← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K110241

# ENDOLOGIX GUIDEWIRE (K110241)

_Techdevice Corporation · DQX · Feb 25, 2011 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K110241

## Device Facts

- **Applicant:** Techdevice Corporation
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** Feb 25, 2011
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Guidewire facilitates placement and exchange of catheters and other instruments in the peripheral vasculature. This guidewire is not intended for use in the coronary arteries or neurovasculature.

## Device Story

Endologix Guidewire is a vascular guidewire used to facilitate catheter placement and exchange in peripheral vasculature. Device consists of stainless steel core wire with platinum coil; coil may cover entire core or distal portion only. Surfaces may be uncoated or coated with PTFE spray or jacket. Platinum Tungsten marker bands may be included for visualization. Device is used by clinicians in clinical settings. Performance is verified through bench testing of tensile strength, torque strength, torqueability, tip flexibility, coating integrity, and corrosion resistance. Changes from predicate include vacuum heat annealing process and alternative labeling.

## Clinical Evidence

Bench testing only. Comparative performance testing included tensile strength, torque strength, torqueability, tip flexibility, coating integrity, and corrosion testing.

## Technological Characteristics

Stainless steel core wire; platinum coil; optional PTFE spray or jacket coating; Platinum Tungsten marker bands. Vacuum heat annealing process. Dimensions/form factor not specified. No software or energy source.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Guidewire ([K053251](/device/K053251.md))
- Guidewire ([K053028](/device/K053028.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KIIO241 P. 1 of 2

TechDevice Corporation January 19, 2011

510(k) Premarket Notification (Special) Vascular Guidewire

# 510(k) Summary

| General Company Information                                        |                                            |                  |
|--------------------------------------------------------------------|--------------------------------------------|------------------|
| Name                                                               | TechDevice Corporation                     |                  |
| Address                                                            | 650 Pleasant Street<br>Watertown, MA 02472 | FEB 25 2011      |
| Contact:                                                           | Gary Bunnewith                             |                  |
| Telephone:                                                         | 617-972-5810                               |                  |
| General Device Information                                         |                                            |                  |
| Product Name:                                                      | Endologix Guidewire                        |                  |
| Common Name:                                                       | Guidewire                                  |                  |
| Classification:                                                    | DQX, Catheter guide wire                   | 21 CFR 870. 1330 |
| Predicate Devices                                                  |                                            |                  |
| Guidewire K053251<br>TechDevice Corporation<br>Watertown, MA 02472 |                                            |                  |
| Guidewire K053028<br>TechDevice Corporation<br>Watertown, MA 02472 |                                            |                  |

### Product Description:

The Guidewire is constructed of a stainless steel core wire and a platinum coil welded/brazed in place over the core wire. The coil may cover the entire core wire or just the floppy distal portion. The exposed core wire and or the coil may be uncoated stainless steel or coated with a PTFE spray or a PTFE jacket. Platinum Tungsten marker bands may be placed under the coil to aid in visualization.

### Indications for Use:

The Guidewire facilitates placement and exchange of catheters and other instruments in the peripheral vasculature. This guidewire is not intended for use in the coronary arteries or neurovasculature.

## Safety and Performance:

Substantial equivalence for this device was based on a comparison of labeling, physical and performance design characteristics as compared to the predicate device, as well as on the results of comparative bench testing. Comparative performance testing included:

- Tensile Strength A.
- Torque Strength B.
- C. Torqueability
- . Tip Flexibility D.
- ய் Coating Integrity
- Corrosion Testing F.

{1}------------------------------------------------

510(k) Premarket Notification (Special) Vascular Guidewire

TechDevice Corporation January 19, 2011

# Substantial Equivalence

The proposed guidewires are substantially equivalent to the currently marketed guidewires which were cleared for marketing by FDA under K053251 and K053258. Substantial equivalence for the proposed devices is based on a companson of materials, dimensional specifications, design characteristics, labeling and indications for use. The two guidewires are of the exact same design and the only changes are the heat annealing process going from ambient air to vacuum and alternative labeling (i.e. naming) requested for the two wires. Physical testing has been conducted to confirm the proposed products performance characteristics.

## Conclusion:

. .

Based on the indications for use, technological characteristics, and comparison to predicate devices, the Guidewires have been shown to be safe and effective for its intended use.

{2}------------------------------------------------

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is positioned to the right of the department's name, which is written in a circular fashion around the logo.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

TechDevice Corporation C/O Gary Bunnewith Director of Quality 650 Pleasant Street Watertown, MA 02472

FEB 2 5 2011

Re: K110241

Trade/Device Name: Endologix Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DOX Dated: January 19, 2011 Received: January 27, 2011

Dear Mr. Garv Bunnewith:

We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. Gary Bunnewith

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

[signature]

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K110241

TechDevice Corporation January 19, 2011

510(k) Premarket Notification (Special) Vascular Guidewire

Page

510(k) Number (if known): Kl | 024 |

Device Name: Endologix Guidewire

Indications for Use:

The Guidewire facilitates placement and exchange of catheters and other instruments in the peripheral vasculature. This guidewire is not intended for use in the coronary arteries or neurovasculature.

Prescription Use ਮ (Part 21 CFR 801 Subpart D) Over-the -Counter Use (21 CFR 807 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CORH, Office of Device Evaluation (ODE)

(Division Sign-Oth
Division of Cardiovascular Devices
510(k) Number/K110241

009

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K110241](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K110241)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
