Lake Region Medical. The logo features a circular graphic to the left of the text "Lake Region". Below the word "Region" is the word "Medical" in a smaller font size. The logo is simple and professional. ### PRODUCT: PRE-FORMED GUIDEWIRE SUBMISSION DATE: March 21, 2013 SUBMISSION TYPE: TRADITIONAL 510(k) ### SECTION 5.0: 510(k) SUMMARY | 5.1 | MANUFACTURER / REGISTRATION INFORMATION | | |-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|--| | Lake Region Medical<br>340 Lake Hazeltine Drive<br>Chaska, MN 55318-1029 USA<br><i>FDA REGISTRATION NUMBER:</i> 2126666 | Contact Person: Mathew Pexa<br>Title: Regulatory Specialist II<br>Telephone: 952-641-8511<br>Fax: 952-448-3441 | | #### 5.2 TRADE NAME (PROPRIETATY NAME) Pre-Formed Guidewire DEVICE COMMON NAMES/USUAL NAMES/CLASSIFICATION NAME 5.3 ### CATHETER GUIDEWIRE (DQX) #### CLASS OF DEVICE 5.4 This type of Guidewire was originally listed as a Class II device by the Cardiovascular (DQX) review panel. | 5.5 IDENTIFICATION OF PREDICATE DEVICE(s) ==================================================================================================================================== | | | |--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|------------------------------| | | 510(k) NUMBER MANUFACTURER | DEVICE NAME | | K930622 | Boston Scientific | Amplatz SuperStiff Guidewire | #### 5.6 DEVICE DESCRIPTION The 0.035" diameter, 260cm-300cm length guidewire is composed of two primary wire components: a core and a coil. Both components are made of stainless steel per ASTM A313. The core wire is a stainless steel wire which forms the inner body of the guidewire. The coil component is the guidewire's outer layer and is a stainless steel wire coated in Green Polytetrafluoroethylene (PTFE). The coil and core components are weld together on the distal and proximal ends, forming the guidewire. The distal end of the guidewire contains a double-curve. | OUTSIDE DIAMETER: | 0.035" | |-------------------|---------------| | LENGTHS: | 260cm - 300cm | | TIP SHAPE: | Double-curve | | TIP SIZES: | Small - Large | #### COMPLIANCE WITH APPLICABLE STANDARDS 5.7 The guidewire is in compliance with ISO 10993-1, ISO 10993-4, ISO 10993-7, ISO 10993-7, ISO 10993-10, ISO 10993-11, ISO 10993-12, ISO 15223-1, EN 980, ISO 11135-1, and ISO 11070. AUG 2 3 2013 {1}------------------------------------------------ # C Lake Regio ### PRODUCT: PRE-FORMED GUIDEWIRE SUBMISSION DATE: March 21, 2013 SUBMISSION TYPE: TRADITIONAL 510(k) ## K130798 Page 2 of 3 ### SECTION 5.0: 510(k) SUMMARY #### 5.8 INTENDED USE STATEMENT The Pre-Formed guidewires are intended to facilitate the introduction and placement of interventional devices within the chambers of the heart, including those used within transcatheter aortic valve procedures. #### 5.8 CONTRAINDICATIONS This wire is not intended for use in the cerebrovasculature or coronary arteries. #### 59 COMPARISON Comparison bench and animal tests were completed on the Boston Scientific Amplatz SuperStiff guidewire with 510(k) number K930622 to determine substantial equivalence. #### 5.10 OUALIFICATION TESTING The conclusions drawn from bench testing, biocompatibility, and a GLP animal study demonstrate the device is as safe, as effective and performs at least as safely and effectively as the legally marketed device. ### BENCH TESTING In order to demonstrate equivalence of the guidewire, Lake Region Medical performed testing to establish requirements. Test pieces were tested and inspected according to established requirements for visual/tactile, dimensional and mechanical attributes. Test methods were developed using FDA Coronary and Cerebrovascular Guidewire guidance and ISO 11070:1998. The following table lists the applicable bench tests performed at baseline and aging include: - Dimensional . - . FDA Device Compatibility - FDA Tensile Strength . - . FDA Tip Flexibility - . FDA Coating Adherence/Integrity - FDA Catheter Compatibility . - . Packaging Study - Particulate . - . ISO Visual - . ISO Fracture - ISO Flex . - . ISO Corrosion Resistance - ISO Strength of Union . - . ISO Radiopacity - Body Stiffness ### BIOCOMPATIBILITY TESTING Biocompatibility testing per ISO 10993 series has been performed on the Pre-Formed guidewires and has been found to be acceptable. - Cytotoxicity . - . Kligman Maximization Test - Irritation / Intracutaneous Reactivity . - Complement Activation Assay - Rabbit Pyrogen 0 - Hemolysis - Thrombogenicity - . Lee and White Coagulation - . Unactivated Thromboplastin Time Assay - Acute Systemic Toxicity Test . - Partial Thromboplastin Time Assay . - USP Physicochemical Test . - . Inhibition and Enhancement ### ANIMAL STUDIES A GLP Animal Study was completed to evaluate customer feedback, performance and safety of the Guidewire compared to the currently marketed device. The studies show the guidewires are substantially equivalent to the legally marketed device. {2}------------------------------------------------ # Lake Reg ### PRODUCT: PRE-FORMED GUIDEWIRE SUBMISSION DATE: March 21, 2013 SUBMISSION TYPE: TRADITIONAL 510(k) #### SUBSTANTIAL EQUIVALENCE DATA 5.11 The Pre-Formed guidewires are substantially equivalent to the Amplatz SuperStiff Guidewire manufactured by Boston Scientific cleared under 510(k) number K930622. The guidewire has similar technological characteristics to the predicate device. The technological differences between the proposed device and the predicate device are: - Larger Proximal Core Diameter . - -Longer Distal Grind - . Lack of intermediate joint - -Stainless steel weld (as opposed to Solder) - Double Curve on distal tip - - -Lack of moveable core The data generated by accepted test methods and comparisons to the predicate device show that the Pre-Formed guidewire is substantially equivalent to the predicate device and the technological differences listed above do not pose any new issues of safety or effectiveness. The Pre-Formed guidewires are substantially equivalent to the Amplatz SuperStiff Guidewire cleared under 510(k) K930622. All test results support the claim of substantial equivalence. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002 August 23, 2013 Lake Region Medical C/O Mr. Mathew Pexa Regulatory Specialist II 340 Lake Hazeltine Drive Chaska, MN 55318-1029 Rc: K130798 > Trade/Device Name: Pre-Formed Guidewires Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guidewire Regulatory Class: Class II Product Code: DQX Dated: August 14, 2013 Received: August 14, 2013 Dear Mr. Pexa: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH docs not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (sec above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-rclated adverse events) (21 CFR 803); good manufacturing practice requirements us set {4}------------------------------------------------ Page 2 - Mr. Mathew Pexa forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, M.A. Hillen for for Bram D. Zuckerman, Ph.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ C Lake R PRODUCT: PRE-FORMED GUIDEWIRE SUBMISSION DATE: March 21, 2013 SUBMISSION TYPE: TRADITIONAL 510(k) ## INDICATIONS FOR USE 510(k) NUMBER (IF KNOWN): K130798 DEVICE NAME: PRE-FORMED GUIDEWIRES INDICATIONS FOR USE: The Pre-Formed guidewires are intended to facilitate the introduction and placement of interventional devices within the chambers of the heart including those within transcatheter aortic valve procedures. This guidewire is not intended for use in the cerebrovasculature or coronary arteries. PRESCRIPTION X USE AND/OR OVER-THE-COUNTER USE (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) PAGE 1 of M.A. Zillmann
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...