← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K101116

# HI-TORQUE PILOT 50 AND 150 GUIDE WIRE WITH HYDROCOAT HYDROPHILIC COATING, HI-TORQUE WHISPER LS AND MS GUIDE WIRE WITH (K101116)

_Abbott Vascular-Cardiac Therapies · DQX · Jun 23, 2010 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K101116

## Device Facts

- **Applicant:** Abbott Vascular-Cardiac Therapies
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** Jun 23, 2010
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The HI-TORQUE PILOT Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The guide wire is also intended to facilitate the placement of compatible stent devices during therapeutic intravascular procedures.

## Device Story

HI-TORQUE PILOT guide wire (0.014" diameter) facilitates placement of balloon dilatation catheters and stents during PTCA/PTA procedures. Device features varying tip stiffness (50, 150, 200) and distal hydrophilic coating to reduce friction. Proximal section coated with PTFE. Available in 190 cm (extendable) and 300 cm (exchange) lengths with brachial/femoral markers for catheter exit indication. Used by physicians in clinical settings (e.g., cath lab) to navigate vasculature. Output is mechanical guidance for interventional devices; aids in successful lesion access and therapeutic device delivery.

## Clinical Evidence

No clinical data. Evidence consists of in vitro bench testing (coating adherence, integrity/particulate, friction) and biocompatibility testing (cytotoxicity, hemolysis, acute systemic toxicity, complement activation, coagulation, intracutaneous reactivity, USP systemic injection, sensitization, rabbit pyrogen, LAL pyrogen, bacterial endotoxins, and in vivo thrombogenicity).

## Technological Characteristics

0.014" diameter guide wire; PTFE-coated proximal section; polyurethane-covered distal section with hydrophilic coating. Available in 190 cm and 300 cm lengths. Mechanical features include varying tip stiffness and shapeable/pre-shaped J-tips. Biocompatibility per ISO/USP standards. Non-electronic, mechanical device.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- HI-TORQUE PILOT Guide Wire ([K030549](/device/K030549.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K101116

JUN 2 3 2010

### 510(k) Summary

The 510(k) Summary is submitted in accordance with 21 CFR §807.92 and the requirements of the Safe Medical Device Act (SMDA) of 1990.

| 1. Submitter's Name           | Abbott Vascular                                                                  |
|-------------------------------|----------------------------------------------------------------------------------|
| 2. Submitter's Address        | 26531 Ynez Road, Temecula, CA 92591                                              |
| 3. Telephone                  | (951) 914-3242                                                                   |
| 4. Fax                        | (951) 914-0339                                                                   |
| 5. Contact Person             | Kay Setzer                                                                       |
| 6. Date Prepared              | April 16, 2010                                                                   |
| 7. Device Trade Name          | HI-TORQUE PILOTTM 50, 150, and 200 Guide<br>Wire Family with Hydrophilic Coating |
| 8. Device Common Name         | Guide Wire                                                                       |
| 9. Device Classification Name | Catheter Guide Wire (DQX)                                                        |
| 10. Predicate Device Name     | HI-TORQUE PILOT Guide Wire (K030549,<br>cleared May 14, 2003)                    |

#### 11. Device Description

The HI-TORQUE PILOT™ Guide Wire with Hydrophilic Coating is a guide wire with a maximum diameter of 0.0140" and is available in a 190 cm extendable length and a 300 cm exchange length.

There are three HI-TORQUE PILOT Guide Wire designs with varying tip stiffness, the HI-TORQUE PILOT 50, HI-TORQUE PILOT 150, and HI-TORQUE PILOT 200 Guide Wires.

The distal tip of the guide wire is available either as a straight tip that is shapeable, or as a pre-shaped "J". The straight shape allows the physician to shape the tip according to his/her preference; the J shapes provide the physician the convenience of a J shape without manual shaping. Brachial and femoral markers are located on the proximal segment of the 190 cm and 300 cm guide wires to indicate when the tip of the guide wire is about to exit the guide catheter.

The proximal section of the wire is coated with polytetrafluoroethylene (PTFE). The distal, polyurethane-covered area of the wire is covered with a hydrophilic coating.

1

{1}------------------------------------------------

#### 12. Indication for Use

The HI-TORQUE PILOT Guide Wire Family is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The guide wire is also intended to facilitate the placement of compatible stent devices during therapeutic intravascular procedures.

#### 13. Technological Characteristics

Comparisons of the new and predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices. The proposed device is similar in design and materials to the currently marketed product in that its core wire, tip coils, and solders remain the same. The hydrophilic coating has been changed but is identical to previously marketed devices.

#### 14. Performance Data

In vitro bench testing, including coating adherence and integrity (particulate testing), and friction testing were conducted on the subject device. These tests were used to verify that the hydrophilic coating has not impacted safety or effectiveness of the device. Other mechanical tests, such as Tensile Strength, Torqueability, Torque Strength, and Tip Flex were leveraged from the predicate device as the mechanical functionality of the device has not changed. Biocompatibility testing was leveraged from predicate devices with identical materials and manufacturing process. Biocompatibility tests included cytotoxicity, hemolysis, acute systemic toxicity, complement activation, coagulation, intracutaneous (intradermal) reactivity test, USP systemic injection test, sensitization, rabbit pyrogen test, LAL pyrogen, bacterial endotoxins, and in vivo thrombogenicity tests. The in vitro bench tests and the biocompatibility tests demonstrated that the HI-TORQUE PILOT Guide Wire met all acceptance criteria and performed similarly to the predicate devices. No new safety or effectiveness issues were raised during the testing program and therefore, the HI-TORQUE PILOT Guide Wire may be considered substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol on the right side. Encircling the bird is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

# JUN 2 3 2010

Ms. Kay Setzer Senior Regulatory Affairs Associate Abbott Vascular Inc. 26531 Ynez Road Temecula, CA 92591

Re: K101116

Trade Names: HI-TORQUE WHISPER™ LS, MS, and ES Guide Wire Family With Hydrophilic Coating HI-TORQUE PILOT™ 50, 150, and 200 Guide Wire Family with Hydrophilic Coating Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: Class II Product Code: DQX Dated: May 27, 2010 Received: May 28, 2010

Dear Ms. Setzer:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your because of the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association in May 28, 1976, the enactment date of the Medical Device Amendments, or to conimered prior to they 2014 11:14 provisions of the Federal Food, Drug, de rices that have been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, merelors, manov of the Act include requirements for annual registration, listing of general connects profilering practice, labeling, and prohibitions against misbranding and de vices, good manufacturing practices, as the grange with and to contract liability addition. Tious note: ODFET as , that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Kay Setzer

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known): صالا | الا | 10 |

Device Name: HI-TORQUE WHISPER™ LS, MS, and ES Guide Wires with Hydrophilic Coating

Indications for Use:

To facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The wire is also intended to facilitate the placement of compatible stent devices during therapeutic intravascular procedures.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mitchell/Blake for B Zuckerman

(Division Sign-Off) Division of Cardiovascular D 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K101116](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K101116)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
