ASAHI SION PTCA GUIDE WIRE MODEL AHW14R001S & AHW14R301S
Applicant
Asahi Intecc Co., Ltd.
Product Code
DQX · Cardiovascular
Decision Date
Sep 22, 2010
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
The ASAHI SION PTCA Guide Wire is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI SION PTCA Guide Wire is not to be used in the cerebral blood vessel.
Device Story
Steerable guide wire (0.014 inch diameter; 180/300 cm lengths) used to facilitate balloon dilatation catheter placement during PTCA/PTA procedures. Constructed with stainless steel core, platinum-nickel/stainless steel coils, and safety wire; distal tip is radiopaque for visualization; hydrophilic coating on distal portion; PTFE coating on proximal section. Operated by physicians in clinical settings to navigate vasculature. Device provides mechanical support and guidance for interventional catheters. Benefits include improved trackability and navigation within coronary and peripheral vessels.
Clinical Evidence
No clinical data. Evidence consists of in vitro bench testing including tensile strength, torque strength, torqueability, tip flexibility, coating adherence, catheter compatibility, and integrity (particulate testing).
Technological Characteristics
Stainless steel core wire; platinum-nickel and stainless steel coils; soldered safety wire. Hydrophilic distal coating; PTFE proximal coating. 0.014 inch diameter. Sterilized. Mechanical device; no software or energy source.
Indications for Use
Indicated for facilitating balloon dilatation catheter placement during PTCA and PTA procedures. Contraindicated for use in cerebral blood vessels.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
K062186 — ASAHI PTCA GUIDE WIRE, FIELDER J, MODEL AGP140000J, AGP140300J · Asahi Intecc Co., Ltd. · Nov 9, 2006
K063819 — ASAHI PTCA GUIDE WIRE, FIELDER FC · Asahi Intecc Co., Ltd. · Jan 26, 2007
K041531 — ASAHI PTCA GUIDE WIRE CONFIANZA PRO · Asahi Intecc Co., Ltd. · Aug 3, 2004
Submission Summary (Full Text)
{0}------------------------------------------------
KIDG578
#### 9.0 510(K) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | Asahi Intecc Co., Ltd.<br>1703 Wakita-cho, Moriyama-ku<br>Nagoya, Aichi 463-0024<br>Japan |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| OFFICIAL<br>CORRESPONDENT | Yoshi Terai<br>President, CEO<br>Asahi Intecc USA, Inc.<br>2500 Red Hill Avenue, Suite 210<br>Santa Ana, CA 92705<br>Tel: (949) 756-8252<br>FAX (949) 756-8165<br>e-mail: yoshi.terai@asahi-intecc.com<br>SEP 2.2 2010 |
| TRADE NAME: | ASAHI SION PTCA Guide Wire |
| COMMON NAME: | Guide Wire |
| CLASSIFICATION<br>NAME: | Wire, Guide, Catheter |
| DEVICE<br>CLASSIFICATION: | Class 2 per 21 CFR §870.1330 |
| PRODUCT CODE | DQX |
| PREDICATE DEVICE: | Asahi - SUOH PTCA Guide Wire - 510(k) K083904<br>Asahi -Asahi PTCA Guide Wire - 510(k) K070945<br>Asahi - JoWire Neo's PTCA Guide Wire - 510(k) K022762 |
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: ·
The ASAHI SION PTCA Guide Wire is a steerable guide wire with a maximum diameter of 0.014 inches (0.36mm) and available in 180 cm and 300 cm lengths. The extension wire is connected to the end of the guide wire outside the body. The guide wire is constructed from a stainless steel core wire with platinum-nickel and stainless steel coils. The coil assembly consists of an inner coil and an outer coil, and there is a safety wire for which is soldered to the inner and outer coils and the core wire. The distal end of the guide wire has a radiopaque tip to achieve visibility, and is available in a straight configuration and can be made to bend easily with the vessel curve. A hydrophilic coating is applied to the distal portion of the guide wire. The proximal section of the quide wire is coated with PTFE.
- 9.
{1}------------------------------------------------
### Indication for Use:
The ASAHI SION PTCA Guide Wire is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The Asahi SION PTCA Guide Wire is not to be used in the cerebral blood vessel.
# TECHNICAL CHARACTERISTICS:
Comparisons of the ASAHI SION PTCA Guide Wire and predicate devices show that the technological characteristics such as product performance, design and intended use are substantially equivalent to the current marketed predicate devices.
The ASAHI SION PTCA Guide Wire is similar in design - device dimensional specifications and manufacturing process, and intended use, operating principle, shelf life and sterilization process are the same and materials that have been used in other predicate devices in that its core wire, tip coils and solders remain the same.
# PERFORMANCE DATA:
All components that come in direct contact with the patient have a long history of use in medical devices and are proven to be biocompatible for use in the vasculature. Furthermore, this submission contains reference to a predicate ASAHI device that uses the same materials as used in the subject device. In vitro bench testing, including tensile strength, torque strength, torqueability, tip flexibility, coating adherence, catheter compatibility and integrity (particulate testing) etc were conducted on the ASAHI SION PTCA Guide Wire, This 510(k) notice includes mechanical and functional bench testing that demonstrates that the ASAHI SION PTCA Guide Wire performs.as intended.
### SUMMARY/CONCLUSION:
The ASAHI SION PTCA Guide Wire characteristics are substantially equivalent to the specified predicate devices and other currently marketed devices for the same indication for use.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Asahi Intecc Co. Ltd. c/o Mr. Yoshi Terai President, CEO Asahi Intecc USA, Inc. 2500 Red Hill Avenue, Suite 210 Santa Ana, CA 92705
SEP 2222010
Re: K100578
Trade/Device Name: ASAHI SION PTCA Guide Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II (two) Product Code: DQX Dated: September 17, 2010 Received: September 20, 2010
Dear Mr. Terai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Yoshi Terai
regulation (21 CFR Part 803), please go to
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D/Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
INDICATIONS FOR USE STATEMENT
INDICATIONS FOR USE STATEMENT 2.0
510(k) Number (if known): _ _ 100578
Device Name: ASAHI SION PTCA Guide Wire
SEP 22 2010
1100528
ي
Indications for Use:
The ASAHI SION PTCA Guide Wire is intended to facilitate the placement of balloon dilatation catheters during percutaneous transluminal coronary angioplasty (PTCA) and percutaneous transluminal angioplasty (PTA). The ASAHI SION PTCA Guide Wire is not to be used in the cerebral blood vessel.
Prescription Use_X____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _
of
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CD, Office of Device/Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K100578
Page of
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.