← Product Code [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX) · K030617

# PT2 GUIDE WIRE AND ADDWIRE EXTENSION WIRE (K030617)

_Boston Scientific Scimed, Inc. · DQX · May 21, 2003 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K030617

## Device Facts

- **Applicant:** Boston Scientific Scimed, Inc.
- **Product Code:** [DQX](/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX.md)
- **Decision Date:** May 21, 2003
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The PT2TM Guide Wires are intended to facilitate the placement of balloon dilatation catheters or other therapeutic devices during PTCA, PTA, or other intravascular interventional procedures. The PT2TM Guide Wires are not intended for use in the cerebral vasculature. The AddWireTM Extension Wire creates an extended guide wire that can be used to exchange out a therapeutic device without removing the original guide wire from the anatomy. When the exchange is complete, the Extension Wire can be detached and the original guide wire can be used in a conventional manner.

## Device Story

PT2 Guide Wire: steerable, polymer-tipped wire; 0.014-inch diameter; Light/Moderate support models; 185/300 cm lengths; radiopaque distal tip (straight or J-tip). Construction: PTFE-coated stainless steel core coupled to tapered Nitinol core; stainless steel coupler; distal stainless steel ribbon; Tungsten-loaded polyurethane sleeve; ICE hydrophilic coating. AddWire Extension Wire: 145 cm length; 0.014-inch diameter; PTFE-coated stainless steel wire with distal connector. Used in clinical interventional settings (PTCA/PTA) by physicians to navigate vasculature and exchange therapeutic devices without removing the primary guide wire. Benefits include procedural efficiency during catheter exchanges. Non-clinical bench testing confirms mechanical integrity and compatibility.

## Clinical Evidence

Bench testing only. Evaluations included tensile strength, combined load, coupler fatigue, torque response, tip prolapse, coating adherence, polymer peel, PTCA catheter compatibility, and biocompatibility. Results verified device meets all minimum requirements.

## Technological Characteristics

Materials: PTFE-coated stainless steel, Nitinol, Tungsten-loaded polyurethane, stainless steel coupler. Design: 0.014-inch diameter, steerable, hydrophilic coating. Extendable proximal section on 185-cm models. Sterilized, single-use.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- PT Graphix ([K962572](/device/K962572.md))
- Balance MiddleWeight ([K021228](/device/K021228.md))
- BMW ([K983033](/device/K983033.md), [K973494](/device/K973494.md), [K971815](/device/K971815.md))
- Extension Wire ([K970376](/device/K970376.md))
- Steerable PTCA GW ([K011968](/device/K011968.md), [K970376](/device/K970376.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K030617

MAY 21 2003

Abbreviated 510(k)
BSC PT2TM
Guide Wire BSC AddWire™ Extension Wire February 25, 2003

## 510(k) Summary per 21 CFR §807.92

| Submitter's Name<br>and Address | Boston Scientific Corporation (BSC)<br>One Scimed Place<br>Maple Grove, MN 55311                                           |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Contact Name and<br>Information | Anne V. Rossi<br>Specialist, Regulatory Affairs<br>Phone: 763-494-2827<br>Fax: 763-494-2323<br>e-mail: Anne.Rossi@bsci.com |
| Date Prepared                   | February 25, 2003                                                                                                          |
| Proprietary<br>Name(s)          | PT2TM Guide Wire<br>AddWireTM Extension Wire                                                                               |
| Common Name                     | Catheter Guide Wire                                                                                                        |
| Product Code                    | 74DQX                                                                                                                      |
| Classification of<br>Device     | Class II, 21 CFR Part 870.1330                                                                                             |
| Predicate Devices               |                                                                                                                            |
| BSC Scimed                      |                                                                                                                            |
|                                 | <i>PT GraphixTM</i> K962572 (Clearance Date December 17, 1996)                                                             |
|                                 | Advanced Cardiovascular Systems, Inc. (ACS)                                                                                |
|                                 | <i>Balance<br/>MiddleWeightTM</i> K021228 May 15, 2002                                                                     |
|                                 | (BMW) K983033 November 10, 1998                                                                                            |
|                                 | K973494 December 12, 1997                                                                                                  |
|                                 | K971815 July 9, 1997                                                                                                       |
|                                 | Lake Region Manufacturing, Inc. (LRM)                                                                                      |
|                                 | <i>Extension Wire1</i> K970376 June 6, 1997                                                                                |
|                                 | <i>Steerable PTCA GW2</i> K011968 July 25, 2001                                                                            |
|                                 | K970376 June 6, 1997                                                                                                       |

<sup>1</sup> The LRM Extension Wire is currently marketed by Boston Scientific under the trade name Trooper™/Patriot™ Extension Wire.

<sup>2</sup> The LRM Steerable PTCA Guide Wire is currently marketed by Boston Scientific under the trade name Forté™ Guide Wire.

{1}------------------------------------------------

K030617

Abbreviated 510(k)
BSC AddWire™ BSC PT2TM Guide Wire
BSC AddWire™™ Extension Wire
BSC AddWire™™ Extension Wire
February 25, 2013

|                                                  | February 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description -<br>Guide Wire            | The PT2TM Guide Wires with ICE® hydrophilic coating are steerable,<br>polymer-tipped guide wires. The PT2TM is available in a nominal<br>diameter of 0.014 inches, two rail support models, Light Support and<br>Moderate Support, and nominal lengths of 185 and 300 centimeters<br>with Brachial and Femoral marks at 90cm and 100cm respectively.<br>The distal two centimeters of all models are radiopaque and available<br>in either a straight shapeable or a pre-formed J-Tip.<br>The 185-centimeter version of the PT2TM is designed with a proximal<br>extension section. This proximal extension section allows connection<br>to the AddWireTM Extension Wire or the similarly designed<br>TrooperTM/PatriotTM Extension Wire.<br>The PT2TM corewire consists of a PTFE coated SS corewire segment<br>coupled to a tapered Nitinol corewire segment. The PT2TM coupler is<br>a stainless steel material. The distal most end of the PT2TM corewire is<br>a SS ribbon.<br>The distal tapered Nitinol corewire segment of the PT2TM is coated<br>with an adhesive pre-coat. The pre-coated corewire is then jacketed<br>with a Tungsten loaded polyurethane sleeve. The entire polymer<br>sleeve of the PT2TM Guide Wire is coated with ICE® Hydrophilic<br>coating. |
| Intended Use of<br>Device - Guide<br>Wire        | The PT2TM Guide Wires are intended to facilitate the placement and<br>exchange of balloon dilatation catheters or other therapeutic devices<br>during PTCA or PTA or other intravascular interventional procedures.<br>The PT2TM Guide Wires are not intended for use in the cerebral<br>vasculature. The devices are provided non-pyrogenic, sterile, and<br>intended for one procedure only.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Technological<br>Characteristics -<br>Guide Wire | The PT2TM Guide Wires utilize similar materials and methods of<br>construction as the currently marketed PT GraphixTM family of guide<br>wires. The differences in construction are a minor change to the<br>polymer sleeve formulation and the addition of a 2-part core wire.<br>The 185-cm models also incorporate an extendable proximal section<br>similar in construction to the BSC FortéTM extendable guide wires.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Non-Clinical Test<br>Summary - Guide<br>Wire     | Testing and evaluation of the PT2TM Guide Wires included tensile,<br>combined load, coupler fatigue, torque response, tip prolapse, coating<br>adherence, visual inspection, polymer peel, PTCA catheter<br>compatibility, and biocompatibility.<br>Test results verified that the PT2TM Guide Wires met all of the<br>minimum requirements and are adequate for their intended use.<br>The PT2TM Guide Wires are considered to be substantially equivalent<br>to guide wires currently marketed by Boston Scientific and ACS based<br>on a comparison of intended use, design and the results of in vitro<br>testing and evaluation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{2}------------------------------------------------

KO30617

Abbreviated 510(k)
BSC PT2T" GSC PT2T" Guide Wire
BSC AddWire™ Extension Wire
BSC AddWire™ Extension Wire
February 25, 2005

| Device<br>Description -<br>Extension Wire            | The AddWireTM Extension Wire is a PTFE coated SS wire with a<br>connector at the distal end that connects to the proximal end of BSC<br>extendable guide wires. The nominal length is 145 cm and the<br>nominal diameter is 0.014 inches. The AddWireTM Extension Wire is<br>similar in design and materials to the LRM Extension Wire that is<br>currently marketed by BSC as the TrooperTM/PatriotTM Extension<br>Wire.                                                                                                                                                                                    |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use of<br>Device -<br>Extension Wire        | The AddWireTM Extension Wire creates an extended guide wire that<br>can be used to exchange out a therapeutic device without removing the<br>original guide wire from the anatomy. When the exchange is<br>complete, the Extension Wire can be detached and the original guide<br>wire can be used in a conventional manner.                                                                                                                                                                                                                                                                                 |
| Technological<br>Characteristics -<br>Extension Wire | The AddWireTM Extension Wires utilize similar materials and methods<br>of construction as the LRM Extension Wire that is currently marketed<br>by BSC as the TrooperTM/PatriotTM Extension Wire.                                                                                                                                                                                                                                                                                                                                                                                                             |
| Non-Clinical Test<br>Summary -<br>Extension Wire     | Testing and evaluation of the AddWireTM Extension Wires included<br>tensile, coating adherence, visual inspection, PTCA catheter<br>compatibility, exchange system connectability, and biocompatibility.<br>Test results verified that the AddWireTM Extension Wire met all of the<br>minimum requirements and are adequate for their intended use.<br>The AddWireTM Extension Wire is considered to be substantially<br>equivalent to the currently marketed BSC TrooperTM/PatriotTM<br>Extension wire based on a comparison of intended use, design and the<br>results of in vitro testing and evaluation. |

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three curved lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 21 2003

Ms. Anne V. Rossi Specialist, Regulatory Affairs Boston Scientific Corporation One Scimed Place Maple Grove, MN 55311

K030617 Re:

> Trade/Device Name: PT2TM Guide Wire and AddWire™ Extension Wire Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire. Regulatory Class: II Product Code: DQX Dated: February 25, 2003 Received: February 26, 2003

Dear Ms. Rossi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application(PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{4}------------------------------------------------

Page 2 - Ms. Anne V. Rossi

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours.

laatu

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

KO 30617

## Indications for Use Statement

| 510(k) Number<br>(if known) | Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name(s)              | PT2TM Guide Wire<br>AddWireTM Extension Wire                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Indications For<br>Use      | The PT2TM Guide Wires are intended to facilitate the<br>placement of balloon dilatation catheters or other<br>therapeutic devices during PTCA, PTA, or other<br>intravascular interventional procedures. The PT2TM Guide<br>Wires are not intended for use in the cerebral vasculature.<br><br>The AddWireTM Extension Wire creates an extended guide<br>wire that can be used to exchange out a therapeutic<br>device without removing the original guide wire from the<br>anatomy. When the exchange is complete, the Extension<br>Wire can be detached and the original guide wire can be<br>used in a conventional manner. |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR §801.109)

Confidential and Proprietary to Boston Scientific Corporation

---

**Source:** [https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K030617](https://fda.innolitics.com/submissions/CV/subpart-b%E2%80%94cardiovascular-diagnostic-devices/DQX/K030617)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
