PV 1900 SYNCHRO 0.016 GUIDEWIRE

K050964 · Boston Scientific - Precision Vascular · DQX · Jun 13, 2005 · Cardiovascular

Device Facts

Record IDK050964
Device NamePV 1900 SYNCHRO 0.016 GUIDEWIRE
ApplicantBoston Scientific - Precision Vascular
Product CodeDQX · Cardiovascular
Decision DateJun 13, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1330
Device ClassClass 2

Indications for Use

The PV 1900 Synchro® 0.016" Neuro Guidewire series of products is intended for peripheral and neurovascular use. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures.

Device Story

PV 1900 Synchro 0.016" guidewire is a sterile, single-use device designed for navigating small, tortuous peripheral and neurovasculature. It features a shapeable Nitinol distal tip and a platinum marker coil for fluoroscopic visualization. The device is coated with a lubricious material to facilitate navigation. It is operated by physicians trained in percutaneous, intravascular procedures. The guidewire is torqued manually to navigate through the vasculature; a torque device is provided to assist in deployment and positioning. An introducer is included to aid insertion into catheter hubs or hemostasis valves. The device facilitates the exchange and positioning of interventional catheters. By providing steerability and torque transmission, it enables access to distal vascular sites, potentially improving the success of diagnostic and interventional procedures.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included dimensional inspection, fatigue assessment, tip flexibility, tip shaping, tensile strength, guidewire compatibility, torsional strength, torqueability, and EO sterilization validation in accordance with ISO 11070:1998 and ISO 10993-1.

Technological Characteristics

0.016" diameter guidewire; Nitinol distal tip (25-35cm); platinum marker coil; lubricious outer coating. Materials verified per ISO 10993-1 for short-term blood contact. Sterilization via ethylene oxide (EO). Manual operation; torque-transmitting design.

Indications for Use

Indicated for general intravascular use, including neuro and peripheral vasculature, to selectively introduce and position catheters and other interventional devices. For use by physicians trained in percutaneous, intravascular techniques.

Regulatory Classification

Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ **K050964** 1 of 2 ## JUN 1 3 2005 ## Summary of Safety and Effectiveness | Amended May 23, 2005 | | | | | |------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|---------------------------------| | General<br>Provisions | Submitter of 510(k)<br>Premarket Notification: | Boston Scientific - Precision Vascular<br>2405 West Orton Circle<br>West Valley City, UT 84119<br>Phone: 801.974.1700<br>Fax: 801.974.1740 | | | | | Contact Person: | Rick Gaykowski<br>VP, Regulatory/Clinical Affairs<br>Quality Systems & Operations | | | | | Device Trade Name:<br>Device Generic Name: | PV 1900 Synchro® 0.016" Guidewire<br>Guide Wire | | | | The predicate devices are listed in the table below. | | | | | | | Device | Manufacturer | 510(k) Number/Date | Pro Code | | Predicate<br>Devices | Synchro™ .014"<br>Synchro™ .010"<br>Transend<br>Agility-16<br>0.016" Guidewire<br>* | Precision Vascular<br>Precision Vascular<br>Target Therapeutics<br>Cordis<br>Terumo<br>This information is assumed based upon our best knowledge. | K032146, 12 Aug 2003<br>K032146, 12 Aug 2003<br>K971254, 01 Jul 1997*<br>K010511, 01 Mar 2001*<br>K913074, 05 Nov 1991* | DQX<br>DQX<br>DQX<br>DQX<br>DQX | | Classification | Class II, 21 CFR 870.1330, Wire, Guide, Catheter 74DQX | | | | | Performance<br>Standards | Performance standards have not been established by FDA under section 514 of the<br>Federal Food, Drug and Cosmetic Act. | | | | | Intended Use | The PV 1900 Synchro® 0.016" Neuro Guidewire series of products is intended for<br>peripheral and neurovascular use. It can be used to selectively introduce and position<br>catheters and other interventional devices within the peripheral and neurovasculature.<br>This device should be used only by physicians trained in percutaneous, intravascular<br>techniques and procedures. | | | | | Device<br>Description | The PV 1900 is a member of the Synchro® Guidewire family of products having a<br>0.016" outside diameter, being a sterile, single use/disposable product, with a<br>shapeable tip which is used to gain intravascular access to and facilitate the positioning<br>and exchange of interventional devices in small diameter, tortuous vasculature for<br>peripheral and neuro diagnostic and interventional procedures. The guidewire can be<br>torqued to facilitate navigation through the vasculature. A torque device, (Boston<br>Scientific (K903606)) is supplied with the wire to facilitate deployment & positioning. A<br>guidewire introducer (B. Braun (K760389)) is also supplied and may be used to aid<br>introduction of the guidewire into the catheter hub and/or hemostasis valve and to | | | | {1}------------------------------------------------ | | gently shape the guidewire's distal flexible tip, if desired, according to standard practice. Neither the guidewire introducer nor the torque device are intended to enter the body. The product is projected to be provided in a 100cm - 300cm length range with 140-180cm being nominal. The Nitinol tip length is projected to be presented in a 25cm - 35cm range. A traditional range of flexibility profiles shall also be provided ranging from support (stiff) to flex (soft). The device is coated on the outer diameter with a lubricious coating over the distal portion of the device. The marker coil is platinum wire at the distal tip of the device to aid visualization under fluoroscopy. The subject device has the ability to access distal, tortuous vasculature, with steerability and torque transmission properties. | |------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological<br>Characteristics | Technological similarities between the PV 1900 Synchro® 0.016" Guidewire and the PVS 1300 Synchro™ predicate remain identical. This is also true for competitive predicate device features including the basal design and dimensions, generic materials & construction, and hydrophilic coating. There are no new questions raised regarding safety or efficacy of the PVS 1300 Synchro® 0.014" Guidewire. | | | Biocompatibility of the PV 1900 Synchro® 0.016" Guidewire materials have been verified in accordance with ISO 10993-1, Biological Evaluation of Medical Devices - Part 1. Materials test results confirmed biocompatibility of the subject device when tested as an external communicating, blood contact, short duration (<24 hours) device. | | Safety and<br>Performance<br>Tests | Performance testing of materials comprising the PV 1900 Synchro® 0.016" Guidewire was conducted in accordance with ISO 11070:1998, Sterile, Single-Use Intravascular Catheter Introducers. Verification testing for the subject device included dimensional inspection, fatigue assessment, tip flexibility, tip shaping, tensile strength, guide wire compatibility testing and performance under simulated conditions. Subject product testing has yielded acceptable safety & performance outcomes. | | | In addition, torsional strength, torqueability, and EO sterilization tests also yielded acceptable results. The results of these tests, in conjunction with the substantial equivalence claims as outlined in the premarket notification, effectively demonstrate the PV 1900 Synchro® 0.016" Guidewires' substantial equivalence to the cited predicate devices. | | Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and safety and performance testing, the subject PV 1900 Synchro® 0.016" Guidewire meets the minimum requirements that are considered adequate for its intended use and is substantially equivalent in design, materials, sterilization, principles of operation and indications for use to current commercially available guidewires/cited predicates. | | | | . {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 1 3 2005 Boston Scientific - Precision Vascular c/o Mr. Rick Gaykowski V.P. Regulatory/Clinical Affairs Quality Systems & Operations 2405 West Orton Circle West Valley City, UT 84119 Re: K050964 PV 1900 Synchro® 0.016" Guidewire Regulation Number: 21 CFR 870.1330 Regulation Name: Guidewire Regulatory Class: Class II Product Code: DQX Dated: May 23, 2005 Received: May 25, 2005 Dear Mr. Gaykowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Rick Gaykowski Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Donna R. Wchner Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K050964 Device Name (Generic): PV 1900 Synchro® 0.016" Guidewire The PV 1900 Synchro® 0.016" Guidewire series of Indications For Use: products are intended for general intravascular use, including the neuro and peripheral vasculature. It can be used to selectively introduce and position catheters and other interventional devices within the peripheral and neurovasculature. This device should be used only by physicians trained in percutaneous, intravascular techniques and procedures. × Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use_ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Duma R. Vacher (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number_K6509664 Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...