The PeriVac Kit is intended for use in pericardial aspiration and drainage in the presence of pericardial effusion or tamponade.
Device Story
PeriVac Kit is a pre-assembled, single-use convenience kit containing 21 components for emergency pericardial aspiration and drainage. Components include needles, guidewire, dilator, and catheter for site preparation, anesthesia, puncture, and fluid collection. Used in clinical settings (Special Procedures Laboratory, Cardiac Catheterization Laboratory, or bedside) by trained physicians. Procedure involves entering the pericardium via progressively larger needles, guidewire, dilator, and catheter; fluid is drained into a graduated bag. Catheter is withdrawn and entry site closed upon completion. Kit facilitates rapid initiation of drainage to relieve tamponade. Output is the physical removal of pericardial fluid. Benefits include convenience, reduced preparation time, and standardized equipment for emergency cardiac procedures.
Clinical Evidence
Bench testing only. Non-clinical tests conducted showed the device met its design-input criteria and was safe and effective for its intended use. No clinical data provided.
Technological Characteristics
Convenience kit containing 21 components for pericardial access. Single-use, sterile (ETO), non-pyrogenic. Includes needles, guidewire, dilator, and catheter. No power source. Packaging consists of thermoform tray, CSR wrap, and header pouch. 2-year expiration.
Indications for Use
Indicated for patients requiring pericardial aspiration and drainage due to pericardial effusion or tamponade.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
# APR 3 0 2004
#### 510(k) Summary of Safety and Effectiveness 3.
#### General Information a.
## Modified Device Information
| Category: | Comments: |
|--------------------------|-------------------------------|
| Sponsor: | Boston Scientific Corporation |
| | 2710 Orchard Parkway |
| | San Jose, Ca 95134 |
| | |
| Correspondent: | Ronald C. Allen, Ph. D. |
| | Manager, Regulatory Affairs |
| | Boston Scientific Corporation |
| | 2710 Orchard Parkway |
| | San Jose, Ca 95134 |
| Contact Information: | E-mail: allenr@bsci.com |
| | Phone: (408) 895-3670 |
| | Fax: (408) 895-2202 |
| Device Common Name: | Pericardiocentesis Kit |
| Device Proprietary Name: | PeriVac Kit |
| Device Classification: | Class II |
## Predicate Device Information
| Predicate Device: | PeriVac Kit |
|-----------------------------------------|-------------------------------|
| Predicate Device Manufacturer: | Boston Scientific Corporation |
| Predicate Device Common Name | Pericardiocentesis Kit |
| Predicate Device Classification: | Class II |
| Predicate Device Classification Number: | 21 CFR $$870.1330$ |
{1}------------------------------------------------
## b. Date Summary Prepared
March 28, 2004
#### Description of Device C.
The PeriVac Kit is designed to provide all of the necessary components for site preparation, local anesthesia, puncture, aspiration and/or drainage, collection, and dressing for the completion of pericardial fluid aspiration. The kit is comprised of finished components and packaged to provide convenience for the user.
The PeriVac Kit is comprised of 21 components pre-assembled and packaged to eliminate delays when initiating emergency pericardial aspiration of fluid. For customer convenience, the PeriVac Kit will be made available in four packaging configurations with the only difference between the kits being the catheter types and the inclusion of Lidocaine and Providone-Iodine solution. This submission addresses two components of the PeriVac Kit.
## d. Intended Use
The intended use for the PeriVac Kit is as follows:
The PeriVac Kit is intended for use in pericardial aspiration and drainage in the presence of pericardial effusion and tamponade.
OR
#### Comparison to Predicate Device e.
See Table I- Comparison of Device Characteristics to Predicate on the following page.
{2}------------------------------------------------
# Table 1 - Comparison of Device Characteristics to Predicate
| | CLEARED KIT | PREDICATE KIT | |
|------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FEATURE | PeriVac Kit | Mansfield Kit | Current Submission |
| 510(k) Reference<br>Number | K032050 | Pre-Amendment | |
| Product Code | DQX | - | DQX |
| Classification Section | 21CFR 870.1330 | 21CFR 870.1330 | 21CFR 870.1330 |
| Classification Name | Convenience Kit | Convenience Kit | Convenience Kit |
| Indications for Use | The Pericardiocentesis Kit is<br>intended for use in pericardial<br>aspiration and drainage in the<br>presence of pericardial effusion<br>or tamponade. | The Pericardiocentesis Kit is<br>intended for use in pericardial<br>aspiration and drainage in the<br>presence of pericardial effusion<br>or tamponade. | The Pericardiocentesis Kit is<br>intended for use in pericardial<br>aspiration and drainage in the<br>presence of pericardial effusion or<br>tamponade. |
| Anatomical Sites | Pericardium | Pericardium | Pericardium |
| Method of Use | The Pericardium is entered via a<br>series of progressively larger<br>positioning of needles,<br>guidewire, dilator, and catheter<br>positioning. Once confirmed in<br>position, fluid contained in the<br>pericardium is drained into a<br>graduated bag. Upon completion<br>the catheter is withdrawn and<br>the entry site closed. | The Pericardium is entered via a<br>series of progressively larger<br>positioning of needles,<br>guidewire, dilator, and catheter<br>positioning. Once confirmed in<br>position, fluid contained in the<br>pericardium is drained into a<br>graduated bag. Upon completion<br>the catheter is withdrawn and<br>the entry site closed. | Same |
| Power Source | None | None | None |
| Sterilization Method | ETO | ETO | ETO |
| | CLEARED KIT | PREDICATE KIT | |
| FEATURE | PeriVac Kit | Mansfield Kit | Current Submission |
| Pyrogen Free | Non-pyrogenic | Non-pyrogenic | Non-pyrogenic |
| Expiration Date<br>(years) | Two year | One year | Two year |
| Single Use Only | Yes | Yes | Yes |
| Packaging | Thermoform Tray, Wrapped in<br>CSR Wrap, and sealed in<br>Header Pouch | Thermoform Tray, Wrapped in<br>CSR Wrap, and sealed in<br>Header Pouch | Thermoform Tray, Wrapped in<br>CSR Wrap, and sealed in<br>Header Pouch |
| Packaging Contents | The 21 components provided in<br>the PeriVac Kit are presented in<br>Table 4. | The 17 components provided in<br>the Mansfield kit are presented<br>on the product label (Appendix<br>D). | The 21 components provided in<br>the PeriVac Kit are presented in<br>Table 4. |
| Contraindications | There are no known<br>contraindications for pericardial<br>aspiration, although recurrent<br>effusion or unresolved<br>tamponade may warrant surgical<br>intervention. | There are no known<br>contraindications for pericardial<br>aspiration, although recurrent<br>effusion or unresolved<br>tamponade may warrant surgical<br>intervention. | There are no known<br>contraindications for pericardial<br>aspiration, although recurrent<br>effusion or unresolved<br>tamponade may warrant surgical<br>intervention. |
| | | | |
| Warnings and Precautions | Inspect all devices and<br>equipment prior to use. Ensure that sterile barrier has<br>not been breached. Careful attention to aseptic<br>technique should be<br>employed. This kit should be used only by<br>persons thoroughly trained in<br>the techniques of pericardial<br>centesis. Take proper care to ensure<br>that patient-contact electrical<br>equipment is properly isolated<br>and grounded. STERILE. For one procedure<br>only. Do not resterilize. CT scan, fluoroscopic, or<br>echocardiographic<br>examinations are<br>recommended to evaluate<br>needle and catheter<br>placement. Pericardiocentesis should be<br>carried out in the Special<br>Procedures Laboratory or<br>Cardiac Catherization<br>Laboratory, utilizing equipment<br>capable of cardiac monitoring.<br>When performed at bedside,<br>electrocardiographic<br>monitoring should be<br>employed continuously. | Inspect all devices and<br>equipment prior to use. Take<br>proper care to ensure that all<br>patient-contact electrical<br>equipment is properly isolated<br>and grounded. Do not wipe<br>catheter with organic solvents<br>(e.g. alcohol, ethers, esters,<br>phenols, etc.). Careful<br>attention to aseptic technique<br>should be employed. | Inspect all devices and<br>equipment prior to use. Ensure that sterile barrier has<br>not been breached. Careful attention to aseptic<br>technique should be<br>employed. This kit should be used only by<br>persons thoroughly trained in<br>the techniques of pericardial<br>centesis. Take proper care to ensure<br>that patient-contact electrical<br>equipment is properly isolated<br>and grounded. STERILE. For one procedure<br>only. Do not resterilize. CT scan, fluoroscopic, or<br>echocardiographic<br>examinations are<br>recommended to evaluate<br>needle and catheter<br>placement. Pericardiocentesis should be<br>carried out in the Special<br>Procedures Laboratory or<br>Cardiac Catherization<br>Laboratory, utilizing equipment<br>capable of cardiac monitoring.<br>When performed at bedside,<br>electrocardiographic<br>monitoring should be<br>employed continuously. Replace the straight catheter<br>with pigtail catheter (sold<br>separately) if drainage is<br>required for up to 24 hours. Do<br>not leave the pigtail catheter in<br>longer than 24 hours. Secure the pigtail catheter to<br>the skin and apply a sterile<br>dressing. At the time of catheter |
| Boston Scientific Corporation<br>Special 510(k) Submission | | CONFIDENTIAL | |
- 12 -Boston Scientific Corporation
Special 510(k) Submission
CONFIDENTIAL
Lagund)
{3}------------------------------------------------
# SPECIAL 510(K) SUBMISSION PeriVac Kit
# CONFIDENTIAL
9 30 h (480) 07 100 100 100 100 100 100 100 1000 1000 1000 1000 1000 100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000
13
{4}------------------------------------------------
{5}------------------------------------------------
#### Summary of the Non-clinical Data f.
Where appropriate, testing conformed to the requirements of 21 CFR Part 58 (Good Laboratory Practices (GLP)). Specifically, non-clinical tests conducted for the Device showed the device met its design-input criteria, and was safe and effective for its intended use.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird with three stylized wing segments.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 3 0 2004
Boston Scientific Corp. c/o Dr. Ronald Allen EP Technologies, Inc. 2710 Orchard Parkway San Jose, CA 95134
Re: K040867
PeriVac Kit. models 4304. 4305. 4314. and 4315 Regulation Number: 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: March 28, 2004 Received: April 2, 2004
### Dear Dr. Allen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market the device, subject to the general controls provisions of Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
{7}------------------------------------------------
Page 2 - Dr. Ronald Allen
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
In addition, we have determined that your device kit contains 5 ml of 1% HCL Lidocaine which are subject to regulation as a drug.
Our substantially equivalent determination does not apply to the drug components of your device. We recommend you first contact the Center for Drug Evaluation and Research before marketing your device with the drug components. For information on applicable Agency requirements for marketing this drug, we suggest you contact:
> Director, Division of Drug Labeling Compliance (HFD-310) Center for Drug Evaluation and Research Food and Drug Administration 5600 Fishers Lane Rockville, Maryland 20857 (301) 594-0101
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation, please contact the Office of Compliance at (301) 594-4648. Also, please note the regulation entitled, "Misbranding by reference to premarket notification'(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers.
{8}------------------------------------------------
Page 3 – Dr. Ronald Allen
International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/dsma/dsmamain.html
Sincerely yours,
Donna R. Lochner
(1) Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Premarket Notification -Indication for Use Statement
Device Name: PeriVac Kit
Indication for Use:
The intended use for PeriVac Kit is as follows:
The PeriVac Kit is intended for use in pericardial aspiration and drainage in the presence of pericardial effusion or tamponade.
Prescription Use______________________________________________________________________________________________________________________________________________________________
Over-the-Counter Use_
(Per 21 CFR §801.109)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IS NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Duana R. Lochner
CHOVE
510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.